Reference membrane location
Download
1 / 5

Reference Membrane Location - PowerPoint PPT Presentation


  • 55 Views
  • Uploaded on

Reference Membrane Location. Reference Membrane. Span Check. Perform 8 minute count Compare mass value with ABS value ABS must be within 5% or BAM will log CAL error (D error code). CALIBRATION MODE. REF MBRN: <. COUNT (Io):. 634000. COUNT (I):. 556234. CAL MASS M:. 0.801 mg/cm2.

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about 'Reference Membrane Location' - gigi


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript


Span check
Span Check

  • Perform 8 minute count

  • Compare mass value with ABS value

  • ABS must be within 5% or BAM will log CAL error (D error code)


Calibration test screen

CALIBRATION MODE

REF MBRN: <

COUNT (Io):

634000

COUNT (I):

556234

CAL MASS M:

0.801 mg/cm2

START

STOP

EXIT

Calibration Test Screen


Hands on workshop reference membrane
Hands On Workshop – Reference Membrane

  • Remove Reference Membrane

  • Inspect Reference Membrane

  • Reinstall Reference Membrane

  • Inspect mechanical movement of Reference Membrane

  • Perform Reference Membrane Count/Mass Test in TEST/CALIBRATE Screen

  • Compare Mass value with ABS value