Therapeutic Ultrasound

Loading in 2 Seconds...

play fullscreen
1 / 27

Therapeutic Ultrasound - PowerPoint PPT Presentation


  • 1078 Views
  • Uploaded on

Therapeutic Ultrasound Clinical Application FDA Labeling Requirements Output frequency Effective Radiating Area (ERA) Beam Nonuniformity Ratio (BNR) Beam profile Date of last service/ calibration should also be posted on the unit Treatment Area US heats a limited area

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about 'Therapeutic Ultrasound' - liam


Download Now An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
therapeutic ultrasound

Therapeutic Ultrasound

Clinical Application

fda labeling requirements
FDA Labeling Requirements
  • Output frequency
  • Effective Radiating Area (ERA)
  • Beam Nonuniformity Ratio (BNR)
  • Beam profile
  • Date of last service/ calibration should also be posted on the unit
treatment area
Treatment Area
  • US heats a limited area
    • About the size of a catsup packet
  • Treatment area should be 2 to 3 times the size of the ERA
  • For larger areas, divide the treatment area into smaller zones
coupling methods

Coupling Methods

Direct CouplingImmersion MethodPad/Bladder Method

coupling methods5
Coupling Methods
  • Ultrasonic energy cannot pass through the air
  • A coupling medium is required
  • Medium should be water-based
  • Coupling method should confirm to the body area
  • The body area should be clean and relatively hair-free
direct coupling
Direct Coupling
  • Gel or Creams
  • Only use approved coupling agents
  • Apply liberally to area
  • Remove air bubbles by passing sound head over area (before power is increased)
direct coupling7
Direct Coupling
  • Move the sound head s-l-o-w-l-y
    • 4 cm/sec
  • Moving the head faster decreases heating
  • If the patient describes discomfort, decrease the output intensity
coupling ability of various media
Coupling Ability of Various Media

Substance Transmission

  • Saran Wrap 98
  • Lidex ge, fluocinonide (.05%) 97
  • Thera-Gesic 97
  • Mineral oil 97
  • US Transmission gel 96
  • US Transmission lotion 90
  • Chempad-L 68
  • Hydrocortisone powder (1%) 29
  • Hydrocortisone powder (10%) 7
  • Eucerin cream 0
  • Myoflex 0
  • White petrolatum gel 0
immersion technique
Immersion Technique
  • Used to treat irregularly shaped areas
  • The limb is immersed in a tub of degassed water
  • If tap water is used, increase the output intensity by 0.5 w/cm2
  • Transducer is held appx. 1” from the body part
  • Avoid the formation of air bubbles
pad bladder method
Pad (Bladder) Method
  • A mass of conductive gel
    • Commercial pads
    • Self-made bladders
  • Conforms to the treatment area
  • Commercial pads help limit the size of the treatment area
output parameters

Output Parameters

Output FrequencyDuration

Duty CycleOutput Intensity

output frequency
Output Frequency
  • Determines the treatment depth
  • 1 MHz Output
    • Deep (5 to 7 cm) tissues
      • Rotator cuff, vastus intermedius, gastroc
  • 3 MHz Output
    • Superficial (up to 3cm deep) tissues
      • Patellar tendon, MCL, brachialis
  • Remember that adipose tissue is transparent to ultrasound
treatment duration
Treatment Duration
  • Depends on:
    • Size of the treatment area
    • Output intensity
    • Therapeutic goals
  • Vigorous heating
    • 1 MHz output
      • 10 to 12 minuts
    • 3 MHz output
      • 3 to 4 minutes
  • Also see “Output Intensity”
duty cycle
Duty Cycle
  • Determine the proportion of thermal and nonthermal effects
    • High duty cycle: Predominantly thermal effects
    • Low duty cycle: Predominantly nonthermal effects
  • Thermal effect used in subacute and chronic conditions
  • Nonthermal effects may be beneficial in acute stages
output intensity 1 mhz
Output Intensity – 1 MHz
  • At a 100% duty cycle
  • 1.75 W/cm2 output
  • Approximately 4 min treatment is required to increase tissue temp 4°C
output intensity 3 mhz
Output Intensity – 3 MHz
  • At a 100% duty cycle
  • 1.5 W/cm2 output
  • Approximately 10 min treatment is required to increase tissue temp 4°C
us electrical stimulation
US/Electrical Stimulation
  • Theoretically combines the benefit of the two treatments
  • Used to treat:
    • Trigger points
    • Muscle spasm
  • Output parameters:
    • Thermal ultrasound
    • Motor level electrical stimulation
contraindications
Acute injuries (100% duty cycle)

Ischemic areas

Areas of impaired circulation including arterial disease

Over areas of deep vein thrombosis

Anesthetic areas

Over cancerous tumors

Over sites of active infection or sepsis

Over the spinal cord or large nerve plexus in high doses

Exposed metal that penetrates the skin (e.g., external fixation devices)

Areas around the eyes, heart, skull, or genitals

Over the thorax in the presence of an implanted pacemaker

Pregnancy when used over the pelvic or lumbar areas

Over a fracture site before healing is complete

Stress fracture sites or sites of osteoporosis

Over the pelvic or lumbar area in menstruating female patients

Contraindications
precautions
Precautions
  • Symptoms may increase after the initial treatments.
  • Use caution when applying ultrasound around the spinal cord, especially after laminectomy.
  • The use of ultrasound over metal implants is not contraindicated
    • Keep the sound head moving
  • Use caution when applying ultrasound over epiphyseal plates of growing bone
instrumentation
Instrumentation
  • Duty cycle: Adjusts between continuous and pulsed ultrasound application
  • Frequency: Selects the depth of penetration
  • Gel warmer: Used to preheat the transmission gel, primarily for patient comfort
  • Intensity: Adjusts the output intensity.
    • The WATT METER displays the output in either total watts or watts per square centimeter.
  • Maximum head temperature: Prevents overheating in case the head is not properly coupled.
  • Pause: Interrupts the treatment but retains the remaining amount of treatment time
  • Power: Turns the unit ON or OFF
  • Start-Stop: Initiates or terminates the production of ultrasound from the transducer.
  • Timer: Sets the duration of the treatment.
  • Watt meter: Displays the output of ultrasound in total watts or watts per square centimeter.
patient preparation
Patient Preparation
  • Establish that no contraindications are present
  • Determine the method and mode of ultrasound application to be used
  • Clean the area to be treated
  • Determine the type of coupling method to be used
  • Identify a treatment area that is 2 to 3 times the ERA
  • For direct coupling: spread the gel over the area to be treated
  • Explain the sensations to be expected during the treatment:
    • Mild to moderate warmth (but not pain or burning) for thermal treatments
    • No subcutaneous sensations during pulsed ultrasound
  • For thermal treatments pre-heating with a moist heat pack will decrease the treatment time required to reach vigorous heating levels
  • Advise the patient to report any adverse, unusual, or painful sensations during the treatment.
initiation of the treatment
Initiation of the Treatment
  • Reduce the INTENSITY to zero before turning on the POWER.
  • Select the appropriate mode for the output (CONTINUOUS or PULSED)
  • Set the WATT METER to displays the appropriate output for the type of treatment
  • Set the TIMER to the appropriate treatment duration
  • Begin slowly moving the sound head over the medium
    • Press the START button to begin the treatment session
    • Units having low BNR may be moved at a slower rate than those with a higher BNR
  • Slowly increase the INTENSITY
    • Keep the sound head moving
  • Move the head at a moderate pace (4 cm per second or slower)
  • Use firm, yet not strong, overlapping strokes
  • If pain is experienced:
    • Move the sound head at a faster rate
    • Use a lower duty cycle
    • Use a lower the intensity
  • If the gel begins to wear away or if the sound head begins sticking on the skin, depress the PAUSE button and apply more gel
phonophoresis application
Phonophoresis Application
  • Preheating the area to increase the absorption of the medication
  • Use only approved ultrasound transmission media.
    • The direct coupling method is recommended
    • Efficacy of phonophoresis using the bladder method has not been established
  • Ensure that the skin is well moistened; avoid areas of dry skin
  • Position the extremity to encourage circulation.
  • Use a continuous output to maximize the effect of phonophoresis
    • Unless the thermal effects of ultrasound are contraindicated
  • Follow the procedures described in “Initiation of the Treatment”
  • After treatment, cover the remaining medication with an occlusive dressing.
termination of the treatment
Termination of the Treatment
  • If the treatment is being terminated prematurely:
    • Reduce the intensity before removing the transducer
  • Clean the remaining gel from the patient’s skin
  • To ensure continuity of treatment sessions:
    • Record the treatment parameters
      • Output frequency
      • Intensity
      • Duration
      • Duty cycle
      • Running count of ultrasound treatments given for this condition
  • Immediately initiate any post-treatment stretching
maintenance
Maintenance

Daily Maintenance

  • Clean ultrasound head and transducer face as recommended by the manufacturer

Monthly Maintenance

  • Check all electrical cords for frays or kinks
  • Check the sound head cable for frays or kinks
  • Clean the transmitter face as recommended by the manufacturer.

Federal regulations require that therapeutic ultrasound units be recalibrated annually by an authorized service technician