1 / 29

ABX Pentra DF 120

ABX Pentra DF 120. Flexible hematology platform . For internal use only. ABX Pentra DF 120. Main features analysis Methods and technologies Pentra ML. Main features. Rack loader. Reagent compartment. Technical software. Processed tubes. Stat Sample Bay. ABX Pentra ML.

lysandra
Download Presentation

ABX Pentra DF 120

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ABX Pentra DF 120 Flexible hematology platform  For internal use only Prepared by L.Castellane

  2. ABX Pentra DF 120 Main features analysis Methods and technologies Pentra ML Prepared by L.Castellane

  3. Main features Rack loader Reagent compartment Technical software Processed tubes Stat Sample Bay ABX Pentra ML Optional slide maker and stainer Prepared by L.Castellane

  4. Main features • CBC, DIFF • IMmature Lymphocyte, Monocyte and Granulocyte • Optional and integrated slide maker and stainer • Expert Validation Station Prepared by L.Castellane

  5. Main features 32 Parameters • CBC (12 parameters) • Wbc, Rbc, Hgb, Hct, Mcv, Mch, Mchc, Rdw, Plt, Mpv, Pct, Pdw • Differential count (12+20 parameters) • Lym, Mon, Neu, Eos, Baso, Aly, Lic (% and #) • IML, IMM, IMG (% and #) • Throughput (12O samples per hour) Prepared by L.Castellane

  6. Main features • 5 reagents • ABX Diluent • ABX Basolyse • ABX Leucodiff • ABX Alphalyse or ABX Lysebio (Cyanide free) • ABX Cleaner • Unique blood control • ABX Difftrol - 3 levels • Target values on disk, multi-language package insert ABX reagents and controls are bar-coded and CE marked Batch number, level detection, small reagent volumes, waste level sensor Prepared by L.Castellane

  7. Main features • 3 types of analyses (with random access) • CBC • DIFF • SPS External and internal bar-code reader Prepared by L.Castellane

  8. ABX Pentra DF 120 Main features analysis Methods and technologies Pentra ML • CBC principles • 5 Diff – Double matrix Prepared by L.Castellane

  9. Methods & technologies RBC/PLT - Impedance method HGB - Cyanmethemoglobin or cyanide free WBC - (Triangulation) -> WBC channel (impedance) -> BASO channel (Specific lyse + impedance) -> 5Diff channel (optical) Prepared by L.Castellane

  10. Methods & technologies 5Diff Principles  • 1. Cytochemistry • Lysing of red blood cells • Staining of nuclei and cytoplasmic granules of leukocytes • Fixing the leukocytes in their original state (48 hours post-draw stability) • 2. Double Hydrodynamic Sequential System (DHSS) • Flow cytometer with hydrodynamic focusing before and after the micro-aperture • 3. Impedance method • Measurement of the cell volume (“X” axis)    • 4. Light absorbance • Internal structure (“Y” axis) Prepared by L.Castellane

  11. Methods & technologies Eosinophils Large Immature Cells Neutrophils Lymphocytes Monocytes Atypical LYmphocytes Double Diff Matrix IMmature Granulocytic line IMmature Monocytic line IMmature Lymphocytic line Prepared by L.Castellane

  12. Methods & technologies Metamyelocyte Myelocyte Promyelocyte Myeloblast IMmature Granulocytes IMG IMG% and # Identification of Granulocytic cell line count of the overall myelemia Prepared by L.Castellane

  13. Methods & technologies Promonocyte Monoblast IMmature Monocyte IMM IMM % and # Identification of Monocytic cell line Count of the precursors of the monocytic line Prepared by L.Castellane

  14. Methods & technologies IMmature Lymphocyte Lymphoblast Prolymphocyte IML IML % and # Identification of Lymphocytic cell line IML and ALY include all abnormal lymphocytic populations Prepared by L.Castellane

  15. Methods & technologies • Routine identification and quantification of 3 immature cell lines involved in leucopoiesis: • Granulocytic line • Monocytic line • Lymphocytic line -> Reduction of the number of slide examinations -> Diagnostic and follow-up tool for rapid decision-making Prepared by L.Castellane

  16. Methods & technologies Ne: (Neutro-eosino) Hyper-segmented neutrophils Hypo-granular eosinophils... ALY: (Atypical Lymphocytes) Atypical Lymphocytes Plasmocytes, hairy Cells... LIC: (Large Immature Cells) Blasts, myelocytes Promyelocytes Metamyelocytes... Mn: (Mono-neutro) Immature granulocytes Band cells... Nl: (Neutro-lympho) Hypo-segmented neutrophils Sezary cells, hairy cells... 1- Priority alarms 2- Separation alarms Prepared by L.Castellane

  17. Methods & technologies No: (Noise) Platelet aggregates, NRBCs Lysis resistant RBCs Ln: (Left-neutro) Cellular debris Ll, Ll1: (Left-lympho) Platelet aggregates, NRBCs Rn: (Right-neutro) Immature granulocytes Rm: (Right-mono) Large monocytes, blasts 3- Zone alarms Prepared by L.Castellane

  18. Methods & technologies CHRONIC MYELOID LEUKEMIA Pentra DF 120 result: NEU 25.4 % EOS 1.8 % LYM 5.1 % MON 8.6 % BAS 0.1 % IMG 49.3 % IMM 9.2 % IML 0.5 % ALY 1.3 % Prepared by L.Castellane

  19. Methods & technologies ACUTE MONOBLASTIC & MONOCYTIC LEUKEMIA Pentra DF 120 result: NEU 11.1 % EOS 0.5 % LYM 9.3 % MON 47.9 % BAS 0.1 % IMG 4.1 % IMM 26.1 % IML 0.9 % ALY 1.1 % Prepared by L.Castellane

  20. Methods & technologies T-CELL PROLYMPHOCYTIC LEUKEMIA Pentra DF 120 result: NEU 12.1 % EOS 0.4 % LYM 6.7 % MON 12.8 % BAS 0.4 % IMG 0.4 % IMM 2.0 % IML 7.4 % ALY 57.8 % Prepared by L.Castellane

  21. ABX Pentra DF 120 Main features analysis Methods and technologies Pentra ML • Specs & functionality Prepared by L.Castellane

  22. Specs & functionality Dedicated PC 17 inch flat colour screen Windows XP CD-ROM burner Automatic rerun Reflex testing Pre-diluted mode Cytology library Onboard user manual Prepared by L.Castellane

  23. Specs & functionality • Mono or bi-directional connection: • ASTM format • Transfer of • Demographic data (blood type - man, woman, child) • Tests requested (CBC, CBC/DIFF, SPS) • Up to 3 ABX Hematology analyzers (Pentra range) can be connected • Query mode • Download capability Prepared by L.Castellane

  24. Specs & functionality • User defined rules applied when : • receiving the worklist, • a sample has been analyzed. • _ Rerun capability • _ Reflex testing • _ Automatic validation • _ Messages, comments…. • Bio-technical validation with : • Patient history • Delta check • Result edition • Manual entry • Recalculation capability Prepared by L.Castellane

  25. Result screens • Bio-technical validation with : • Patient report, run, re-run, patient history, graphs, alarms. All displayed on the same screen. • Delta check • Result edition • Manual entry • Recalculation capability 1- Title bar 2- Patient and Graphics area 3- Messages 4- Patient Comments 5- Technical Comments 6- Extended Diff 7- Result area 8- Report 9- Runs Prepared by L.Castellane

  26. Specs & functionality Hematovision IV included A cell cytology library Prepared by L.Castellane

  27. ABX Pentra 120 series Prepared by L.Castellane

  28. ABX Pentra 120 series Prepared by L.Castellane

  29. Thank you Prepared by L.Castellane