1 / 19

R&D status of 13inch HAPD

In this talk, we present the R&D status of the 13-inch HAPD, a useful photon detector for the 2km detector in T2K. The HAPD offers excellent performance compared to PMTs, with single photon sensitivity and a time resolution of 190ps. We also discuss the development of a dedicated readout system for the HAPD and its potential applications in other fields.

ronaldalee
Download Presentation

R&D status of 13inch HAPD

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. R&D status of 13inch HAPD T. Abe, H. Aihara, M. Iwasaki, T. Uchida (Univ. of Tokyo) M. Tanaka, (KEK) Y. Kawai, H. Kyushima, M. Suyama (HPK) M. Shiozawa (ICRR)

  2. Introduction • We believe HAPD is an useful photon detector for 2km detector in T2K. • We have developed large (13-inch) HAPD. • In this talk, we present the R&D status of the HAPD.

  3. HAPD HAPD uses avalanche diode as e multiplier. # of parts < 1/10 of PMT

  4. 190ps σ Theme • Excellent HAPD performance than PMT’s • Single photon sensitivity • Single photon time resolution ~ 190ps (s) • Readout system development dedicated for HAPD • Possible application to other field

  5. Bombardment: x4500@20kV Ne/sNe=67 Avalanche Gain: X30 Total Gain:~105 Principle of HAPD Operation HAPD PMT 1st Dynode Gain: x5 Ne/sNe=2.2 Incident Light Incident Light Photoelectron Photoelectron Large gain @ 1st mult. stage  good single photon detection 10~20kV Dynodes Avalanche Diode No Dynodes (No TTS)  large HAPD with good time resolution Total Gain:~107 Total Gain:~105

  6. Property of HAPD • Single photon sensitivity large gain at the first stage of multiplication • Good time resolution No TTS in dynodes • Cost reduction and easy Quality control simple structure without dynodes • Lower gain Need a dedicated readout system

  7. t f HAPD system (Q, T) QP.H. TTime ≥1Gsps MOF 13-inch HAPD Pre-amplifier ASIC (Low noise) Waveform sampling (Fast sampling) +Digital filter (Noise Suppression) HAPD performances are evaluated using this system

  8. 13inch HAPD performances • Energy measurement related items • Comparison between HAPD and PMT • Dynamic range • Position dependence of relative gain • Timing measurement related items • Comparison & # of P.E. dependence • HV dependence

  9. HAPD and PMT 1P.E. 2P.E. 3P.E. 4P.E. 1P.E. 5P.E. 2P.E.(?) HPD(HV=12kV bias=330V) PMT(13inch)

  10. :Sample A :Sample B :Sample C Dynamic range (no preamplifier) 3000p.e. equivalent input 110 HV=+20kV, Bias=290V for Sample A 370V for Sample B 390V for Sample C Light Source: Pulsed Laser (PW: ~70ps, λ: ~400nm) 100 90 80 Output/input [%] 70 60 -3% Deviation at Output Charge of 100pC (~3000p.e. input signal intensity equivalent at total gain of ~2x105) 50 40 1 10 100 1000 Output Charge [pC]

  11. 18cm ~65deg. Position dependence of gain for photons<5 Gain uniformity:s~2%

  12. 45 deg 70 deg Time resolution (large HAPD) Time resolution = Transit time variation Timing resolution Center

  13. Timing resolution Spot Laser illuminated HV=12kV bias=330V Better st No dependence of # of p.e. 190ps

  14. Photoelectron transit time Center Bias=290V Light Source: Pulsed Laser (PW: ~70ps, λ: ~400nm) 45 deg 70 deg Position dep. of Transit time < 1 nsec

  15. HAPD vs. PMT

  16. Under development f FPGA 1Gsps AMC t Pre-amplifier ASIC Analog memory cell (AMC) + FPGA for Digital Filter f HAPD electronics status (Q, T) ≥1Gsps MOF

  17. HAPD electronics status • Preamplifier (low noise) We achieve intrinsic ENC to be 3400 leading S/N~60 for the HAPD gain of 2X105. • Analog Memory Cell (fast sampling) AMC + slow FADC is cheaper than fast FADC. Sampling speed of 1GHz and signal depth of 512ns are obtained. • FPGA We start to develop a dedicated FPGA board for DSP.

  18. Application for 2km detector • The 13inch HAPD might be very big for 2km detector. • We will develop HAPD to apply for 2km detector in the following way • Small HAPD (8inch) will be fabricated in the next fiscal year. • Multi-pixel HAPD (13inch) can obtain fine segmentation without increasing # of HAPD (low cost).

  19. Summary • We develop a 13inch HAPD and confirm • Excellent single photon st~190ps. • Excellent single photon sE~44%. • Total gain of > 2X105. • We develop a dedicated readout electronics for the HAPD.