An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. IPN OrsayR&D on Calorimeters Jean PeyréGöteborg - October 2008 IPNO-RDD-Jean Peyré

  2. 1.Milano conclusions in 20062.Lund conclusions in 20073.Magnetic Field4.Temperature dependence5.Xtalk of XP14D56.Quality of manufacturers7.Calorimeter Design8.Demonstrator9.Conclusions IPNO-RDD-Jean Peyré

  3. 22x22x22 22x22x220 44x22x200 66x22x200 XP5300B 5,74% 6,70% 5,86% 6,57% APD S8664-1010 8,23% 9,39% XP1912 8,05% 9,40% 10,33% 12,28% Energy Resolutions CsI(Tl)+VM2000+APD/PMT+137Cs MILANO oct 2006 IPNO-RDD-Jean Peyré

  4. Use of XP1452 foreseen in 2007 MILANO oct 2006 • XP1452  38 mm (Active area 4x272mm2), green extended bialkaly • Quantum efficiency 30% at 420nm, 20% at 550nm • Size from 35mm to 50mm Next tests with XP1452 XP1452 IPNO-RDD-Jean Peyré

  5. Conclusions of Milano Oct 2006 MILANO oct 2006 Calculated Resolutions Xtal 22x22x220 @550nm IPNO-RDD-Jean Peyré

  6. Conclusions of Milano Oct 2006 MILANO oct 2006 • Collected light through long crystals is not constant. Tests with non collimated 137Cs source give resolutions greater that 15%. • The choice of VM2000 seems obvious • With XP1912, it seems difficult to have a resolution better than 9%-10% at 662keV on long crystals because: • Area covered by PMT is only 36% with 22x22 output face • Intrinsic resolution of crystal is quite high (~4/5%) • Photocathode is not green extended bialkaly • Resolution obtained with PMT XP1912 and APD are quite comparable until 3.25MeV • No work has been done on surface quality (polished/rough) • Use of XP1452 seems promising • Tests on LaCl3/LaBr3 crystals • Litrani simulations to be explored IPNO-RDD-Jean Peyré

  7. 1.Milano conclusions in 20062.Lund conclusions in 20073.Magnetic Field4.Temperature dependence5.Xtalk of XP14D56.Quality of manufacturers7.Calorimeter Design8.Demonstrator9.Conclusions IPNO-RDD-Jean Peyré

  8. Results for PMT + 137Cs Crystal from St-Gobain 22x22x220 LUND june 2007 Crystal 22x22x220 Before modification by St Gobain Output face polished and others unpolished After modification by St Gobain Polished (brown), unpolished (red) IPNO-RDD-Jean Peyré

  9. Results for PMT + 137Cs Crystal from St-Gobain 22x22x220 LUND june 2007 IPNO-RDD-Jean Peyré

  10. Results for PMT + 137Cs Crystal from St-Gobain/Amcrys/Scionnix 22x22x220 LUND june 2007 IPNO-RDD-Jean Peyré

  11. St-Gobain Amcrys Scionnix 22x22x22 5,74% 5.96% 5,75% 22x22x220 6,71% 6,33% 6,29% Energy Resolutions CsI(Tl)+VM2000+PMT+137Cs Crystal from different companies with XP5300B LUND june 2007 IPNO-RDD-Jean Peyré

  12. Energy Resolution for some Crystals LUND june 2007 CsI(Na) 137Cs XP5300B 22x22x22 60Co CsI(Tl) 22x22x22 LaCl3 Diam25x25 56Co Energy resolution LaBr3 Diam25x25 IPNO-RDD-Jean Peyré

  13. Conclusions of Lund June 2007 LUND june 2007 CsI(Tl) • For CsI(Tl), the basic resolution for different manufacturers is quite comparable for small Crystals, but the wrapping and opticical treatement for long crystals seem at that time to be a very difficult science for St Gobain • Energy resolution of 5/5.5% seams reachable with XP1452 / XP14D5 at 1MeV New Crystals • Even if Labr3/LaCl3 is very promising, St Gobain seems at that time to have great difficulties to make long crystals. • Good Energy and Time resolution on small Crystals • There are no competitors on this market at that time • Are they interested by designing special crystals for physics experiments ? IPNO-RDD-Jean Peyré

  14. 1.Milano conclusions in 20062.Lund conclusions in 20073.Magnetic Field4.Temperature dependence5.Xtalk of XP14D56.Quality of manufacturers7.Calorimeter Design8.Demonstrator9.Conclusions IPNO-RDD-Jean Peyré

  15. Magnetic field R3B R3B Calorimeter Target Xm Zc 2.2T Magnet Yc Xc Beam axis Ym 14° Zm 1450mm (0,0,0) m More than 600 Gauss IPNO-RDD-Jean Peyré

  16. Magnetic field R3B R3B Calorimeter Target Xm Zc 2.2T Magnet Yc Xc Beam axis Ym 14° Zm 1450mm (0,0,0) m Iron cylinder IPNO-RDD-Jean Peyré

  17. Magnetic field R3Bwith shielding 10cm thick (R<50cm) Magnetic field in calorimeter coordinate for different Yc Coordinates (cm) Magnetic field out of shielding > 1500 Gauss Yc=65 Max value of Magnetic field in each vertical plane Yc Yc (cm) B without shielding(Gauss) B with shielding 2cm(Gauss) B with shielding 10cm(Gauss) IPNO-RDD-Jean Peyré

  18. 1.Milano conclusions in 20062.Lund conclusions in 20073.Magnetic Field4.Temperature dependence5.Xtalk of XP14D56.Quality of manufacturers7.Calorimeter Design8.Demonstrator9.Conclusions IPNO-RDD-Jean Peyré

  19. Temperature dependance Michaël JOSSELIN CONCLUSION Variation of Xtal+detector response at 20°C (according to NIm papers) • PMT + CsI(Tl) → +0.2 to 0.4%/°K • APD + CsI(Tl) → +2.3 to 3.8 %/°K IPNO-RDD-Jean Peyré

  20. 1.Milano conclusions in 20062.Lund conclusions in 20073.Magnetic Field4.Temperature dependence5.Xtalk of XP14D56.Quality of manufacturers7.Calorimeter Design8.Demonstrator9.Conclusions IPNO-RDD-Jean Peyré

  21. XP1452 Test of new PMt XP14D5 4 Channels PMt 2 Channels PMt XP14D5 IPNO-RDD-Jean Peyré

  22. Xtalk between 2 channels saw line 3mm +saw line 2mm 3mm 3 PMT’s -1 with window thickness 3mm -1 with window thickness 2mm -1 with window thickness 3mm + saw line IPNO-RDD-Jean Peyré

  23. CH1 CH2 CH1 CH2 Xtalk between 2 channels We want to measure the Crosstalk between the two channels of the PMT when placing a crystal in front of one of the two channels. Case A S1a=G1*NbPhé1a S2a=G2*NbPhé2a NbPhé= Number of Photoélectrons created on each part of the photocathode & collected by first dynode G= global Gain for each Channel (PMT+ChargePream+Shaper) S= signal for each channel CH Cross-Talk CH1->CH2 NbPhé2a/NbPhé1a=S2a/S1a*G1/G2 Case B S1b=G1*NbPhé1b S2b=G2*NbPhé2b Cross-Talk CH2->CH1 NbPhé1b/NbPhé2b=S1b/S2b*G2/G1 IPNO-RDD-Jean Peyré

  24. Xtalk between 2 channels Cross-Talk CH2->CH1 Cross-Talk CH1->CH2 Are supposed equal (estimation of ±12% variation between both) Cross-Talk = (S1b/S2b*S2a/S1a)1/2 Xtalk1= standard Crosstalk Xtalk2= standard Crosstalk measured with saw line filed with white silicon The standard production of PMT will certainly provide a quite constant crosstalk between channels: easy to correct IPNO-RDD-Jean Peyré

  25. 1.Milano conclusions in 20062.Lund conclusions in 20073.Magnetic Field4.Temperature dependence5.Xtalk of XP14D56.Quality of manufacturers7.Calorimeter Design8.Demonstrator9.Conclusions IPNO-RDD-Jean Peyré

  26. Xtals ordered from other manufacturers New crystals received :- Amcrys (Russia) and Siccas (China) IPNO-RDD-Jean Peyré

  27. Results CsI(Tl)+VM2000+XP5300B+137Cs SICCAS Xtal Amcrys Xtal IPNO-RDD-Jean Peyré

  28. Energy Resolutions CsI(Tl)+VM2000+PMT+137Cs 22x22x22 22x22x220 44x22x200 66x22x200 XP5300B Amcrys Amcrys St-Gobain St-Gobain 5,96% 6,33% 5,86% 6,57% 19x38x19 19x38x110 19x38x130 XP5300B Amcrys Amcrys Amcrys 6,1% 6,3% 6,4% Measurements can be inproved as we have received new PMT’s XP1452 -> XP14D5 100μA/lm-> 150 μA/lm XP1452 Amcrys Amcrys Amcrys 7,1% 7,34% 7,5% IPNO-RDD-Jean Peyré

  29. Energy Resolutions CsI(Tl)+VM2000+PMT+137Cs Measures made with optimized PM (Photonis #1014 tube) and collimated sources IPNO-RDD-Jean Peyré

  30. Measurements with APD’s Many difficulties to obtain low noise signals if the APD touches the hole made for the crystal => bad resolution g source 137Cs FWHM > 10 % Signal + noise of an APD too near from the hole EM shield box CsI (Tl) crystal APD IPNO-RDD-Jean Peyré

  31. Measurements with APD’s Influence of the gain / shield The APD 12 mm inside the shield box 12 mm IPNO-RDD-Jean Peyré

  32. 1.Milano conclusions in 20062.Lund conclusions in 20073.Magnetic Field4.Temperature dependence5.Xtalk of XP14D56.Quality of manufacturers7.Calorimeter Design8.Demonstrator9.Conclusions IPNO-RDD-Jean Peyré

  33. Common R&D for the EXL and R3B Calorimeters Basic design : CsI(Tl) & Studies on alternative solutions IPNO-RDD-Jean Peyré

  34. 3.0° 2.0° 1.7° 1.5° 1.2° Σ 24.0° Σ 16.0° Σ 16.0° Σ 13.6° Σ 12.0° Σ 9.6° 90° 106° 74° Polar angles 130° 60.4° 48.4° 38.8° 3.0° 2.0° 1.7° Beam 1.5° 1.34° 1.2° Target Barrel EndCap Common R&D for the EXL and R3B Calorimeters Design guided by Doppler shift IPNO-RDD-Jean Peyré

  35. Possible Calorimeter Design - R&D Example of Panda calorimeter IPNO-RDD-Jean Peyré

  36. Possible Calorimeter Design - R&D Square PMT (2 per envelop) Tests on a prototype Rectangular Crystals IPNO-RDD-Jean Peyré

  37. 1.Milano conclusions in 20062.Lund conclusions in 20073.Magnetic Field4.Temperature dependence5.Xtalk of XP14D56.Quality of manufacturers7.Calorimeter Design8.Demonstrator9.Conclusions IPNO-RDD-Jean Peyré

  38. Prototype R3B-EXL IPNO-RDD-Jean Peyré

  39. Prototype R3B-EXL Rectangular Crystals Square PMT (1 per 2 crystals) We have ordered 18 CsI(Tl) crystals from Amcrys and 9 PMts from Photonis. All material in November at IPNO (CsI(Tl)+Dual PMT). Base design made at IPN Orsay. IPNO-RDD-Jean Peyré

  40. Prototype R3B-EXL In 2008, construction of a prototype of 18 Xtals [CsI(Tl)] with 9 square PMT • Simulation and Tests of the prototype with different thickness of material in between the crystals depending on mechanical and physics considerations. IPNO-RDD-Jean Peyré

  41. Prototype R3B-EXL We have received 18 CsI(Tl) crystals from Amcrys IPNO-RDD-Jean Peyré

  42. Prototype R3B-EXL Test signal input Ch1 output Ch2 output Ch1_CPA output Bases are ready Low voltage Still waiting for PhotoMultiplier Tubes Commun HV Ch2_CPA output IPNO-RDD-Jean Peyré

  43. Prototype R3B-EXL Ch1_CPA output Bases are ready Low voltage Still waiting for PhotoMultiplier Tubes IPNO-RDD-Jean Peyré

  44. 1.Milano conclusions in 20062.Lund conclusions in 20073.Magnetic Field4.Temperature dependence5.Xtalk of XP14D56.Quality of manufacturers7.Calorimeter Design8.Demonstrator9.Conclusions IPNO-RDD-Jean Peyré