An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. PHOS, the ALICE PHOton Spectrometer The scientific objectives The design Status & Performance Daicui Zhou (for the ALICE/PHOS Collaboration) Institute of Particle Physics, Central China Normal University, Wuhan,China QM2006 Daicui Zhou

  2. PHENIX Phys. Rev. Lett. 88 (2002) 022301 The role of photons • SPS and RHIC data have shown the importance of the photon signal to diagnose the properties of matter formed in heavy-ion collisions. • Jet quenching first observed through high ptp0 suppression • Observation of high pt direct photon, constraining pQCD • Observation of low pt photon excess which could be interpreted in terms of thermal photons emission WA98, Phys. Rev. Lett. 85 (2000) 3595 PHENIX, Nucl-Exp/0605005 QM2006 Daicui Zhou

  3. … and at LHC • Extending the photon measurement in the very hard regime ( ~ 100 GeV/c) • More favorable environment for direct photon identification (larger g/p0, isolation cuts) • Provide a calibration of the jets through photon-tagged jet measurements • A hotter and longer lived QGP should emit a measurable thermal radiation QM2006 Daicui Zhou

  4. Requirements • High quality photon identification in a hadron dominated environment (dN/dy ~ 2000-4000) • CPV for h± and e± - g discrimination • time of flight for h (mainly n and n) - g discrimination • Shower shape and isolation cut for p0 - g discrimination • This can be provided by • A high granularity EM calorimeter providing excellent energy and position resolution and timing measurement • Associated to a charged particle detector QM2006 Daicui Zhou

  5. ALICE-PHOS Detector ALICE EMCA: 5 arrays of 56×64 Lead tungstate crystals (small radiation length and Moliere radius) CPV: multi-wire chamber with pad readout matching the granularity of EMCA PHOS QM2006 Daicui Zhou

  6. Avalanche Pin-Diode Crystal Strip unit Preamplifier module Support structure Technical data • 17920 PWO crystals • distance to IP: 4.6m • coverage in pseudo-rapidity: |Dh| < 0.12 • coverage in azimuthal angle: DF < 100° • crystal size:22×22×180 mm3 • Depth: 20X0 • photo readout: APD + CSP • operating temperature: -25 oC PHOS Module QM2006 Daicui Zhou

  7. Readout and electronics • Cold volume: • PWO strips 2x16=32 channels array • APD+CSP • FEE boards working at room temperature 112 FEEs / 8 TRUs / 4 RCUs / 3584 CSP-APD QM2006 Daicui Zhou

  8. Signal Processing Scheme NIM A565(2006) 768 CERN/ALICE 99-4, ALICE TDR2 QM2006 Daicui Zhou

  9. FEE Card • 32 channel FEE card • V1.0 prototype in Aug. 2004 and test beam at CERN • V1.1 version in Jan. 2005 and review in May 2005 • 130 cards production and testing at Wuhan in Jan 2006 • Commissioning and test beam at CERNfrom June-Sept 2006 FEE card hardware properties • 32 ch.dual gain shapers τ=1μs • RMS noise 2 us: 615 e-(3.1MeV) • 14 bit dyn range 5 MeV-80GeV • 32 APD bias regulators ±0.1V • Fast OR (2*2 Xtals) for trigger • Board controller FPGA (PCM) • USB controller • TPC-like readout and control bus • DAQ and DCS via RCU • 5.5 Watt,349*210 mm2 NIM A565(2006) 768 CERN/ALICE 99-4, ALICE TDR2 FEE Card QM2006 Daicui Zhou

  10. QM2006 Daicui Zhou

  11. 1st Module beam test at PS-T10 3584 crystals 112 FEE cards Cooling system First PHOS module Assistant cooling ALICE-PHOS first Module QM2006 Daicui Zhou

  12. Calibration Stability • The balance test of ratio of different APD channels was carried from 2 calibrations runs under the condition that the prototype was warmed up and cooled down again • The balance value of ratio of the gains: • Mean  0.99 •   0.02 Gain 1 / Gain 2 QM2006 Daicui Zhou

  13. Energy and Position resolution <(E)/E> ~ 3%,((E)/E) ~ 0.1% @ 2 GeV <x> ~ 2.7mm @ 2 GeV QM2006 Daicui Zhou

  14. Timing resolution measurement with the electron beam. Standard start-stop method with an external trigger Invariant mass spectrum + 12C → 0 + X → 2 + X ~ 0.5 ns at E > 1.5 GeV (0) =4.7 MeV/c2 Timing and Mass Resolution Hysayuki Torii’s poster for ALICE-PHOS(138) QM2006 Daicui Zhou

  15. Installation plan • The first PHOS module will be installed in ALICE for the first pp run in 2007 • The 2nd and 3rd modules will be installed during the first winter shutdown after pp run • The 4th and 5th modules will be installed for the first long heavy-ion run QM2006 Daicui Zhou

  16. ALICE-PHOS project-organization - China, Beijing, China Institute of Atomic Energy - China,Wuhan,Central ChinaNormal University, Institute for Particle Physics - China,Wuhan, Huazhong University of Science and Technology - Czech Republic,Prague, Academy of Science of the Czech Republic, Institute of Physics - Germany,Münster, Westfälische Wilhelms Universität, Institute für Kernphysik - Japan, Hiroshima, Hiroshima University - Norway, Bergen, University of Bergen, Department of Physics - Norway, Oslo, University of Oslo, Department of Physics - Poland, Warsaw, Soltan Institute for Nuclear Studies - Russia, Moscow, Russian Research Center ‘Kurchatov Institute’ - Russia,Protvino, Institute for High Energy Physics - Russia, Sarov, Russian Federal Nuclear Center ‘VNIIEF’ - Russia, Dubna, Joint Institute for Nuclear Research - Switzerland, Geneva, CERN All the PHOS colleagues are acknowledged for the slides Thanks ! QM2006 Daicui Zhou

  17. Importance of the TOF • Different time resolutions of the TOF system were studied to estimate the contamination of the photon spectrum by hadrons. TOF 2ns/E reduces significantly the contamination from hadrons below ~0.1 QM2006 Daicui Zhou