Fiber Optic Sensorized Tools for Cardiology Applications

Fiber Optic Sensorized Tools for Cardiology Applications. July 7 th , 2008 Yong-Lae Park Seok Chang Ryu. Catheter Ablation. Technique used to treat cardiac arrhythmia (an abnormal heart rhythm). Destroys parts of the electrical pathways causing the arrhythmias. Radiofrequency ablation

morrisa
Download Presentation

Fiber Optic Sensorized Tools for Cardiology Applications

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Fiber Optic Sensorized Tools for Cardiology Applications July 7th, 2008 Yong-Lae Park Seok Chang Ryu

  2. Catheter Ablation • Technique used to treat cardiac arrhythmia (an abnormal heart rhythm). • Destroys parts of the electrical pathways causing the arrhythmias. • Radiofrequency ablation • Cryotherapy ablation

  3. Atrial Fibrillation • Atrial fibrillation is the commonest arrhythmia. • 30,000 to 35,000 catheter ablations for AF performed in the U. S. in 2006

  4. Need for Contact Force/Tactile Sensing • Contact force is one of the crucial factors which determine the lesion size. • A large survey reported (8745 patients) a complication rate of 6%. • 20% of the complications were caused by a perforation of the heart .

  5. Fiber Optic Force Sensing in Ablation Catheter • Multi-axis force sensing using embedded FBG sensors

  6. TactiCath from EndoSense • Fiber optic sensor within the catheter tip • Axial and lateral force measurement • Sensitivity of 1g • 10 Hz data acquisition rate

  7. 3 mm 1 mm 3 mm 0.5 mm Dynamic Tactile Sensing • Sensing texture of heart wall to collect more information on target areas. • Active sensing using piezo-actuator

  8. FBG Interrogator Piezo-actuator FBG Phase Locked Loop (PLL) Feedback Control • Lock the vibration frequency at resonant frequency • IFOS I-Sense can measure FBG signal at higher than 5 kHz.