E N D
4. Reporting Medical Device Problems Why?
Who?
When?
How?
5. Why Report Medical Device Problems? Health professionals have a moral and ethical obligation to minimize harm to patients, improve their skills and support their hospitals pursuit of patient safety and quality of care
Reporting medical device failures and related adverse effects helps identify and prevent similar events in the future
Reporting allows analysis of cause and focused corrective action
6. Why Report Medical Device Problems? Without reporting and sharing information health professionals, biomedical engineers, materials managers and procurement personnel are handicapped in selecting and purchasing medical products
Without reporting health professionals may not be able to identify the need for additional training of physicians and nurses
Without reporting the Saudi Food and Drug Authority cannot identify deficient products and prevent their import or sale and this increases the probability that hospitals will harm patients, waste time and money
7. Why Report Medical Device Problems? Health professionals and hospitals need to minimize the risk of lawsuits by patients and families and loss of reputation caused by injuries and deaths related to medical devices
Saudi suppliers and manufacturers need organized feedback to improve their judgment about products, processes, components and materials so they can deliver safer and higher quality products to hospitals and device users
8. What Types of Adverse Events Are Caused by Medical Devices? Injuries to patients and hospital personnel
Deaths of patients and hospital personnel
Environmental damage
Lawsuits by patients or their families
New expenses for repair or retraining
Abandonment of a product and loss of the investment
9. The Causes of Medical-Device-Related Incidents Device failure
Device interaction
User error
Maintenance error
Packaging error Tampering
Support system failure
Environmental factor
Idiosyncratic patient reaction
10. Typical Causes of Device Failure Design/labeling error
Manufacturing error
Software deficiency
Random component failure
Power-supply failure
Failure of accessory
11. Mechanisms of Device Related Injury & Death Overdose
Suffocation/barotrauma
Infection
Embolism (gas/particulate)
Skin lesion (puncture/cut/burn Electrocution
Fire
Performance failure
Crushing
Exsanguination
12. Most Frequently Reported Harmful Devices Anesthesia machines
Cardiac interventional catheters
Cardiopulmonary bypass systems
Defibrillators
Dialysis systems
Electrosurgical (surgical diathermy) units
Disposable surgical trocars
IV pumps
Surgical staplers
Ventilators
13. Observation Of these top 10 harmful devices, half (5) are used only in the operating theatre.
3 additional ones may be used in the operating theatre.
3 are used in almost all areas of the hospital.
The most frequently reported problem device is the infusion pump.
3 of the 10 devices are highly dependent on physician technique with very simple devices (i.e., interventional catheters, trocars, and staplers)
14. Conclusion 7 of these 10 devices require main voltage electrical energy.
But in 6 of these 7 devices, almost all reported problems are mechanical.
One device, the defibrillator, has frequently reported battery- and power-supply problems (i.e., too little rather than too much electricity).
Focus attention on mechanical problems because the electrical safety issues are highly exaggerated.
Maintenance error or failure to inspect is rarely a cause of harm.
Focus attention on training because most harm involving devices is caused by operator error.
15. What Inhibits Reporting of Medical Device Adverse Events? Perception of the event
Information-sharing culture
Fear of authoritarian superiors
Fear of blame and punishment
Lack of a general incident reporting system, (critical to risk management)
Failure to investigate the event and incorporate what was learned into training, revised clinical procedures and more effective selection and procurement of medical devices
16. Educate Device Users Have them:
Explore www.mdsr.ecri.org
Read case histories of deaths and injuries caused by devices they use in their specialty and why they occurred
Download and publicize safety posters for specific devices
17. Who Should Report? It depends on the incident reporting system and reporting channels in your hospital
In most Western hospitals reporting is done by the nursing staff through the incident reporting system and the incident comes to the attention of the risk management department and biomedical engineering department
Typically the risk manger or biomedical engineer reports the incident to the device regulatory agency
18. When? As quickly as possible before memories fade
While the scene of the incident and evidence is preserved so the risk manager and biomedical engineer can examine the device, take notes and if appropriate take photographs and try to determine the cause
19. Protect the Patient & Staff, But Preserve the Evidence! The first priority is to prevent further harm to patient or staff
The next highest priority is to protect the physical environment e.g. from fire
The last critical priority is to preserve the evidence so cause can be determined
20. Preserve the Evidence for Analysis Do not move equipment, or accessories unless needed to project patient or staff from harm
Treat the location of the event as if it were a crime scene
Do not changes control settings on any equipment
Do not detach or dispose of any accessories or consumables or single use products such as cables, catheters, electrodes. tubes, humidifiers, etc. Their presence, juxtaposition and connections may prove critical in understanding the event
Make sketches or take photographs as appropriate
Document who was there, who did what, what happened. etc.
21. Things Not To Do Never, ever send or release implicated devices to suppliers or manufacturers until the analysis is complete and cause is determined. Once the device is out of your control it can be altered or be lost
Do not assume that suppliers are on your side. If investigation requires the help of the supplier be sure it is done in the hospital with your risk manager, biomedical engineer and involved health professionals and, if you wish, a representative of the SFDA present
Do not assume, once you have completed your investigation, that nothing else need be done. If litigation is possible lock the device and its accessories in an area with controlled access and preserve it and related documentation and photographs as if it will undergo additional analysis and be introduced as evidence in court
22. Things to Do Prevent adverse events by carefully selecting and purchasing medical devices. You now have immediate access to the most comprehensive up-to-date information in the world on the quality, safety and cost-effectiveness of such products via SFDA and ECRI
When choosing equipment give special attention to ergonomics and human factors design (aviation examples)
Emphasize training and retraining
Share information openly
23. How? We will now demonstrate:
How to report a medical device event to the Saudi Food & Drug Authority
How to search SFDA-ECRI supplied databases for adverse event information
How to use SFDA-ECRI supplied databases to improve selection and procurement of medical devices