3730 自動定序儀 每週一清洗步驟流程

3730 自動定序儀 每週一清洗步驟流程. 依照上圖指示 先 Water wash 冷水共 3 次 後 Water wash 溫水共 3 次 總共 water wash 6 次 PS: 溫水水溫約 38℃ 使用無菌燒杯取 Milli-Q water 約 80ml 微波加熱後隔水降溫 , 取溫度計測量溫度後才可使用. Flush. 待以上步驟完成後 即可將舊的廢液盤更換 包括 :buffer.water.waste 廢液盤 PS: 記得 polymer.1X running buffer 皆需要先回溫 30 分鐘後才可更換喔 !.

chacha
Download Presentation

3730 自動定序儀 每週一清洗步驟流程

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. 3730自動定序儀 每週一清洗步驟流程

  2. 依照上圖指示先Water wash冷水共3次後Water wash溫水共3次總共water wash 6次PS:溫水水溫約38℃使用無菌燒杯取Milli-Q water約80ml微波加熱後隔水降溫,取溫度計測量溫度後才可使用

  3. Flush

  4. 待以上步驟完成後即可將舊的廢液盤更換包括:buffer.water.waste 廢液盤PS:記得polymer.1X running buffer皆需要先回溫30分鐘後才可更換喔!

  5. 每個月月初定期清洗polymer block syrring1.取無菌空針,抽取20ml Milli-Q water2.準備ㄧ個乾淨的空瓶裝廢液3.syrring fitting and waste fitting稍微轉開1/2~1/34.從Syrring fitting處,將water push 進去(PS:速度1秒1滴,不可過快)

  6. 說明完3730機本清洗步驟後簡單舉例來討論,當發生下列異常時正確處理流程之步驟為何及如何找出問題點?~~動腦時間到囉~~說明完3730機本清洗步驟後簡單舉例來討論,當發生下列異常時正確處理流程之步驟為何及如何找出問題點?~~動腦時間到囉~~

  7. 狀況ㄧ~ 此array view約後2/3處皆有黃色之backgroung

  8. 看過上圖後,正確之處理步驟為何? 需觀察哪些地方來協助找出問題點呢!

  9. 檢查及處理步驟~ 1.首先將觀察機台狀況 *3730及軟體是否有亮紅燈或Error顯示 *機台溫度.室內溫度.EPT current.Laser.CCD Tm.有無漏膠 *此份Array前後之毛細管狀況是否正常 2.到Run history處將此份樣品相關畫面用小畫家照相存檔下來 Ex:Array view/EPT view/Instrument view/Event and Log 3.確認樣品來源及客戶別 4.確認反應人員及沉澱上機人員操作有無問題 5.確認近期內有無更換反應/沉澱/上機之試劑 6.開啟分析軟體觀察定序報告是否有問題

  10. 7.區別好問題點為樣品本身或是機台-毛細管問題7.區別好問題點為樣品本身或是機台-毛細管問題 8.處理問題—重新run樣品測試或是Run HiDi確定毛細管狀態 9.並於上機本內以紅筆記錄異常狀況及處理流程 慢慢變厲害囉 但是還有ㄧ題喔… …….嘿嘿

  11. 案例二~3730機台及collection softward軟體亮紅燈 並顯示Error:Leak detected during capillary filling/Run failured

  12. 請大家動動腦,說出正確處理步驟囉…………………..88請大家動動腦,說出正確處理步驟囉…………………..88 別緊張,提示來囉…………..

  13. 樣品正在分析中的instrument view~ EP Voltage:8.0~8.5 KV EP Current:180~200 uA Laser Power:25~25.5 mW Laser Current:5.0~5.5 A Buffer Heater:30~35 ℃ Cell Heater:48.5~49 ℃ Oven Tm:60 ℃

  14. 正常之EPT vuew~ EPT Current Oven Tm Ovent Peiltier Tm Buffer Tm Cell Heater EPT Voltage CCD Tm Laser Power Ambient Tm Lser Current

  15. EPT current代表之意義~

  16. Event Log記錄~

  17. Polymer Upper Tubing Buffer Jar

  18. 想到答案了嗎? 你是哪一個呢………………加油喔