Validated data and removal of bias through Traceability to SI units

This article discusses the importance of validating data and removing bias in measurements by ensuring traceability to SI units. It explores key concepts such as traceability, SI units, uncertainty of measurement, accuracy, precision, and stability. The article also highlights the need for improved quality assurance in optical radiation measurement and the use of electrical substitution radiometry.

Download Presentation

Validated data and removal of bias through Traceability to SI units

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Validated data and removal of bias through Traceability to SI units Nigel Fox Centre for Optical and Analytical Measurement Dec 03

  2. Resolution adopted by CEOS Plenary 14 (Nov 2000) • 1/ All EO measurement systems should be verified traceable to SI units for all appropriate measurands. • 2/ Pre-launch calibration should be performed using equipment and techniques that can be demonstrably traceable to and consistent with the SI system of units, and traceability should be maintained throughout the lifetime of the mission. Traceability – Property of the result of a measurement or the value of a standard whereby it can be related to stated references, usually through an unbroken chain of comparisons all having stated uncertainties

  3. SI units – The coherent system of units adopted and recommended by the General Conference of Weights and Measures (CGPM). Reproducibility of results of measurements – Closeness of the agreement between the results of measurements of the same measurand carried out under changed conditions of measurement. Vocabulary for International Metrology (VIM) • Uncertainty of measurement – Parameter, associated with the result of a measurement, that characterises the dispersion of the values that could reasonably be attributed to the measurand. • Accuracy of measurement – Closeness of the agreement between the result of a measurement and a true value of the measurand. • Uncertainty of measurement – Parameter, associated with the result of a measurement, that characterises the dispersion of the values that could reasonably be attributed to the measurand. • Precision – No metrological definition except to state that it should never be used in the context of “accuracy” and, because of possible confusion its use, should normally be avoided in metrological applications. • Error of measurement – Result of a measurement minus a true value of the measurand • Stability – Ability of a measuring instrument to maintain constant its metrological characteristics with time. • Precision – No metrological definition except to state that it should never be used in the context of “accuracy” and, because of possible confusion its use, should normally be avoided in metrological applications. • Traceability – Property of the result of a measurement or the value of a standard whereby it can be related to stated references, usually through an unbroken chain of comparisons all having stated uncertainties. • Repeatability of results of measurements – closeness of the agreement between the results of successive measurements of the same measurand carried out under the same conditions of measurement.

  4. Convention of the Meter – established 1875

  5. SIM SI Traceability: The Mututal Recognition Arrangement (MRA) CCPR • CCPR Key comparisons • Spectral Irradiance • Spectral Responsivity • Luminous intensity • Luminous Flux • Spectral transmittance • Spectral diffuse reflectance • (total hemispherical)

  6. Monitoring and interpreting the Earth’ systems Incoming Solar Radiation Drives all the processes of the Earth System and potentially damaging (UV) to Biosphere (Human health) Solar Reflected Radiation Atmosphere - Aerosol (size & distribn) - Clouds - Pollution (impact on health) Water - Pollution (originator) - Algae plumes Land - Useage / condition - Type/quantity of vegetation - Minerals - Carbon & hydrological cycles Governments – Treaties, Tax, Planning Thermal Emitted Radiation Atmosphere –Atmospheric chemistry Water – Temperature Land – Fires, Volcanoes, Pollution, Spatial variability requires good stability and SNR (signal to noise ratio) from a single sensor - but long term studies “climate change” need accuracy and consistency Engineering specification of SNR should equate to accuracy

  7. LOS 1998 IEEE Trans G & RS Need for improved Quality Assurance Need for improved Quality Assurance Difficulties - bias between sensors - instruments change on launch and degrade in-orbit (gain and spectral) Requirement - baseline for climate studies - improve models - prediction of weather systems - identify crops from weeds - global warming - Man or Nature? - detection of change - monitoring the treaties - auditing carbon sinks - efficiency of carbon sinks - QA of operational services (GMES) - instrument synergy - application of correction for atmosphere loss - lack of cohesion between networks and “ground truth” validation data (atmosphere an exception) - models inadequate - no consistent statements of uncertainty or confidence. To strengthen the evidence

  8. Traceability chain for optical radiation measurement

  9. Electrical Substitution Radiometry a 100 yr old technology When thermometer temperature T=To=TE then Po=PE Optical power =Po Absorbing black coating Electrical Heater Power = PE Copper disk Thermal shroud When T =To=TE then Po=PE Optical power =Po Mechanical cryogenic cooler “Fridge” (T = 20 K) Shutter Cooling improves sensitivity by 1000 X Absorbing cavity (~ 0.99999) Electrical Heater Power = PE Principle of Cryogenic radiometry

  10. 25 yrs of cryogenic radiometry at NPL

  11. Primary standard lamp Satellite Pre-flight Calibration Spectroradiometer Traceability ?? Laser Cal interval ~100nm Working standard lamp Solar illuminated Diffuser Cal Lab Primary lamp Laser Cal interval ~0.1 nm Lamp Vicarious Cal Lab working std Lamp Radiance Temperature Atmosphere/ Model User Cal Lamp Radiance continuum (Planck) User Instrument Data products Spectroradiometer Spectroradiometer Spectroradiometer Fundamental constants (SI) Primary standard cryogenic radiometer Satellite In-flight Calibration Photodiode (spectral responsivity Filter Radiometer Ultra High Temperature Black Body (3500 K) Spectroradiometer (multi-band filter radiometer Spectral Radiance/Irradiance calibrations LAND OCEAN ATMOSPHERE

  12. Traceability for in-flight / in-situ / vicarious calibration Spectral Radiance - lamp illuminated spheres - Filter radiometers (spectroradiometers) - Irradiance source + diffuser Lamp + spectralon or …. Sun + spectralon or …. Sun + Moon Via models / atmosphere correction to satellite for cal/val (radiances) To bio/geophysical quantities (refelectances) Spectral Reflectance - in-situ absolute ratio (using radiometers) - Ratio to “standard” reflector/diffuser

  13. Validated data products require all processing steps and data to be QA – Accredited? • Pre-flight • User specification • Instrument build compliance • Calibration? • Post-launch • In-flight checks • Ground “Truth” comparison • Inter-sensor cross calibration • Processed data released • “validated” • Uncertainty statement? Rare for all these activities to have been independently reviewed and/or audited

  14. Global Monitoring Environment and Security (GMES)Joint initiative of ESA and EU Aim:to establish “operational services” for Earth Observation data to meet needs of key stakeholders , public services, private industry, academia and the citizen with a view to financial self-sufficiency. Success requires: - Combination of data from many sources, (satellites, in-situ, aircraft) - Efficient production of cost effective, reliable, data products / maps - Data must provide the evidence to allow decisions to be taken with confidence. - Innovation in measurement and analysis Reliability:Implies Quality assurance and statements of confidence associated with data (not only for end users but also “operational service” providers Users generally assume QA Robust evidence requires robust QA

  15. Infrastructure for innovation in measurement, validation and QA of EO data • Transfer standards • Comparisons • Innovation on techniques • Measurement & test protocols • International link • Independence Post-launch airborne NPL ++ Modelling & Data processing NIST ++ In-situ Calibration QA Traceability Audit Validation Academia Pre-flight Advice Private Industry Public sector

  16. Summary • Primary scales, transfer standards and techniques now allow high accuracy to be achieved for both pre-flight and vicarious calibration (particularly for radiance) • All aspects/steps of producing EO data products needs validation and traceability (instrument calibration and algorithms/models) • Consistent presentation and breakdown of uncertainty budgets • Flexibility to allow innovation • Regular comparisons to evaluate biases • Establish well characterised ground targets as “reference standards” • Develop improved “in-flight” calibration methods e.g TRUTHS (Fox et al Proc. SPIE 4881, p395 2003) For Earth Observation to provide the evidence to support policy requires the industry and its data to be as robust as traditional industries