An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Validation on the basis of Commission Decision 2002/657/EC by Dr. Michael Scheutwinkel International Federation forConsulting GmbH IFC

  2. Did you sleep well?

  3. Validation shall demonstrate that the analytical method complies with the criteria applicable for the relevant performance characteristics (= is fit for purpose). Commission Decision 2002/657/EC, validation Banned substances (no permitted limits) MRL substances Screening methods Confirmatory methods

  4. Commission Decision 2002/657/EC Table 10

  5. Commission Decision 2002/657/EC

  6. Commission Decision 2002/657/EC

  7. Commission Decision 2002/657/EC

  8. Commission Decision 2002/657/EC

  9. CRL Guidance Paper 7 December 2007 … state of the art…

  10. CRL Guidance Paper 7 December 2007 MRPLs for MRL substances … state of the art…

  11. CRL Guidance Paper 7 December 2007 MRPLs for banned substances … state of the art…

  12. Commission Decision 2002/657/EC

  13. Commission Decision 2002/657/EC

  14. Commission Decision 2002/657/EC

  15. Commission Decision 2002/657/EC

  16. How to establish CCa …

  17. Commission Decision 2002/657/EC

  18. How to establish CCa … for banned substances In practice … • Calibration curves with very low ao values and low SDwiR are common. That mean CCa gets unreasonable low. In this case CCa is calculated as • CCa = C1 + 2.33 x SDwiR

  19. How to establish CCa … for MRL substances CCa = (fortification at MRL level + 1.64 x SDwiR) Calibration curve: 0, 0.5, 1.0, 2.0 and 2.5-fold MRL We need the within-laboratory reproducibility taken from the three series of checking the internal reproducibility at MRL-level Recovery corrected

  20. Commission Decision 2002/657/EC Three times the signal to noise ratio can be used as decision limit.

  21. Commission Decision 2002/657/EC

  22. For MRL substances CCa = fortification at MRL level + 1.64 x SDwiR CCß = fortification atMRL + 3.28 x SDwiR

  23. Commission Decision 2002/657/EC

  24. Commission Decision 2002/657/EC

  25. Commission Decision 2002/657/EC

  26. Commission Decision 2002/657/EC

  27. How to establish CCß … For banned substances CCß = CCa + 1.64 x SDwiRCCa For MRL substances CCß = MRL + 3.28 x SDwiRMRL

  28. For MRL substances CCa = fortification at MRL level + 1.64 x SDwiR CCß = fortification atMRL + 3.28 x SDwiR

  29. Commission Decision 2002/657/EC

  30. Commission Decision 2002/657/EC

  31. Commission Decision 2002/657/EC

  32. Commission Decision 2002/657/EC

  33. Tools Truneness Specifity Stability Calibration curve • Recovery Repeatability Decision limit CCα Detection capability CCβ Within-laboratory reproducibiity Applicability Reproducibility Ruggedness