JEM-X dead time, grey filter, double triggers, and all that… S. Brandt DNSC, March 2007

JEM-X dead time, grey filter, double triggers, and all that… S. Brandt DNSC, March 2007. Dead time calculation. HW dead time: HW-trigger is disabled for 5.3 m s SW dead time: SW-trigger is disabled to read amplifiers, depending on which selection criteria the event will meet/fail.

gella
Download Presentation

JEM-X dead time, grey filter, double triggers, and all that… S. Brandt DNSC, March 2007

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. JEM-X dead time, grey filter, double triggers, and all that…S. Brandt DNSC, March 2007

  2. Dead time calculation • HW dead time: HW-trigger is disabled for 5.3 ms • SW dead time: SW-trigger is disabled to read amplifiers, depending on which selection criteria the event will meet/fail. • 17 ms for grey filter and ‘buffer full’ • 79 ms for crude rise time and high anode signal • 179 ms for analysis requiring read-in of all amplifiers • Let L denote the fractional live time, L = (1 – deadtime) • Grey filter introduces an additional dead time by allowing a certain transmission T =(1+GF)/32 • To correct a rate for dead time and grey filter: Rcorrected = Rraw / L / T • Correct if all works as expected SDAST Meeting,March 2007 JEM-X grey filter etc. Søren Brandt page 2

  3. Grey filter mechanism and randomness • Grey filter has 32 settings • Transmission, T = (GF+1)/32 • Grey filter is acting “up-front” on HW triggers, based on a table with 32 entries to ensure randomness of accepted events (so they say) • We have 32 tables, each with 32 on/off bits, maximally spread out • T=100%, Table 31: 1111 1111 1111 1111 1111 1111 1111 1111 • T=50%, Table 15: 1010 1010 1010 1010 1010 1010 1010 1010 • T=3%, Table 0: 1000 0000 0000 0000 0000 0000 0000 0000 • So for T=3%, after proccessing an event, we must ALWAYS wait for 31 HW triggers before processing • This has implications for the distribution of times between events, i.e. grey filter mechanism does NOT appear random after all SDAST Meeting,March 2007 JEM-X grey filter etc. Søren Brandt page 3

  4. Probability of processed event following processed event • Entries in 32 bit grey filter are always maximally spread out • For T<50% a processed (accepted) event is ALWAYS following an rejected due to grey filter Truly random JEM-X random SDAST Meeting,March 2007 JEM-X grey filter etc. Søren Brandt page 4

  5. Distribution of delta-times (log plots) GF=31 GF=2 Dead time effects GF=31 GF=2 SDAST Meeting,March 2007 JEM-X grey filter etc. Søren Brandt page 5 Ticks 122 micro-sec Ticks 122 micro-sec

  6. Double triggers • Number of HW-triggers are greater than what is expected from SW-triggers and dead time calculations • Excess could be caused by double triggers (not statistically independent) • SW trigger is disabled at least 79 ms for normally processed events and 17 ms for grey filter events • Double trigger rate is around 20% (going slightly up) • Possible cause: trails of ionization from particles • Double trigger rate goes dramatically up, when the drift voltage (VC) is lowered (slow drift) • Relative double trigger rate is lower, when only 1 anode segment is active (geometry of tracks) SDAST Meeting,March 2007 JEM-X grey filter etc. Søren Brandt page 6

  7. Relative HW trigger rate (%) Double triggers observed in VC test of orbit 262 Nominal setting VC setting HW trigger rate Dead time corrected SW trigger rate (%) SW trigger rate (%) SDAST Meeting,March 2007 JEM-X grey filter etc. Søren Brandt page 7

  8. Relative double trigger rate as function of anode setting All anodes on D=23% Only 1 anode on D=12% SDAST Meeting,March 2007 JEM-X grey filter etc. Søren Brandt page 8

  9. Dead time correction problems when grey filter active • Fewer accepted events than expected are seen when grey filter is active • SW triggering is enabled 17 ms after a grey filter rejected event • In some cases we may then see the second part of a double trigger • Try to estimate the additional correction factor from available data • Better understanding of the double trigger is needed SDAST Meeting,March 2007 JEM-X grey filter etc. Søren Brandt page 9

  10. Additional correction factor • 10% correction for GF<16, less for GF>16 • Proportional to probability of processed event following GF event Relative Crab rate as function of grey filter setting SDAST Meeting,March 2007 JEM-X grey filter etc. Søren Brandt page 10

  11. Upcoming Crab calibration • Grey filter test (10 ks) • map the additional correction factor in detail as function of GF setting • Anode section and grey filter test (8 ks) • Enable various combinations of anode sections for GF=31 and GF 15 • Measure double trigger rate as function of anode section geometry • Drift voltage test (8 ks) • Determine double trigger rate and performance as function of drift voltage • Check gas contamination • Establish calibration with higher drift voltage (4x25 ks, IBIS pointing) • Determine electronic efficiency relative to nominal setting • Check imaging distortion relative to nominal setting SDAST Meeting,March 2007 JEM-X grey filter etc. Søren Brandt page 11

  12. Conclusions • Grey filter rejection not as random as we thought • Possible impact on power density spectra • New deadtime correction needed – 10% effect! • function of grey filter • Function of background rate? • Is the drift velocity going down? • Crab test will show, hopefully • Can we improve performance by increasing drift voltage? • Crab calibration will show SDAST Meeting,March 2007 JEM-X grey filter etc. Søren Brandt page 12