An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. What are we measuring with EEG and MEG James Kilner

  2. Outline • Why use EEG or MEG • What are EEG and MEG • A little history of EEG and MEG • What are we measuring

  3. Why use MEG or EEG?

  4. The measurements EEG MEG

  5. A little background - EEG

  6. "The electroencephalogram represents a continuous curve with continuous oscillations in which ... one can distinguish larger first order waves with an average duration of 90 milliseconds and smaller second order waves of an average duration of 35 milliseconds [Beta waves]. The larger deflections measure at most 150 to 200 microvolts...." A little background - EEG Hans Berger 1929

  7. “Caton has already published experiments on the brains of dogs and apes in which bare unipolar electrodes were placed either on the cereral cortex and the other on the surface of the skull. The currents were measured by a sensitive galvanometer. There were found distinct variations in current, which increased during sleep and with the onset of death strengthened, and after death became weaker and then completely disappeared. “ Hans Berger 1929

  8. A little background - MEG In 1963 Gerhard Baule and Richard McFee of the Department of Electrical Engineering,Syracuse University, Syracuse, NY detected the biomagnetic field projected from the human heart. They used two coils, each with 2 million turns of wire, connected to a sensitive amplifier. The magnetic flux from the heart will generate a current in the wire. Very noisy signal.

  9. Biomagnetic fields are very small compared to urban noise.

  10. In the late 1960’s David Cohen, at MIT, Boston recorded a clean MCG in an urban environment. This was possible due to: 1) Magnetically shielding the recording room. 2) Improved recording sensitivity. (The introduction of SQUIDS)

  11. MCG MEG

  12. SQUIDS The SQUID (Superconducting Quantum Interface Device) was introduced by James Zimmerman in the late 1960s. It is an ultrasensitive detector of magnetic flux. It is made up of a superconducting ring interrupted by one or two Josephson Junctions. It has been demonstrated that fields of the order of fT are within the scope of a SQUID.

  13. Brian Josephson, winner of the Nobel Prize for physics in 1972, demonstrated in 1962 the existence and very particular characteristics of the 'tunnel effect' that can be produced between two superconducting materials separated by a thin insulating layer. This is called a Josephson Junction. A Josephson Junction enables the flow of electrons – even in the absence of any external voltage.

  14. SQUIDS The SQUID has two generators of current flow. One a ‘classical’ current flow caused by the bias current across the SQUID The second is caused by the Josephson Junction and flows around the ring. The magnetic field measured causes a change in the current flow around the ring which will effect the overall current flow across the SQUID.

  15. The sensitivity of the SQUID to magnetic fields may be enhanced by coupling it to a superconducting pickup coil having greater area and number of turns than the SQUID inductor, alone. This pickup coil is termed a "flux" transformer". The pickup coil is made of superconducting wire and is sensitive to very small changes in the magnitude of the impinging magnetic flux. The magnetic fields from the brain causes a supercurrent to flow. Magnetometers First Order Gradiometer

  16. Inside the MEG dewar SQUIDs Sensors (Pick up coil)

  17. What are we measuring with EEG and MEG?

  18. Mainly Noise! Blinks, eye-movements, subject movement, heartbeat, sweat, electrode movement, swallowing, muscle activity, breathing, mains electricity, environmental noise, system noise ….. and a signal related to neuronal activity.

  19. The main source of the extracranial magnetic fields is current flow in the long apical dendrites of the cortical pyramidal cells. NOT action potentials. A distal excitatory synapse will induce a dipolar dendritic current towards the soma of the pyramidal cell, meaning that the electricity is flowing in one direction along the entire length of the dendrite, which therefore may be considered an electric dipole. Pyramidal neurons constitute nearly 70% of neocortical neurons, and the cells are oriented with their long apical dendrites perpendicular to the brain cortex. There are more than 100,000 of these cells per square millimeter of cortex. - Sink + Source

  20. The right hand rule

  21. Unlike EEG, MEG is only sensitive to tangential components. However the brain is not a sphere so all areas will have some tangential components.

  22. Summary • MEG and EEG are related measures • They mainly measure noise! • The neuronal signal reflects post-synaptic potentials of thousands of pyramidal cells. • One is not a better measure than the other – it will depend on what you are interested in.