L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse Villa Cagnola, Gazzada (VA) - 13 Giugno 2009. IGRT VOLUMETRICA. marta.scorsetti@humanitas.it. Introduction. IGRT: tools for verifying patient position and adapting treatment plans . 

shubha
Download Presentation

marta.scorsetti@humanitas.it

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse Villa Cagnola, Gazzada (VA) - 13 Giugno 2009 IGRT VOLUMETRICA marta.scorsetti@humanitas.it

  2. Introduction IGRT: tools for verifying patient position and adapting treatment plans .  Increases accuracy and precisionbut unfortunately it also increasestreatment time(because it requires additional time for image acquisition evaluation, patient repositioning and adaptation) . L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  3. Introduction IMRT also increases treatment delivery time by requiring a large number of beam directions and increases monitor units (MU)  Time is important ! Is there a procedure that is better and more time efficient ? L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  4. VMAT • VMAT (volumetric modulated arc therapy) is a treatment planning and delivery platform using a single 360 deg gantry Arc : • time efficient • higly conformal dose distribution • superior dosimetric accuracy L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  5. RapidArc RA is a planning and delivery technique which aims to : • improve the degree of target coverage • improve OARs and healthy tissue sparing compared to other IMRT solutions • reduce significantly the treatment time (beam on time) per fraction L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  6. RapidArc • Total MU • Gantry position • X-ray fluence • Gantry rotation speed • MLC shape • Shorter Treatment Times • Improves patient comfort • Limits unwanted patient movement • Decreases target motion • Better Dose Conformity and Tissue Sparing • Fewer MUs • Less chance of secondary cancers L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  7. RT outcome RT global outcome Multi-immaging for planning: CT, PET, MRI, 4D-PET/CT Mobile laser IMRT Portal imaging 4D-RT Micro-MLC NEW: Volumetric arc therapy L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  8. WorkFlow with RapidArc Target and OAR delineation I II QA III IV Multi-modality imaging (CT,PET,MRI,4D-PET/CT) Treatment Inverse Treatment Planning L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  9. RapidArc: an example Need to control patient re-positioning 70Gy 35Gy • PTV1:70 Gy, PTV2: 54 Gy • 410 MU (2 Gy/day) • 135 s (2 full arcs) L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  10. Repositioning Not personalised systems Personalised systems L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  11. IGRT On Board Imaging - OBI X-Ray tube kV flat pannel L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  12. IGRT - OBI 3D CBCT-simulation CT matching 2D kV-kV image matching L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  13. IGRT - CBCT L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  14. Integrated RapidArc&IGRT 65 s RapidArc: High dose gradient Fast (65 s, 1 arc) Prostate motion Courtesy of Lynn Purdy, Calypso Medical Technlogies Low Intra-motion Only Inter-motion is required OBI L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  15. Pathologies@Humanitas Prostate Abdominal lesion H&N Mesothelioma Lung vertebra L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  16. Pathologies@Humanitas vertebra L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  17. Pathologies@Humanitas DTot=66 Gy (2Gy/d) vertebra Objectives: Dose elevata ed omogeneità al target Limitare dose al polm controlat L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  18. RapidArc@Humanitas Installation 6 October – 10 October Acceptance tests 13 October – 31 Ottobre Training 3 November – 15 November 1st patient 21 November Total patients 56 @10/6/09 Patients per month Patients per anatomical region L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  19. RapidArc@Humanitas 14 cases solitary abdominal mets 6 x 7.5 Gy, single arc RA vs IMRT vs Conf arcs Bignardi et al. IJROBP 2009 L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  20. RapidArc@Humanitas RA 27 x 2 Gy = 54 Gy 15/18 MV 9 fields IMRT 2 isocentric arcs L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  21. H&N: repositioning Planning L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  22. H&N: repositioning SimulationCT CB_CT: day6 CB_CT: day12 L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  23. Prostate: repositioning Simulation TC 70 Gy 95% PTV2 95% PTV1 20 Gy L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  24. Prostate: repositioning CB_CT: day1 70 Gy 95% PTV2 95% PTV1 20 Gy L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  25. Prostate: repositioning CB_CT: day4 70 Gy 95% PTV2 95% PTV1 20 Gy L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  26. Prostate: repositioning CB_CT: day14 95% PTV2 70 Gy 95% PTV1 20 Gy L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  27. PET - CBCT • CBCT • PET (plan_CT-PET) Courtesy: L.Cozzi - Bellinzona L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  28. Advantages of RapidArc • Fast treatment ( < 2 minute treatments) • Enhances patient comfort • Enhances safety and quality • Allows sufficient time for proper IGRT • Reduces errors due to internal organ motion and repositioning • Reduce risk of trade-off against cell repair in toumours • Allows definition of hypofractionation schemes in selected indications • Allows spread of complex SIB techniques with dose escalation and acceleration schemed • Allows reintroduction of iperfractionation schemes (more fractions per day at low dose per fraction) CB_CT: day14 L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  29. Conclusions • RA and volumetric IMAT proved to be a clinical step forward in terms of conformal avoidance • Clinical productivity : waiting lists, throughput, costs • Patient's eye view :  less intra-fraction movement, less internal organ deformation, better compliance • Doctor's eye view : more selective and precise treatment in a shorter time, increases the possibility to use non conventional fractionation (hypofractionation, iperfractionation, SIB) • Tumour's eye view : cell dynamics ( fast response to local damage), volumetric effects • Robust QA process in place • Higly active research programmes running at main institutions CB_CT: day14 L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009

  30. L’offerta tecnologica in Radioterapia: Equilibrio tra appropriatezza clinica e utilizzo delle risorse - 13 Giugno 2009 marta.scorsetti@humanitas.it Grazie