EPICS IOC using WindowsXP-based Oscilloscope for BPM DAQ System in KEK e-/e+ Linac

EPICS IOC using WindowsXP-based Oscilloscope for BPM DAQ System in KEK e-/e+ Linac. Masanori Satoh (KEK Linac) June 13-16, 2011 EPICS Collaboration Meeting in NSRRC. KEK e-/e+ Linac and Operation. Lepton Accelerator Complex in KEK Tsukuba Site.

Download Presentation

EPICS IOC using WindowsXP-based Oscilloscope for BPM DAQ System in KEK e-/e+ Linac

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. EPICS IOC using WindowsXP-based Oscilloscope for BPM DAQ Systemin KEK e-/e+ Linac Masanori Satoh (KEK Linac) June 13-16, 2011 EPICS Collaboration Meeting in NSRRC M. Satoh/ KEK Linac Control

  2. KEK e-/e+ Linacand Operation M. Satoh/ KEK Linac Control

  3. Lepton Accelerator Complex in KEK Tsukuba Site • Injector Linac: 8 GeV e-/3.5 GeV e+, 50 Hz, 0.1~10 nC • KEKB : 8 GeV e-/1400 mA, 3.5 GeV e+/2000 mA • PF: 2.5 GeV e-/450 mA • PF-AR: 6.5 GeV e-/60 mA • SuperKEKB : 7 GeV e-, 4 GeV e+ 2014 ~ • KEKB : Shutdown in June 2010

  4. PF Simultaneous Top-up for 3 rings • Beam current stability since Apr. 2009- KEKB e-/e+: 1 mA (~ 0.05%) : e-: 12.5 Hz, e+: 25 Hz • - PF: 0.05 mA (~ 0.01%) : 0.5 Hz KEKB e- KEKB e+ Luminosity M. Satoh/ KEK Linac Control

  5. BPM andOld DAQ System M. Satoh/ KEK Linac Control

  6. Linac BPM System • BPM x 100 for 600-m-long Linac • Stripline-type (four electrodes) • Position and Charge measurement • Orbit and Energy Feedback • DAQ System • x 20 DAQ Stations (One DAQ station typically handle BPMx4) • Digital Oscilloscope and VME CPU M. Satoh/ KEK Linac Control

  7. Oscilloscope (Tektronix TDS680B) • VME CPU (OS-9/ MC68060 50 MHz) • 5 GSa/s, 8bits, 2ch, 1 GHz BW, GPIB(low noise immunity) • DAQ rate:~1 Hz Vy1 Vy2 Vx2 Vx1 M. Satoh/ KEK Linac Control

  8. Analogue signal from BPM (KEKB e-) 7 ns M. Satoh/ KEK Linac Control

  9. Orbit and Energy FB Orbit Display Panel (SAD/Tk) HLA layer Middle server layer BPM DAQ layer M. Satoh/ KEK Linac Control

  10. New System(WindowsXP-based Oscilloscope) M. Satoh/ KEK Linac Control

  11. BPM DAQ Upgrade • Difficulty of maintenance (Old oscilloscope/VME) • Fast DAQ is strongly required (50 Hz) for simultaneous top-up operation • Hardware replacement in summer 2007 • Software upgrade => EPICS IOC in summer 2008 M. Satoh/ KEK Linac Control

  12. WindowsXP-based Oscilloscope (Tektronix DPO7104) • 10 GSa/s, 8bits, 4ch, 1 GHz BW, Ethernet connection • DAQ rate:> 50 Hz • EPICS IOC running on Oscilloscope (R3.14.9) 0.1/1 nC (200 mV/div) 10 nC (1 V/div) M. Satoh/ KEK Linac Control

  13. Oscilloscope performance M. Satoh/ KEK Linac Control

  14. EPICS PV Name Example • LIiBM:SP_A1_B8_1:XSNGL:KBE • LIiBM:BPM name:Property:Beam mode • Beam mode • KBE (KEKB e-) • KBP (KEKB e+) • PFE (PF e-) • ARE (PF-AR e-) • and more (Study mode for each beam) M. Satoh/ KEK Linac Control

  15. Synchronization w/ Beam mode EPICS PV name @ Event generator • It works well, so far • When network traffic is increased, …? M. Satoh/ KEK Linac Control

  16. (Python) PF KEB e- KEB e+ HLA layer Blade server NAS Archiver Middle server layer BPM DAQ x 24 (Win.XP Scope. EPICS IOC) M. Satoh/ KEK Linac Control

  17. Summary and Future Plan • The new BPM DAQ system has been developed: • EPICS IOC (WindowsXP-based digital oscilloscope) • Old system (20) has been replaced by new one (24) • Successful in stable beam position measurement (50 Hz) • Similar system was adopted for BPM DAQ of KEKB-BT • Towards to SuperKEKB injector upgrade: • Higher resolution (100 mm -> 10 mm?) • Several candidates are now under consideration • Libera Briliance Single Pass? • VME-based new one? M. Satoh/ KEK Linac Control

  18. Thank you for your attention! M. Satoh/ KEK Linac Control