IAEA quality audits in radiotherapy

IAEA quality audits in radiotherapy. Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics Section, Division of Human Health. IAEA audits in radiotherapy. SINCE 1969: IAEA/WHO TLD postal dose audits of radiotherapy beam calibration (45 years of operation):

Download Presentation

IAEA quality audits in radiotherapy

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics Section, Division of Human Health

  2. IAEA auditsin radiotherapy • SINCE 1969:IAEA/WHO TLD postal dose audits of radiotherapy beam calibration (45 years of operation): 130 countries, >2000 hospitals, >10000 beam checks • SINCE 2005: Quality Assurance Team for Radiation Oncology (QUATRO) • SINCE 2010: TPS end-to-end audits

  3. 1969-2013: N = 10226 , m = 1.008, SD = 0.08, 83% res. within 5% limit ±5% 2012-2013: N = 1271 , m = 1.005, SD = 0.068, 92% res. within 5% Currently >90% results are within 5% acceptance limit

  4. TLD results within the 5% limit

  5. TLD results: use of CoPs (2000-2013) TLD results 2000-2013 126 deviations beyond 20% excluded

  6. IAEA TPS audit procedures Set of practical tests for dosimetry calculations reflecting basic treatment techniques in a typical radiotherapy hospital IAEA scientific secretary: S. Vatnitsky

  7. IAEA supported TPS audit in Europe • Eight European countries conducted TPS audits with the IAEA support • Approx. 60 hospitals participated • Approx. 200 datasets have been collected for beam/TPS algorithm combination • Several dosimetry and TPS-related problems were identified and fixed

  8. Comprehensive RT audit: QUATRO

  9. QUATRO audit procedures • Auditors: RO, MP, RTT • Typically 5 days per RT centre • Entrance briefing • Assessment: tour of facility, staff interviews, review & evaluation of procedures and documentation, measurements, tests, observation of practical work • Exit briefing: feedback to the department, preliminary recommendations, questions, discussion.

  10. QUATRO missions lower middle income upper middle income high income • Training of auditors and regional QUATRO workshops in all regions • 70 QUATRO missions to date: Africa – 6; Asia – 18; Europe – 30 + 3 re-audits; Latin America – 13

  11. Europe: recommendations by QUATRO • Support staff continuous development • Strengthen QA programmes and procedures • Equipment: get more, use better • Increase staffing levels: MP, RO, RTT • Staff requires training: RTT, RO, MP • Improve treatment protocols • Improve structure and organization of cancer centres

  12. THANK YOU

More Related