1 / 46

Craig Moore Radiation Physics Service Hull & East Yorkshire Hospitals NHS Trust hullrad.uk

Optimisation of Computed Radiography chest imaging utilising a digitally reconstructed radiograph simulation technique. Craig Moore Radiation Physics Service Hull & East Yorkshire Hospitals NHS Trust www.hullrad.org.uk. Introduction - Literature.

aneko
Download Presentation

Craig Moore Radiation Physics Service Hull & East Yorkshire Hospitals NHS Trust hullrad.uk

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Optimisation of Computed Radiography chest imaging utilising a digitally reconstructed radiograph simulation technique Craig Moore Radiation Physics Service Hull & East Yorkshire Hospitals NHS Trust www.hullrad.org.uk

  2. Introduction - Literature • Lots of publications have shown that patient anatomy is the limiting factor in the reading normal structures, and detection of lesions (lung nodules) in chest images • Bochud et al 1999 • Samei et al 1999, 2000 • Burgess et al 2001 • Huda et al 2004 • Keelan et al 2004 • Sund et al 2004 • Tingberg et al 2004 • European wide RADIUS chest trial (2005)

  3. Introduction - Literature • Chest radiography is now generally considered to be limited by the projected anatomy • Patient anatomy = anatomical noise So if we want to optimize digital system for chest imaging, vital that anatomical noise is present in the images!!!

  4. Introduction • However, the radiation dose/image quality relationship must not be ignored • Doses must be kelp ALARP • ICRP 2007 • IR(ME)R2000 – (required legally in the UK) • We would therefore want system (quantum) noise present in an image for dose reduction studies

  5. Digitally Reconstructed Radiograph (DRR) • Hypothesis: • Can use CT data of humans to provide realistic anatomy (anatomical noise) • Clinically realistic computerized ‘phantom’ • Simulate the transport of x-rays through the ‘phantom’ and produce a digitally reconstructed radiograph (DRR – a simulation of a conventional 2D x-ray image created from CT data) • Add frequency dependent system noise post DRR calculation • Add radiation scatter post DRR calculation • Validate • Use for optimization studies

  6. DRR Algorithm: Virtual Patient • Virtual patient derived from chest portion of real CT datasets • Voxel resolution = 0.34 x 0.34 x 0.8 mm • CT number converted to linear attenuation coefficient (LAC) using tissue equivalent inserts • Measure mean CT No. in each insert • We know elemental composition of each so can derive LAC • Can derive relationship between CT No. and LAC

  7. Energy absorbed in CR phosphor Final CT axial batch (i.e. slices 681 to 700) CT axial batch 2 (i.e. CT slices 21 to 40) PA slice N CT axial batch 1 (i.e. CT slices 1 to 20) PA slice 1 CT dataset re-orientated in the ‘PA’ direction Intensity of X-rays exiting is calculated X-ray attenuated exponentially through CT dataset using a ray casting method of DRR calculation X-ray spectra derived from IPEM 78

  8. Scatter Addition • Must add to DRR generated image • DRR algorithm does not calculate scatter • Measured scatter in CR chest images using lead pellet array • Use chest portion of RANDO phantom

  9. Noise Addition Based on slightly modified work by Bath et al, 2005 lung spine diaphragm Corrected noise image Uniform noise image

  10. Lung Nodule Simulation • Added artificial nodules to the CT data prior to X-ray projection • Baed on work by Li et al 2009

  11. Human DRR v Human CR DRR CR

  12. DRRs: 50 kVp v 150 kVp

  13. CR DRR DRR (c) (a) (b) Scatter Rejection • The DRR algorithm can produce images with scatter rejection DRR No Rejection DRR Grid CR Grid

  14. DRR CR (a) (b) Obese Patients • DRR algorithm can also produce images of large/obese patients

  15. Validation • Decided to validate with RANDO and real patient images: • Histogram of pixel values • Signal to noise ratios (SNR) • Important because signal and noise affects the visualisation of pathology • Tissue to rib ratios (TRR) • Pixel value ratio of soft tissue to that of rib • Important as rib can distract the Radiologist from detecting pathology

  16. a c b d Phantom Histograms

  17. Phantom Histograms

  18. PATIENT - Histograms Typical histogram of average patient DRR Typical histogram of average patient CR image

  19. SNRs • Good agreement in lung, spine and diaphragm areas of chest • Maximum deviation approx 15% • Mean deviation = 7% • Addition of frequency dependent noise not perfect: • CR system noise is ergodic (changes with time) • Noise added here is a snapshot (and so not ergodic) • However, quantum noise dominates over ‘ergodic noise’ so not such an issue • DQE is NOT constant with dose variation in image

  20. Validation - TRRs • Good agreement • Within 2% • As tube potential increases TRR decreases • Due to rib attenuating higher percentage in incident photons at lower potentials than soft tissue, thus forcing up TRR

  21. Validation - Radiologists • Have told me DRR images contain sufficient clinical data to allow diagnosis and subsequent optimisation • They have scored the images out of 10 • ‘are the images sufficiently like real CR images? • Average score of 7.8

  22. Conclusions – DRR Algorithm • DRR computer program has been produced that adequately simulates chest radiographs of average and obese patients • Anatomical noise simulated by real human CT data • System noise and scatter successfully added post DRR generation that provides: • SNRs • TRRs • Histograms • in good agreement with those measured in real CR images • Provides us with a tool that can be used by Radiologists to grade image quality with images derived with different x-ray system parameters • WITHOUT THE NEED TO PERFORM REPEAT EXPOSURE ON PATIENTS • ACCEPTED FOR PUBLICATION IN THE BJR • AUGUST 2011

  23. Optimisation of CR Chest Radiography using DRR Generated Images • In Hull chest exposure factors were not standardised (historical reasons!!!!) • Three main hospital sites: • 60 kVp & 10 mAs • 70 kVp & 5 mAs • 80 kVp & 5 mAs • In the last 6 months, four expert image evaluators have scored DRR reconstructed images • Two Consultant Radiologists • Two Reporting Radiographers • Scoring criteria based on European guidelines (CEC)

  24. What to Optimise? • Optimum tube potential for Average Patients (70 kg ± 10 kg) • Without scatter rejection (as per Hull protocol) • With scatter rejection (grid and air gap) • Optimum tube potential for Obese patients • Without scatter rejection • With scatter rejection • Is Scatter rejection indicated? • Dose reduction?

  25. Scoring Criteria • Images scored on a dual PACS monitor system • Image on right hand screen held at a constant kVp • Images on left hand screen displayed from 50 to 150 kVp in steps of 10kVp (approx) • Image 1 = 50 kV • Image 2 = 60 kV • Image 10 = 150 kV • Test images scored against reference image • All images matched effective dose

  26. [1] European guidelines on quality criteria for diagnostic radiographic images. CEC European Commission EUR 16260 EN (Luxembourg 1996)

  27. Tube Potential Optimisation

  28. Results – Average Sized Patients, no scatter rejection VGAS = average of Radiologists results

  29. Results – Average Patients with scatter rejection - grid

  30. Results – Average Patients with scatter rejection – air gap

  31. Results – Obese Patients without scatter rejection

  32. Results – Obese Patients without scatter rejection • Very weak trend for better image quality with higher kVp • Lower kVps probably ‘worse’ due to combination of: • Lack of penetration through obese patient • Increased scatter from obese patient (scatter to cassette changes very little with kVp) • This is not so with average patients • Less tissue (fat) so more radiation penetration • Less scatter from fat • It is likely that poorer radiation penetration and increased scatter from obese patient outweighs the inherent benefit of photoelectric contrast obtained from lower kVps

  33. Results – Obese Patients with scatter rejection - grids

  34. Results – Obese Patients with scatter rejection – air gap

  35. Scatter Rejection

  36. Scatter Rejection vs no scatter rejection – average patients

  37. Scatter Rejection vs no scatter rejection – average patients • Superior image quality with scatter rejection technique • Grids performed much better than air gap • Statistically significant differences • BUT: • Image evaluators were asked if increase in dose due to use of grids was justified, even with better image quality • Answer in 100% of cases was NO • SCATTER REJECTION FOR AVERAGE PATIENTS IS NOT INDICATED

  38. Exposure time for average patients without scatter rejection and low tube potentials? • Scatter rejection not indicated • Therefore low kVps should be used (remember the graph) • European guidance recommends exposure times < 20 ms • Can we achieve this with low kVps?? • Modern Philips X-ray generator: • For lgM = 2 (Agfa CR specific Dose Indicator) • With 630 mA, all kVps are possible • Max exp time = 16 ms • At the expense of increased tube loading

  39. Scatter Rejection vs no scatter rejection – obese patients

  40. Scatter Rejection vs no scatter rejection – obese patients • Superior image quality with scatter rejection technique • Grids performed much better than air gap • Statistically significant differences • BUT: • Image evaluators were asked if increase in dose due to use of grids was justified, even with better image quality • Answer in 100% of cases was YES • ANTI SCATTER GRID USE FOR OBESE PATIENTS IS INDICATED

  41. Exposure times for obese patients with an anti-scatter grid • Remember that low tube potentials were superior with an anti-scatter grid!!!! • Need exposure times < 20ms • Is this possible with low kVps and scatter grid for lgM = 2?? • Scatter grid is a focused grid so have to use it in a fixed range of FDD • Nominal distance = 140 cm FDD • Range allowed = 115cm – 180cm • At 140 cm FDD lowest exp time = 17.6ms @ 109 kVp • At 115cm FDD lowest exp time = 20 ms @ 90 kVp • So are limited to 90 kVp

  42. Dose Reduction

  43. Dose Reduction? • Images were also presented at different doses

  44. Dose reduction? • Results suggest doses can be reduced by around 50% before image quality begins to suffer and become unacceptable • Therefore could half exposure mAs

  45. Conclusions – Use of DRR algorithm to optimise CR chest imaging • Average patients: • No scatter rejection is indicated • Therefore, low kVps (< 102 kVp) are indicated • Can have exposure times < 20 ms for all kVps • Obese patients • Anti-scatter grid is indicated • So low kVps (<102kVp) should be used • For exposure times < 20ms, are limited to 90 kVp • Doses • As low as 50% reduction possible • ACCEPTED FOR PUBLICATION IN BJR

  46. Optimisation & Standardisation in Hull? • Agreed with Consultant Radiologists: • 60 kVp & 10 mAs • After a ‘settling in’ period: • 60 kVp & 8 mAs • Want to go lower than 8mAs eventually!!! • Implication on patient dose? • Using PCXMC effective dose calculation software: • 80 kVp/5 mAs = 0.011 mSv • 60 kVp/8 mAs = 0.006 mSv • Approx 45% drop in effective dose

More Related