1 / 25

SBRT IN THE EVERYDAY CLINICAL SETTING

SBRT IN THE EVERYDAY CLINICAL SETTING. William H. Hinson, Ph.D. Department of Radiation Oncology Wake Forest University Baptist Medical Center Winston-Salem, N.C. SBRT Team at WFUBMC. Physician Volker Stieber William Blackstock Kevin McMullen Physics William H. Hinson William T. Kearns

roch
Download Presentation

SBRT IN THE EVERYDAY CLINICAL SETTING

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. SBRT IN THE EVERYDAY CLINICAL SETTING William H. Hinson, Ph.D. Department of Radiation Oncology Wake Forest University Baptist Medical Center Winston-Salem, N.C.

  2. SBRT Team at WFUBMC Physician • Volker Stieber • William Blackstock • Kevin McMullen Physics • William H. Hinson • William T. Kearns Whole cast of support staff

  3. Essentials for SBRT • 3-D Imaging device • Treatment planning system • Treatment delivery system • Precise targeting (stereotactic coordinate system) • Patient positioning verification

  4. Specialized Devices for SBRT • Novalis • Cyberknife • Accelerator-based IGRT (Trilogy, Synergy)

  5. Essentials for SBRT • 3-D Imaging device • Treatment planning system • Treatment delivery system • Precise targeting (stereotactic coordinate system) • Patient positioning verification

  6. WFUBMC SBRT3-D Imaging Device • Phillips PQ 5000 CT simulator • Single slice acquisition • External lasers for isocenter placement • AcQSim software

  7. WFUBMC SBRT“New” 3-D Imaging Device • GE Discovery ST CT/PET Simulator • 8 slice, helical scan, LightSpeed (Ultra) CT scanner • External lasers for isocenter placement • Advantage Simulation Workstation software

  8. WFUBMC SBRTTreatment Planning System • Phillips-ADAC 3D RTP system • Accurate small field dosimetry • Standard plan consists of 9 coplanar beams with negative margins

  9. WFUBMC SBRT • Varian 2100 SCX accelerator • 6MV photon beam • 120 leaf MLC • Portal Vision

  10. WFUBMC SBRTStereotactic Patient Positioning • Elekta Stereotactic bodyframe • Cartesian coordinate fiducials • Diaphragm breathing control • Patient must pass “YMCA” test • “Treat the box” approach

  11. Stereotactic Patient Positioning • Precise targeting requires a stereotactic coordinate system or image-based targeting • Patient positioning verification is a MUST.

  12. WFUBMC SBRTPatient Repositioning Verification • Elekta bodyframe lasers (2 chest points plus leg marker • Serial CT image sets on day of simulation • “GTV” is defined from the fusion of 2 image sets (i.e. “4-D GTV”)

  13. WFUBMC SBRTPatient Repositioning Verification • On day of treatment - SSD checks and orthogonal isocenter portal images • Physician approval of isocenter images required before treatment proceeds. • Total time on linac is about 1 to 1.5 hours

  14. WFUBMC SBRTOrthogonal Portal Images Fraction #1 Fraction #2 Fraction #3

  15. WFUBMC SBRTPatient Positioning Verification • Neuro-surgical Stealth Station computer system • Stereo IR camera • Fiducials placed on SBF and patient • System detects shifts in patient, relative to the SBF

  16. WFUBMC SBRTPatient Positioning Verification *Shifts mainly in axial direction on 3mm thick slices.

  17. WFUBMC SBRT • Through March, 2005, 70 SBRT patients have been treated using the Elekta Bodyframe. • Of these 70 patients, 30 patients have been treated in a single fraction on our Stereotactic Body Radiosurgery protocol. • Others have received between 2 and 4 fractions.

  18. WFUBMC SBRS PROTOCOL CCCWFU 99502: “A Phase I/II Dose-Escalation/Efficacy Study of Palliative Extracranial Radiosurgery using the Elekta Stereotactic Body Frame System.” Eligibility criteria: • The presence of a well-circumscribed tumor (primary or metastatic) on contrast-enhanced CT scan with a maximum diameter of 6 cm. • Age  18 years with a life expectancy  3 months. • No chemotherapy or immunotherapy were allowed at least 3 weeks prior to or planned for 4 weeks after treatment.

  19. WFUBMC SBRS PROTOCOL • Patients were stratified to dose levels by tumor volume • Phase I Endpoint: Acute toxicity using NCI Common Toxicity Criteria 3.0 • Phase II Endpoint: Local control by CT RECIST Criteria (Response Evaluation Criteria In Solid Tumors)

  20. WFUBMC SBRS PROTOCOL • 28 patients have been enrolled on study. 12 (43 %) of these patients have had PET imaging for a minimum of 3 months. We report the results of these patients. • For the PET group, median follow-up so far is 377 days (171-739 days). Median survival has not yet been reached. 3 deaths have occurred. For these 3 patients, median survival from time of treatment was 250 days.

  21. WFUBMC SBRS PROTOCOL • Local control at 3 months was 100 %. • The median change in tumor diameter at 3 months was -14 % (-59% to 0 %) for a median RECIST of SD. • The median corrected SUV change at 3 months was -57 % (-13% to + 11%).

  22. WFUBMC SBRS PROTOCOL • 8 patients have had no progressive disease (PD). • For 7/8, the SUV decreased and had not increased by the last follow-up. • 4 patients have had PD by RECIST. • 3/4 had a corresponding rise in correct SUV, which preceded the RECIST progression by a median of 71 days (71-75 days). • The median time to RECIST failure for these patients was 238 days (175-319 days). • Overall, the correlation between RECIST control and PET control was 0.71.

  23. WFUBMC SBRS PROTOCOL • In the acute phase (< 3 months), there appears to be no significant inflammatory response by PET. • In the late phase, (> 3 months) PET progression appears to precede RECIST PD by 2.4 months. This may have implications for the timing of subsequent therapy (e.g. chemotherapy).

  24. POSSIBLE IMPROVEMENTS • Respiratory gating • On-board KV imaging • Cone beam imaging

  25. THANK YOU