Presentation during QIBA COPD Lung Density Reference and Consistency Standards Work Group Call. Philip F. Judy June 8, 2011. COPDGene Reference Standard Model 2. Spectral effects 200 mAs – B31f – 5mm – Varied tube potential. Observations.

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about 'Presentation during QIBA COPD Lung Density Reference and Consistency Standards Work Group Call' - upton


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
presentation during qiba copd lung density reference and consistency standards work group call

Presentation during QIBA COPD Lung Density Reference and Consistency Standards Work Group Call

Philip F. Judy

June 8, 2011

slide5
Spectral effects

200 mAs – B31f – 5mm – Varied tube potential

observations
Observations
  • CT numbers inside the phantom of foams identified by NIST are sensitive to spectral differences
  • CT number air inside the phantom is less sensitive to spectral differences than foams identified by NIST
  • CT number of acrylic inside phantom may useful characterizing the spectral effects on CT number inside the phantom
observations gammex stack
Observations:Gammex Stack
  • Substantial variation CT number of densest foam (-684) from sample to sample.
  • Variation caused by mAs and reconstruction kernel less than 1 HU
  • Limitation: Variation could be associated with spatial variation
  • Unexplained effect at lowest foam density
observations comparison scan of stack with scan of single phantom
Observations: Comparison scan of stack with scan of single phantom
  • Still is substantial variation CT number of densest foam (-684) from sample to sample.
  • Systematic effect observed.
  • Limitation: Effect could be associated with difference reconstruction field of view.