MATHEMATICS AND CARDIOLOGY: PARTNERS FOR THE FUTURE. Suncica Canic Department of Mathematics University of Houston. RESERVOIR. Compliance Chamber 1. LVAD. Compliance Chamber 2. PRESSURE TRANSDUCERS. LVAD DRIVING CONSOLE. INLET VALVE. CATHETER. OUTLET VALVE.

stasia
Download Presentation

MATHEMATICS AND CARDIOLOGY: PARTNERS FOR THE FUTURE

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. MATHEMATICS AND CARDIOLOGY: PARTNERS FOR THE FUTURE Suncica Canic Department of Mathematics University of Houston

  2. RESERVOIR Compliance Chamber 1 LVAD Compliance Chamber 2 PRESSURE TRANSDUCERS LVAD DRIVING CONSOLE INLET VALVE CATHETER OUTLET VALVE

  3. RESERVOIR Compliance Chamber 1 LVAD Compliance Chamber 2 PRESSURE TRANSDUCERS LVAD DRIVING CONSOLE INLET VALVE CATHETER OUTLET VALVE

  4. RESERVOIR Compliance Chamber 1 LVAD Compliance Chamber 2 PRESSURE TRANSDUCERS LVAD DRIVING CONSOLE INLET VALVE CATHETER OUTLET VALVE

  5. PROBLEM FLUID-STRUCTURE INTERACTION BETWEEN BLOOD FLOW AND ARTERIAL WALLS IN HEALTHY AND DISEASED STATES ANALYSIS OF FLUID-STRUCTURE INTERACTION CAN: 1.Help predict initiation of disease 2.Helpimprove treatment of disease Prostheses design for non-surgical treatment of AAA and CAD • aortic abdominal aneurysm (AAA) repair • coronary artery disease (CAD) repair. 33

  6. DIFFICULT PROBLEM TO STUDY: MULTI-PHYSICS AND MULTI-SCALE NATURE Plasma • BLOOD has complicated rheology: red blood cells, white blood • cells and platelets in plasma (relevant at small scales) • VESSEL WALLS have complex structure: intima, media, adventitia • (+ smaller scales layers); different mech. char. • Challenging to model. • INTERACTION (COUPLING) exceedingly complicated. Red Blood Cells Platelets White Blood Cells

  7. COUPLING BETWEEN BLOOD FLOW AND VESSEL WALL MOTION • NONLINEAR COUPLING: density of the arterial walls is roughly the same as • density of blood • TWO TIME SCALES: fast traveling waves in arterial walls and • slow bulk blood flow velocity • COMPETITION BETWEEN “HYPERBOLIC” AND “PARABOLIC” EFFECTS • (wave propagation vs. diffusion) • algorithms developed for other applications, e.g. aeroelasticity, UNSTABLE; • novel ideas and algorithms needed • resolving both scales accurately requires sophisticated methods • resolving the two different effects requires different techniques

  8. TRADITIONAL SOFTWARE FOR BLOOD FLOW SIMULATION • ASSUMES FIXED VESSEL WALLS • PAST 10 YEARS: intensified activity in fluid-structure interaction studies due to development of new mathematical tools (beginning with earlier work of Peskin (1989).) • CURRENT METHODS (far from optimal): • computationally expensive (implicit, monolithic schemes, commercial software) • OR • suffer from stability problems (explicit, loosely coupled algorithms) CHALLENGES: -3D simulations of larger sections of cardiovasc. sys. - complicated geometries - complicated tissue models ACTIVE AREA OF RESEARCH in the years to come

  9. COMPREHENSIVE STUDY OF FLUID-STRUCTURE INTERACTION IN BLOOD FLOW(medium-to-large arteries: laminar flow and Re away from the turbulent regime) ANALYSIS • Fundamental properties of the interaction and of the solution. • Derivation of new closed, effective models. COMPUTATION • Design of a numerical algorithm (“kinematically coupled”) with a novel operator splitting approach (hyperbolic/parabolic) with improved stability properties. • Models allowing two different structures (stent modeling). • Fluid-cell-structure interaction algorithm VALIDATION AND TREATMENT TEXAS MEDICAL CENTER HOUSTON • Experimental validation. • Application to AAA repair and coronary angioplasty with stenting. 33

  10. Treats more than 5.5 million patient visits annually; 73,600 employees 37 million sq feet of space 46 institutions (hospitals, educational, service) http://www.texmedctr.tmc.edu/root/en/GetToKnow/FactsandFigures/FactsAndFigures.htm * * * * *

  11. THE TEXAS HEART INSTITUTE Dr. Denton Cooley: Founder of THI Pioneer of Heart Transplants HOUSTON (July 13, 2007)U.S. News & World Report ranked the Texas Heart Institute among the nation's top ten heart centers for the 17th consecutive year. Michael Ellis DeBakey born SEPTEMBER 7, 1908. • pioneer in the field of cardiovascular surgery • pioneer in surgical treatment of AAA • 2006 (age 97): Dr. DeBakey treated for AAA; his procedure • oldest patient ever to undergo this treatment • hospital recovery lasted 8 months Dr. DeBakey Dr. Cooley

  12. PROJECTS • Aortic Abdominal Aneurysm (AAA) • Optimal stent design for non-surgical treatment of AAA • Compliancy • Geometry • Graft Permeability • Experiments • Coronary Artery Disease (CAD) • Tissue engineered stents for coronary angioplasty • Auricular chondrocytes lining of artificial surfaces • Stent Design for CAD and heart valve replacement

  13. What is abdominal aneurysm? Aneurysm: dilatation of an artery Mortality: 90% for out-of-hospital rupture (Experimental) Nonsurgical Procedure: - Developed for high-risk patients - Performed using catheterization Complications: - Stent and stent graft migration (20.2%) - Change in shape (56%) - Formation of new aneurysms near anchoring - Graft limb thrombosis - Permeable grafts-> aneurysm growth

  14. STUDY OPTIMAL PROSTHESIS DESIGN FOR AAA REPAIR • METHODS • EXPERIMENTAL MEASUREMENTS OF PROSTHESES • MECHANICAL PROPERTIES (Ravi-Chandar, UT Austin) • MATHEMATICAL MODELING OF PROSTHESES MECHANICS AND • DYNAMICS • COMPUTER SIMULATIONS • EXPERIMENTAL VALIDATION

  15. RESULTS LEAD TO NEW STENT-GRAFT DESIGN MODELING AND COMPUTATION PRODUCED: • RESULTS FOR FLEXIBLE bare Wallstent. • Wallstent 10 times more elastic than aorta: large radial displacements ANGIO • large stresses and strains near anchoring (possibility of migration)PLAY MOVIE • POOR PERFORMANCE NO LONGER USED • RESULTS FOR FABRIC-COVERED STENT-GRAFTS • graft is stiff; elastic exoskeleton tends to pulsate: possibility for suture breakage • stiff graft: elevated local transmural pressure COMPARISONMOVIE • NON-UNIFORM STIFNESS MINIMIZES STRESS AT ANCHORING Next slide [1] Canic, Krajcer, Lapin, Endovascular Today (2006) [2] Canic, Krajcer, Chandar, Mirkovic, Lapin, Texas Heart Institute Journal (2005) [3] R. Wang and K. Ravi-Chandar, Mechanical response of an aortic stent I and IIJournal of Appl. Mechanics, (2004.) [4] SIAM News, Vol. 37 No. 4 (2004) Dana McKennzie

  16. AAA Walstent (compliant)

  17. Suggested Opt. Design in[1] (NEW) Variable stiffness Limbs diameter around 0.7 of main body diameter Larger main body diameter Longer main body Low shear stress rates in the limbsmovie AneuRx Stent-Graft (OLD) Uniformly stiff Limbs diameter less than 0.5 of main body High shear stress rates in the limbs Small limb diameter implies high SSR refs • RESULTS FOR BIFURCATED STENT-GRAFT DESIGN [1] Canic,Krajcer,Lapin, Endovascular Today (Cover Story) May 2006. Show paper

  18. INFLUENCING STENT-GRAFT INDUSTRY NEW Endologix Stent-Graft (2007) AneuRx Stent-Graft Our results show this geometry will have lower limb thrombosis rates.

  19. MATHEMATICAL MODELING AND COMPUTATION SUGGEST IMPROVED DEVICE DESIGN DETECT DEVICE’S STRUCTURAL DEFICIENCIES

  20. MEDICAL PROBLEM • Coronary stenosis • constriction or narrowing of coronary arteries • coronary arteries supply oxygenated blood to the heart. • 12,600,000 Americans suffer from CAD • 515,000 die from heart attacks • caused by CAD each year (NHLBI) • Treatment: • coronary angioplasty

  21. COMPLICATION: RESTENOSIS • related to the development of neo-intimal hyperplasia • scar tissue in response to mechanical intervention with material of poor biocompatibility • 35% after angioplasty without stent • 19 % with stent (R. Kurnik) movie

  22. Biocompatibility/hemocompatibility • (Dr. Doreen Rosenstrauch) • endothelial cells • optimal lining but not easily accessible, harvested or isolated • genetically engineered smooth muscle cells (similar) • auricular chondrocytes (ear cartilage) (with Dr. Rosenstrauch) • - genetically engineered to produce NO • - easily accessible: minimally invasive harvesting • - superior adhearance (collagen) • - good results with LVADs (Dr. Rosenstrauch, Scott-Burden et al.)

  23. STENT COATING Day 2 100x 200x 400x Day 3 100x 200x 400x • Cardiovascular Surgery Research Lab– Texas Heart Institute • Marie Ng • Boniface Magesa • Doreen Rosenstrauch • Arash Tadbiri

  24. RESULTS: • optimize initial seeding for fast complete coverage of stent • study initial cell loss under flow conditions (pre-conditioning) (cell rolling • and detachment) Show results USE MATHEMATICS AND COMPUTATION TO REDUCE THE EXTENT OF EXPERIMENTAL INVESTIGATION TO OPTIMIZE THE PRODUCTION OF CELL-COATED STENTS

  25. PRE-CONDITIONING PHASE: CELL ROLLING AND DETACHMENT Fluid velocity=const. t = 0 Period boundary conditions Fluid velocity=0 No-slip boundary condition Fluid velocity=const. t > 0 Fluid velocity=0 MATHEMATICAL AND COMPUTATIONAL ALGORITHM FLUID-PARTICLE INTERACTION ALGORITHM Glowinski,Pan et al., J. Comp. Phys. (2001) DYNAMIC ADHESION ALGORITHM Hammer and Apte, Biophys.J. (1992)

  26. Number of cells = 80 Mesh size h for the velocity=0.1 mm (using P1 element) Cell size (ellipsas)= 2 x 1.6 mm Mesh size h for the pressure=0.2 mm (using P1 elements) Channel length=400 mm Each cell occupies 20x16 mesh blocks. Dual core AMD Opteron 275 @ 2.2 GHz : 11h 30min 4 sec (not parallelized) Adhesion Algorithm coupled with Fluid-Particle Interaction Algorithm • RESULTS • Cell detachment in the pre-conditioning stage (stochastic bond dynamics) Shear rate (1/s) Viscosity(g/cm s) Detachment % • observed chondrocyte sliding in simulations • (experimentally verified!!) • captured cell detachment (initial linear growth • experimentally verified) 0.01 100 0 0.01 200 25 0.05 5 0 0.05 8 10 0.05 9 30 Click inside the picture to run the movie: (blood:0.03 ; 100 in dog’s coronaries) • NEXT • Optimize pre-conditioning by varying shear rate and fluid viscosity

  27. NEXT: • study behavior of cell-coated stent inserted in a compliant vessel • (latex tube; in vitro testing) complex hemodynamics conditions: • MODELING: Fluid-Cell-Structure Interaction Algorithm

  28. MATHEMATICAL PROBLEM FLUID-CELL-STRUCTURE INTERACTION CELLS Auricular chondrocytes Cell adhesion and detachment Hammer’s adhesion dynamics algorithm

  29. MATHEMATICAL FLUID-STRUCTURE INTERACTION IN BLOOD FLOW

  30. MODELS FLUID-STRUCTURE INTERACTION BETWEEN COMPLIANT WALLS Linearly elastic membrane Linearly elastic Koiter shell Linearly viscoelastic membrane Linearly viscoelastic Koiter shell Nonlinearly elastic membrane Large arteries Medium arteries 3D Linearly Elastic Pre-Stressed aaaaaaTHICK-WALL Tube Incompressible Stokes eqns. Incompressible Navier-Stokes

  31. (membrane) (viscoelastic membrane) LARGE ARTERIES & MODERATE Re Fluid: S(t) h Structure: (Long. displ. neglig) R0 W(t) z Coupling: Through the kinematic and dynamic lateral boundary conditions : (1) Continuity of the velocity (2) Balance of contact forces: Fstructure = -Ffuid

  32. BENCHMARK PROBLEM IN BLOOD FLOW • The fluid equations (incompressible, viscous, Navier-Stokes) • on the domain with a moving boundary • The structure equations (viscoelastic membrane/shell) • The lateral boundary conditions (coupling) • The inlet and outlet boundary conditions: • The initial conditions:

  33. REVIEW • Bioengineering/math hemodynamics literature (numerical methods): • Groups: EPFL & Milano (Quarteroni et al.), NYU (Peskin and McQueen), Stanford (Taylor&Hughes(UT)), U of Pitt (Robertson), New Zeland (Hunter, Pullan), Eindhoven (de Haar), UC-San Diego (YC Fung), Graz University of Technology (Holzapfel), Cambridge (TJ Pedley), University of Trieste (Pedrizzetti), Technical University Graz (Perktold, Rappitsch), WPI(Tang) • METHODS: Immersed Boundary, ALE, Fictitious Domain, Lattice Boltzman, Coupled Momentum Method, … Commercial Software (ANSYS,ADINA,…) • Many issues remain open • Mathematical fluid-structure interaction (existence/stability proofs): • H.B. daVeiga; Esteban, Chambolle, Desjardins, Grandmont; LeTallec; M. Padula,V. Solonnikov. • Existence for 3D benchmark problem with physiological data remains open • S.Canic, T. Kim, G. Guidoboni: existence for an effective model (2007)

  34. DERIVATION OF A CLOSED, REDUCED, EFFECTIVE MODEL WHEN e=R/L << 1 ANALYSIS NUMERICAL SIMULATION WEAK FORMULATION ENERGY ESTIMATE A PRIORI SOLUTION ESTIMATES EXPERIMENTAL VALIDATION ASYMPTOTIC EXPANSIONS HOMOGENIZATION Nonlinear EXISTENCE REDUCED (EFFECTIVE) EQUATIONS Moving boundary New information CONVERGENCE || SOLUTION – solutione|| ? ERROR ESTIMATES RESULT: small (coronary) arteries (Stokes equations, linear coupling) SIAM Appl. Dyn Sys. 2003.

  35. 10% of R ENERGY EQUALITY where A PRIORI ESTIMATES (when inertial forces dominate viscous forces)

  36. Show movie Transport of R+h with average fluid velocity Fluid diffusion is dominant in the r-direction THE REDUCED EQUATIONS 0th-order approximation: NOTE: nonlinearity due to the fluid-structure coupling dominates the nonlinearity of fluid advection. INITIAL and BOUNDARY DATA Dominanat smoothing well-posedness novel numerical algorithm for benchmark problem COMING SOON: Kinematically coupled scheme

  37. linear elasticity viscoelasticity COMPARISON WITH EXPERIMENTS (human femoral artery) Measured pressure-diameter response (Armetano et al.*): Numerical simulation using the reduced (Biot) model nonlinear elasticity in-vitro measurement LINEARLY VISCOELASTIC CYLINDRICAL MEMBRANE [1] SIAM J Multiscale Modeling and Simulation 3(3) 2005. [2] Annals of Biomedical Engineering Vol. 34, 2006. [3] SIAM J Applied Mathematics 67(1) 2006. coming soon: user-friendly software posted onwww.math.uh.edu/~canic(Tambaca&Kosor) *[1] Armentano R.L., J.G. Barra, J. Levenson, A. Simon, R.H. Pichel. Arterial wall mechanics in conscious dogs: assessment of viscous,inertial,and elastic moduli to characterize aortic wall behavior. Circ. Res. 76: 1995. * [2] Armentano R.L., J.L. Megnien, A. Simon, F. Bellenfant, J.G. Barra, J. Levenson. Effects of hypertension on viscoelasticity of carotid and femoral arteries in humans. Hypertension 26:48--54, 1995.

  38. VELOCITY MEASUREMENTS ANDCOMPARISON WITH NUMERICAL SIMULATIONS (S. Canic, Dr. C. Hartley, Dr. D. Rosenstrauch, J. Chavez, H. Khalil, B. Stanley ) Research Laboratory at THI • mock flow loop with compliant walls and pulsatile flow pump • pulsatile flow pump: HeartMate LVAD • compliant tubbing: custom made latex (Kent Elastomer Inc.) • ultrasonic imaging and Doppler methods: measure velocity & displacement • high frequency (20 MHz) crystal probe was used • non-dairy coffee creamer dispersed in water to enable reflection

  39. RESERVOIR Compliance Chamber 1 LVAD DRIVING CONSOLE LVAD PRESSURE TRANSDUCERS Compliance Chamber 2 INLET VALVE CATHETER OUTLET VALVE

  40. CONCLUSIONS • made progress in understanding fluid-structure • interaction in blood flow; in the design of numerical • methods to capture the interaction, and in the design of • stents and stent-grafts for CAD and AAA treatment • sophisticated mathematics can help improve • design of vascular devices, and give an insight • into the hemodynamics of cardiovascular interventions • problems arising in cardiovascular interventions • can drive the development of sophisticated mathematics

  41. COLLABORATORS Cardiologists: Mathematicians: Dr. Z. Krajcer, M.D., THI Dr. D. Rosenstrauch, M.D. THI Dr. A. Mikelic, U of Lyon 1, FR Dr. J. Tambaca, U of Zagreb, HR Dr. G. Guidoboni, U of H, U of Ferrara, IT Engineering/Measurements Math/Sci. Computing: Dr. K. Ravi-Chandar, UT Austin Dr. R. Glowinski, U of Houston Dr. T.-W. Pan , U of Houston Dr. D. Mirkovic, MD Anderson Cancer Center Dr. C. Hartley, Baylor College of Medicine Students: J. Hao,S. Lapin, T.B. Kim, B. Stanley, M. Kosor, T. Josef (Rice), J. Gill (Rice), K. Mosavardi (UT Health Sci. Houston), J. Chavez, H. Khalil, K. Vo, R. Patel, C. Chmielewski (UH&NCState), H. Melder, A. Young (Penn State), D. Roy,Y. Barlas, K. Buss (UH), E. Delavaud & J. Coulon (U of Lyon1), D. Lamponi (EPFL)

  42. University of Houston, MD Anderson Library • THANKS: • The National Science Foundation • The National Institutes of Health (joint with NSF: NIGMS) • Roderick Duncan MacDonald Research Grant at St. Luke’s Episcopal Hospital, Houston • Texas Higher Education Board (ATP Mathematics) • Kent Elastomer Products Inc. • UH Mathematics Department Summer Research Grant • Medtronic Inc. Final note: For the first, the FDA might require the use of math modeling and numerical simulations for peripheral prostheses design FDA