Progress on Table Top Neutron Source for the Calibration of Nuclear Recoil Detectors

Loading in 2 Seconds...

play fullscreen
1 / 13

Progress on Table Top Neutron Source for the Calibration of Nuclear Recoil Detectors - PowerPoint PPT Presentation


  • 60 Views
  • Uploaded on

Progress on Table Top Neutron Source for the Calibration of Nuclear Recoil Detectors. Dante Nakazawa with Prof. Juan Collar For Physics 335 with Prof. Mark Oreglia. Reminder of Project.

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about 'Progress on Table Top Neutron Source for the Calibration of Nuclear Recoil Detectors' - komala


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
progress on table top neutron source for the calibration of nuclear recoil detectors

Progress on Table Top Neutron Source for the Calibration of Nuclear Recoil Detectors

Dante Nakazawa

with Prof. Juan Collar

For Physics 335 with Prof. Mark Oreglia

reminder of project
Reminder of Project
  • Facility of Table Top Neutron Source : Calculation of neutron response of scintillating crystals. Ability to discern the direction of recoil.

Proof of Principle

  • Calculate neutron energy spectrum of 252Cf neutron source with time-of-flight spectroscopy.
  • Using this initial neutron energy scattering angle to calculate the energy of the recoiling atom (Na or I).
  • Measure the known scintillation response of a NaI(Tl) crystal to neutron recoils w.r.t. gamma recoils.
experimental setup
Experimental Setup

NaI(Tl) Coupled to

Photomultiplier Tube

Shielding

Pb

Polyethylene

Borated Silicone

NaI(Tl) Crystal

D1A

252Cf

60 cm

D2

D1B

6LiI Crystal

60 cm

D3

when does a fission occur
When does a fission occur?
  • Each fission from 252Cf emits 20 gammas and 4 neutrons
  • PNNL solution : try to cover at least 3 of the source with gamma detectors (also NaI in our case) and trigger from a coincidence.

NaI(Tl) Coupled to

Photomultiplier Tube

252Cf

shielding

coincidences
Coincidences

PMT voltage

D1A

D1B

time

event rate estimation
Event Rate Estimation
  • 252Cf source :
    • Decay Rate 0.1737 GBq (109 decays per second)
    • 3.1% into fissions, 5.385 MBq
  • Measured :
    • 337.5 kBq, roughly 10% of events. This seems fine considering not all gammas, incident on the phototubes, will be above triggering threshold.
logic electronics
Logic Electronics

D1A

D1B

10x Amp

10x Amp

discriminator

discriminator

AND Logic Unit

coupled the 2 nd and 3 rd detectors
Coupled the 2nd and 3rd Detectors

D2

D3

6LiI Crystal

NaI(Tl) Crystal

Fast PMT (response < 1 ns)

High Speed Amplifier

6 lii ideal for 3 rd detector
6LiI : Ideal for 3rd Detector
  • Do not need PSD to discriminate neutrons from gammas. Use the (n,alpha) reaction.
measured 6 lii spectrum with a multi channel analyzer
Measured 6LiI Spectrum with a multi channel analyzer

Y-88 gamma lines at 898 keV and 1.836 MeV

Co-60 gamma lines at 1.1 and 1.3 MeV

Am-Be

gamma line at 4.4 MeV, neutrons at 4-8 MeV

Cf : 80% of gammas are below 1 MeV. (n,alpha) reaction

in 6Li peaks at 3 MeV so pulse height discrimination will work.

problems
Problems
  • Shifting baseline from Timing PMTs (D1A and D1B). Probably need to look for coincidences offline in the data analysis phase.
  • 6LiI Detector we have may not have the efficiency necessary for nanosecond timing needed for 252Cf spectrum measurements. Need to build a liquid or plastic scintillator and use pulse shape discrimination.
to do
To do :
  • Once we have timing coincidence nailed down and another scintillator, we can acquire spectrum of 252Cf.
  • Develop the Data Acquisition Software (LabVIEW):
    • Trigger from 3rd detector once well shielded against background neutrons. Store waveform of D2 and coincidence detectors.
    • Check timing information from D3, D2, and coincidence logic for energy conservation per event.
    • Calculate nuclear recoil energy from initial energy (time-of-flight) and scattering angle.
    • Integrate the pulse of 2nd detector.
    • Acquire NaI(Tl) gamma response and calculate the scintillation efficiency.