An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Cryodetector Readout for Dark Matter Searches Stuart Ingleby Cryodetectors Group, Oxford

  2. Cryodetector readout • Direct dark matter searches • Liquid noble gas • Cryodetectors • Cryogenic readout techniques • Low impedance – SQUIDs • High impedance – NTD/Ge sensors • Light detectors

  3. Cosmological evidence of dark matter • Baryon-to-photon ratio constrained • BBN • CMB power spectrum • Matter density constrained • Supernova redshift • CMB • Baryon acoustic oscillations • Overall • Baryons ~4% • Dark matter ~23% • Dark energy ~73%

  4. Astronomical evidence of dark matter • Galactic rotation curves • Expect 1/√r velocity curve • Observe ~linear • ‘Halo’ of DM • Alternative gravities • Bullet cluster • Collision of clusters • Observe galaxies, gas and overall mass separately • Consistent with CDM model

  5. Cryodetector experiments • Detect WIMP scattering • Nuclear recoils • Extensive shielding • Deep underground labs • Discrimination • Exclude electron events • Determine scattered nucleus

  6. Recent results • Exclusion plot • Long exposures and low event rate • Exclude more parameter space • Constrained MSSM theory • Filled area [1] • Aim 10-10pb (=10-46cm2) • Larger detectors • Lower backgrounds CRESST 2007 EDELWEISS II ZEPLIN III XENON10 CDMS SuperCDMS (dashed) [1] Trotta et al. 2008

  7. CRESST methods • 300g CaWO4 crystal • Phonons & scintillation at ~10mK • Light absorbed in separate silicon/sapphire wafer • Tungsten SPT in s/c transition • Coincident measurement of phonon & light • Recoils identified by quenching factor

  8. Low-impedance readout:SQUIDs • SQUID • Parallel Josephson junctions • V proportional to flux enclosed • Input coil • Current meter • S/C film stabilised within transition • Current biased • Small ∆T; large ∆I • Current read out using SQUID • SQUID voltage channel low-impedance

  9. Cryogenic cabling • CRESST SQUID cabling • Bespoke twisted-wire woven cables (right) • £400 / channel • Etched metal foil cabling • Conducting track defined by photolithography (below) • £60 / channel

  10. UV exposure unit Etching bath Laminator Etched metal foil cabling • Oxford Physics Photofabrication Unit • Phototool masks area to be etched • UV exposure • Developed to produce photo-resist layer • Etching removes resist-free areas • Max width 40cm • Max length 3m • New 1.2m laminator • Extra length can be achieved with multiple pressing- lower yield

  11. Etched cables Foil with photo-resist pre-etching Cable design for SQUID readout • Maximum yield • Even track width • Radiussed tracks • Teardropped contacts • 15 cables / etched sheet • Simplicity • Surface mount connectors • Durability • Laminate cover layer • Straight fold-free cables • Reinforcement of vulnerable areas

  12. Heatload • Larger detector mass • Lower heatload / channel • Choice of materials • Practical constraints • Resistivity measurements • Heatload calculations • Etched steel cables offer 20 x lower heatload

  13. Installation in K400 • Cryodetectors Lab Oxford • 6-channel SQUID system • Mounted at 4K • 2 x 12-channel etched foil cable • Custom hardware • Compact SQUID mount • Built around existing readout • Copper baffles for etched foil cables • SCSI connector box • Vacuum tight PCB flange with high channel density

  14. Low noise SQUID readout • SQUID baseline noise • Testing cryostat in Oxford • Intrinsic SQUID noise ~1 pA/√Hz (=1.2 μV/√Hz) • CRESST cables 1.55 pA/√Hz • Steel foil cables ~2.5 pA/√Hz • Extra noise • Nyquist noise on voltage channel?

  15. G B A B A B A A A B C D C D C D C C D H EDELWEISS method • Ge crystal 320g • 20mK operation • Phonon signal • NTD/Ge thermometer • Ionisation signal • ‘ID’ detector • Interleaved electrodes for charge capture • Fiducial volume • Reject surface events

  16. Cabling design for NTD/Ge • Readout for NTD/Ge • High impedance • Capacitance • Limits bandwidth • Microphonics • Mounted 4K – 10mK • Heatload minimised • Radiopurity

  17. Radiopurity measurements • Radiopurity tests • On samples of materials used • From GERDA, NEMO, CUORE experiments • Kapton has high 40K content • Steel wiring does not appear significant • 7.1% steel by mass • Polyethylene napthalate (PEN) suitable alternative • Prototyping and testing

  18. Light Detectors • CRESST light detector • Silicon on sapphire wafer • Cryodetector • Separate SQUID readout • Stabilised separately to phonon detector • High sensitivity • 20eV • Photomultiplier tube • Operated within cryostat • Simple high-impedance readout • Radiopurity • Light guides • HV supply • Voltage divider • Voltage multiplier

  19. HV supply for cold PMT • Cockcroft-Walton voltage multiplier • As seen in particle accelerators • Resistive voltage divider • Dissipative components add heatload • Possible noise on DC HV • Voltage multiplier chain • Can be designed and run efficiently at optimum frequency • Single-frequency supply can be chosen outside signal range • 2.9kV generated at 4K from 15V supply

  20. Component testing • Performance simulation • Approximate formulae available • Software simulation • Efficiency • Drop voltage • Transformer • Low-T component testing • Transformer • MPP • Capacitors • Polystyrene • Diodes • Silicon 1N4007

  21. Time / μs Number of events Pulse height / V Prototype PMT module • Installation of CWG-PMT module • Preliminary 57Co spectra taken at 300K • Detailed study of PMT performance for EURECA WP

  22. Future cryodetectors • Ton-scale experiments • EURECA • Greater exposure • Larger detector mass • Lower cost readout per module • Lower heatload per readout channel • Simplicity & reproducibility for mass production • Excellent discrimination • Ionisation • EDELWEISS ID detectors • Scintillation • Low-temperature light detectors