1 / 19

Daniel R. McAlister and E. Philip Horwitz PG Research Foundation, Inc.

The Separation of Beryllium from Spectral Interfering Elements in Inductively Coupled Plasma – Atomic Emission Spectroscopic Analysis. Daniel R. McAlister and E. Philip Horwitz PG Research Foundation, Inc. 8205 S Cass Avenue, Suite 109 Darien, IL 60561. Problems with Current Method.

miriam
Download Presentation

Daniel R. McAlister and E. Philip Horwitz PG Research Foundation, Inc.

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The Separation of Beryllium from Spectral Interfering Elements in Inductively Coupled Plasma – Atomic Emission Spectroscopic Analysis Daniel R. McAlister and E. Philip Horwitz PG Research Foundation, Inc. 8205 S Cass Avenue, Suite 109 Darien, IL 60561

  2. Problems with Current Method Interfering elements in the AES spectrum of Beryllium Beryllium lines very intense  method is very sensitive for the determination of beryllium Interfering lines from other elements could lead to false positives.

  3. Uptake of Selected Elements on Dipex Resin Several EXC materials evaluated Dipex offered the most promising beryllium retention characteristics

  4. Uptake of Selected Elements on Dipex Resin Single column should remove all ICP-AES interferences Load pH 2-4 Rinse 0.2 M HNO3 Strip Be with 4 M HNO3 Potential for 2-10 x concentration of Be depending on sample size

  5. Uptake of Selected Elements on Dipex Resin Resin selective for Be over other alkaline earth metal ions Ca and Mg common matrix impurities

  6. Uptake of Selected Elements on Dipex Resin Fe(III) strongly retained Al(III) and Pb(II) have similar retention to Be(II) Cu(II) more weakly retained Large amounts of Fe(III) could interfere with Be(II) uptake

  7. Uptake of Selected Elements on Dipex Resin Resin selectively retains Be(II) over most other divalent metal ions Be(II) can be separated from samples containing large quantities of divalent metal ion impurities

  8. Proposed Method for Beryllium Purification Prepare samples as before (Digest with H2SO4/H2O2, dilute with HNO3) Neutralize samples to pH 1-2 with sodium aceate Buffers to maximum pH of 4.5 Monitor pH with methyl violet or crystal violet or pH strip pH Range over which separation is effective

  9. Elution of Be and Selected Elements on Dipex Resin

  10. Elution of Be and Selected Elements on Dipex Resin

  11. Elution of Be and Selected Elements on Dipex Resin

  12. Simulated Sample Results Successfully separated Be from samples digested with H2SO4/H2O2 Several different digested methods employed in Be analyses (HCl, H2SO4, HNO3, HF, H2O2, HClO4) High levels of impurities

  13. Digestion Methods using HF

  14. Separation of Be from Large Amounts of Interferences

  15. Separation of Be from Large Amounts of Interferences

  16. Uptake of Selected Metal Ions on LN2 and LN3 Resins U(VI), Th(IV), Ti(IV) retained from all [HNO3] LN2: Strongly retains Fe(III) LN3: Be not retained from pH < 2 Choice of LN2/LN3 depends on desired operating conditions and amount of Fe in samples

  17. Elution of Be on LN2 and LN3 Guard Columns

  18. Effect of Large Quantities of Uranium LN2 and LN3 Resins effectively increase the capacity for Uranium With LN2, the GC remains connected through the load, rinse and strip With LN3, the GC can be removed following the rinse

  19. Conclusions Efficient, Reliable method for purifying Be from all ICP-AES spectral interfering elements has been found using a single column Method is compatible with current monitoring and sample digestion methods Method is robust and performs over a wide pH range Inserting a LN2 or LN3 Resin guard column increases U capacity without changing the chemistry or significantly decreasing Be yields. Working to develop methods to isolate Be from high levels of other impurities

More Related