An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The Synchrony IOL is not currently FDA-approved OPTICAL QUALITY OF DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS 1 Fundacion Valle del Lili, Cali-Colombia 2 Universidad Industrial de Santander, Bucaramanga-Colombia 3 Universidad Autonoma de Bucaramanga, Bucaramanga-Colombia 4 Abbott Medical Optics Inc, Irvine, USA Financial Disclosure a Consultant Abbott Medical Optics b Scientist at Abbott Medical Optics Inc. • Andrea Galvis, MD 1,3 a • Ivan Ossma, MD MPH MSc 1,2, a • Sanjeev Kasthurirangan, PhD 4, b

  2. 2 Synchrony Design • Anterior Optic 32 Diopters • Posterior Optic Divergent (negative)

  3. Study Purpose • Compare Optical Quality Pseudophakic Eyes • Monofocal IOL • Synchrony Dual Optic Accommodating IOL • Parameters • Objective iTrace aberrometry • Custom Matlab Software to calculate MTF • Subjective Contrast Sensitivity Testing • Optec 6500

  4. Study Population • iTrace aberrometry • Cycloplegia • 3 mm zone • Matlab Software to calculate MTF • Monofocal IOL : 15 subjects (30 eyes) • Synchrony IOL : 16 subjects (32 eyes)

  5. iTrace - Matlab conversion

  6. Results

  7. Difference 0.01 um Objective Measurements3 mm pupil

  8. Best-focus MTF 3 mm pupil Differences are less than 10%

  9. 9 Mesopic Contrast Sensitivity p=0.61 Wilcoxon Rank Sum Test

  10. Discussion • Dual Optic Design DOES NOT affect optical or retinal image quality of the eye • Objective evidence (iTrace, MTF) • Subjective evidence (CSF) • Contrast sensitivity of Synchrony is equivalent to that of monofocal IOLs

  11. Discussion • Effect of the slight spherical aberration on Synchrony would only yield 0.42 D of increased depth of focus • Calculations based on Rocha’s work * • The improved near vision of Synchrony patients MOST LIKELY due to true accommodation * Rocha KM. Using Adaptive Optics Simultion to Explore the Effect of Higher Order Aberrations on Depth of Focus for the Correction of Presbyopia. Wavefront Congress 2008

  12. Conclusion • Quality of Vision of Synchrony IOL similar to standard monofocal IOLs • Measured Subjectively • Measured Objectively • Improved Near Vision with Synchrony • Not related to increased depth of focus