Ferroelectric-RTA Standard Operation Procedures

Ferroelectric-RTA Standard Operation Procedures. By David Wu. Check to see if the system is OK. Check the gas valves. Check to see if the leak valve in the mechanical pump is closed. If no, close it. Check to see if the system is on operation condition ( green label ).

Download Presentation

Ferroelectric-RTA Standard Operation Procedures

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Ferroelectric-RTA Standard Operation Procedures By David Wu

  2. Check to see if the system is OK • Check the gas valves. • Check to see if the leak valve in the mechanical pump is closed. If no, close it. Check to see if the system is on operation condition (green label). • Check the foregoing records in the working sheet.

  3. Turn on the system • User utility permit & personal password to turn on the power of monitor. • Turn on the power of computer. • Turn on the power of the RTA system. • Turn on the power of mechanical pump. • After that, the screen shows the window of main menu. • Main menu: • The main menu offers 6 options: • 1.Edit recipes: to create, modify, display or delete a recipe. • 2.Run Process: to run a process. • 3.Maintenance: for manual control of the system. • 4.Historicals: to display saved processed and alarm events. • 5.Parameters: to enter calibration tables, configuration parameters, etc. • 6.Exit: to return to “Windows” and to terminal control.

  4. Edit recipes • When the EDIT mode is selected, the screen displays the recipe programming screen. • •Each recipe is made of : • –A name (up to 10 characteristics), • –An associated calibration table to be chosen according to the sample to process, • –A comment, • –Thirty steps. • •

  5. For each step the operator can select: • –Temperature, • –Duration of step, • –Temperature control mode, • –To open or not the gas valve, • –To open or not the vacuum valve. • Caution • When using power mode there is no temperature control and you may damage your samples, or melt them in a few seconds if power is too high. • The temperature range is 400 to 900 ℃. The time range is 1 to 5 min. However, when the temperature is higher than 700 ℃, the duration must be kept to lower than 3 min. The duration time at 900 ℃ is limited to lower then 1 min

  6. Loading wafer • Select maintenance in the main menu. • Open purge valve. • When the chamber pressure reach 1 atm, open the chamber door and load process wafer. • Open vacuum valve to pump down chamber pressure until the pressure below 30 mbar. • Caution: • –Only ferroelectric materials, such as BST, PZT, SBT etc., are allowed.

  7. Run process • Select Run process in the main menu. • Download recipe. • Start process. • During process, the screen displays all the parameters: • Recipe name, current step, calibration table name, setpoint, temperature from pyrometer and thermalcouple, power to the lamps, and valve status. • Whenever emergency occurs during operation, select End of process to shut down the process.

  8. End of process and unloading wafer • After the process is completed, a window about cooling information appear automatically. • After the window disappear, open the chamber door and unload the wafer. • Clean the chamber by alcohol. • Close the chamber door and pump down the chamber until the pressure below 30 mbar.

  9. Turn off the system • Close all the valves including gas valve and vacuum valve. • Select exit in the main menu. • Turn off the computer. • Turn off the system. • Turn off the mechanical pump. • User utility permit & personal password to turn off the monitor power.