slide1 n.
Download
Skip this Video
Loading SlideShow in 5 Seconds..
M. D. A NDERSON Cancer Center PowerPoint Presentation
Download Presentation
M. D. A NDERSON Cancer Center

Loading in 2 Seconds...

play fullscreen
1 / 1

M. D. A NDERSON Cancer Center - PowerPoint PPT Presentation


  • 143 Views
  • Uploaded on

A. A. C. B. B. C. Gamma Imaging with DigiRad: 111 In-DTPA-PEG-Annexin V. Gamma Imaging with m.CAM: Example Image. Gamma Imaging with m.CAM: 111 In-DTPA-Annexin V vs. 111 In-DTPA-PEG-Annexin V.

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about 'M. D. A NDERSON Cancer Center' - spiro


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
slide1

A

A

C

B

B

C

Gamma Imaging with DigiRad:

111In-DTPA-PEG-Annexin V

Gamma Imaging with m.CAM: Example Image

Gamma Imaging with m.CAM:

111In-DTPA-Annexin V vs. 111In-DTPA-PEG-Annexin V

Mice with MDA-MB-468 tumors without treatment (A), treated with PG-TXL 4 d earlier (B), or C225 4 d earlier (C). Mice were then injected with 111In-DTPA-PEG-Annexin V. Images were acquired 48 h after radiotracer injection. Arrow: Tumor.

Mice with MDA-MB-468 tumors without treatment (A), treated with PG-TXL 4 d earlier (B), or PG-TXL + C225 1 d earlier (C). Mice were then injected with 111In-DTPA-Annexin V (A&B) or with 111In-DTPA-PEG-Annexin V (C) .

First 2-D image acquired with the m.CAM. Twelve nude mice containing an 111In radiolabeled compound were imaged simultaneously at 48 hr post-injection. Image acquisition parameters were: 5 min static, 1024x1024 matrix, zoom = 1 (pixel size = 0.6 mm), and three simultaneous energy window settings (top to bottom: 172 keV/15%, 247 keV/15% and combined).

M. D. ANDERSON Cancer Center

Nuclear Imaging (M.Cam)

Richard E. Wendt III, Ph.D. and William D. Erwin, M.S.

Small Animal Cancer Imaging Research Facility

Overview of Modality

The Siemens Medical Solutions, USA m.CAM™ gamma camera is designed for imaging pharmaceuticals labeled with gamma emitting radionuclides, in a range of laboratory animals, from mice to small primates or dogs. It is an ideal instrument for biomedical research laboratories to quantitatively measure the biodistribution of radiolabeled diagnostic and therapeutic compounds in vivo, in either 2-D or 3-D (SPECT) mode.

Instrument Description

The Siemens m.CAM consists of a single High Definition Digital Detector (HD3) mounted onto a portable stand. The HD3 head may be rotated over 90 degrees, for imaging at any angle between horizontal (face up) to vertical (face out). The HD3 technology provides improved spatial and energy resolution, uniformity and count rate performance, in an energy independent fashion, compared to conventional detector head designs. This is especially important when imaging small objects such as mice.

Performance and Capabilities

The maximum imaging field-of-view (FOV) is 53.3 cm (21”) x 38.7 cm (15.25”) with an intrinsic resolution of 3.2 mm. The m.CAM is equipped with a Low Energy High Resolution (LEHR) collimator for imaging radionuclides such as Tc-99m (140 keV), and a Medium Energy Low Penetration (MELP) collimator for radionuclides such as In-111 (172, 247 keV). The system resolution is 4-20 mm (6.3 mm at 10 cm, LEHR with Tc-99m) depending upon gamma ray energy, subject size and collimator. Achievable image pixel sizes range from 19.18 mm (32x32 image matrix, zoom factor = 1) down to 0.187 mm (1024x1024 image matrix, zoom factor = 3.2). Inherent system deadtime is 5.74 sec, resulting in a count rate data loss of 20% at 38,400 counts per second (cps).

System sensitivity (Tc-99m, 15% energy window) is 172 cpm/Ci for the LEHR collimator, and 237 cpm/Ci for the MELP collimator. Up to six independent energy windows (50 keV – 600 keV) can be selected.

The system is capable of acquiring conventional 2-D (planar) static images and dynamic images (for cardiac imaging or perfusion imaging). A Tomographic Accessory for 3-D SPECT imaging of mice and rats is currently under evaluation. Data output formats are Siemens ICON, Interfile 3.3, and DICOM 3.0. Images can be converted to several standard formats as well (TIFF, JPEG, etc.) Data can be provided to investigators via ftp or on Zip (100 Mb), Jaz (1Gbyte), CD-R, and CD-RW physical media.

Potential Applications

Single animal imaging: mouse through small primate or dog

Simultaneous imaging of several animals: Up to 12 mice have been imaged at once, yielding higher throughput, consistent deadtime characteristics across a cohort, and synchronized dynamic data across a cohort