1 / 28

Interoperabilität von patientennahen Medizingeräten

Interoperabilität von patientennahen Medizingeräten . openSDC Update. Joint meeting of IEEE EMBS 11073 & HL7 Health Care Devices (DEV) WG , 2014/05/27, Stefan Schlichting. openSDC Update Agenda. openSDC Update Convergence Project Standardization Roadmap. openSDC update History.

beata
Download Presentation

Interoperabilität von patientennahen Medizingeräten

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Interoperabilität von patientennahen Medizingeräten openSDC Update Joint meeting of IEEE EMBS 11073 & HL7 Health Care Devices (DEV) WG, 2014/05/27, Stefan Schlichting

  2. openSDC Update Agenda openSDC Update Convergence Project Standardization Roadmap An architecture for distributed systems of medical devices in high acuity environments

  3. openSDC updateHistory • Last WG Meeting we proposed • “an architecture for distributed systems of medical devices in high acuity environments” for standards adoption. An architecture for distributed systems of medical devices in high acuity environments

  4. openSDC updateWhat is openSDC? • Extensible openSDC protocol stack for clinical workplace service-oriented medical device architectures • BICEPS (Message & Service Model) • MDPWS & DPWS (Transport) An architecture for distributed systems of medical devices in high acuity environments

  5. openSDC updateopenSDC Reference Implementation • openSDC downloads since January WG Meeting • New Version beta03 will be out end of Q2/2014 • See http://sourceforge.net/projects/opensdc/ An architecture for distributed systems of medical devices in high acuity environments

  6. openSDC updateBICEPS changes since last WG Meeting • Update of the BICEPS model V2 • Get closer to the classic DIM • Incorporate change request from OR.NET • Bug fixes The BICEPS model V3 will be published in openSDC beta03. Expect a BICEPS V3.1 in Q3/2014 (openSDC beta04) • Optimized representation of the transmitted messages An architecture for distributed systems of medical devices in high acuity environments

  7. openSDC Update Agenda openSDC Update Convergence Project Standardization Roadmap An architecture for distributed systems of medical devices in high acuity environments

  8. Convergence Project • Summary of the convergence project An architecture for distributed systems of medical devices in high acuity environments

  9. Convergence ProjectUse Cases An architecture for distributed systems of medical devices in high acuity environments

  10. Convergence ProjectWhat is Common? • Discussed • Discovery • Security: IPSecvs TLS/DTLS • QoS for Transport: IntServvsDiffServ • Example for Discovery • openSDC may use DNS, but only with the focus of name resolution. An architecture for distributed systems of medical devices in high acuity environments

  11. Convergence ProjectHarmonization potential • Convergence vs. Coexistence? An architecture for distributed systems of medical devices in high acuity environments

  12. Convergence ProjectCoexistence - Multi-protocol Gateway model • Medical Device Standards-based Clinical “Ecosystem” (aka Clinical Gateway) integrating an IEEE 11073 Mgr-Agt pair with a BICEPS-based Proxy-Supervisor pair. Common MDIB An architecture for distributed systems of medical devices in high acuity environments

  13. Convergence ProjectCoexistence - Multi-protocol Gateway model • BICEPS Clinical Gateway part Common MDIB BUS-API BICEPS Supervisor Proxy BICEPS Supervisor An architecture for distributed systems of medical devices in high acuity environments

  14. Convergence ProjectCommon MDIB - Alert models • 11073 Alert Package model BICEPS model An architecture for distributed systems of medical devices in high acuity environments

  15. AlertSystem • Detects AlertConditionsand - as appropriate - generates AlertSignals. • ALARM SYSTEM (From 60601-1-8) • parts of ME EQUIPMENT or a ME SYSTEM that detect ALARM CONDITIONS and, as appropriate, generate ALARM SIGNALS An architecture for distributed systems of medical devices in high acuity environments

  16. Convergence ProjectCommon MDIB – Alert Models • AlertCondition • Contains the information about a potentially or actually hazardous situation.Example: An AlertLimit has been exceeded, or a Sensor has been unplugged. • LimitAlertCondition is derived from AlertCondition • ALARM CONDITION(From 60601-1-8) • state of the ALARM SYSTEM when it has determined that a potential or actual HAZARD exists An architecture for distributed systems of medical devices in high acuity environments

  17. Convergence ProjectCommon MDIB – Alert Models • AlertSignal • Contains information about the way an AlertCondition is communicated to someone.It is generated by an AlertSystem to indicate the presence or occurence of an AlertCondition. • ALARM SIGNAL(From 60601-1-8) • type of signal generated by the ALARM SYSTEM to indicate the presence (or occurrence) of an ALARM CONDITION An architecture for distributed systems of medical devices in high acuity environments

  18. Convergence ProjectCommon MDIB – Alert Models • AlertSignal Example • AlertSignal could be a lamp (-> Visible Manifestation) on a remote device, suchas the nurses handheld (-> SignalDelegationSupported), which starts flashingwhen the heart rate is exceeding 150bmp (-> ConditionSignaled) for more than2 seconds (-> DefaultSignalGenerationDelay) and keeps flashing, until thenurse confirms the alarm, even if the AlarmCondition is not present anymore(-> Latching). An architecture for distributed systems of medical devices in high acuity environments

  19. Convergence ProjectCommon MDIB – Alert Models • AlertStatesfor • AlertSystem • AlertCondition • AlertSignal • Alert Processing • In order to retrieve the overall alert state of an MDS, a client has to subscribe to all AlertSystems. • An AlertCondition has a rank that represents the overall rank of the Alarm Condition in the MDS. • An AlertSignal has a slot that represents the overall ordering number of the signal in the MDS. • In order to retrieve the overall alert condition state of an MDS, a client has to subscribe to all AlertConditions. • In order to retrieve the overall alert signal state of an MDS including remote signals, a client has to subscribe to all AlertSignals. An architecture for distributed systems of medical devices in high acuity environments

  20. Convergence ProjectSummary • Stefan’s View • Clinical Gateway seems feasible incl. a common MDIB. • Only limited value for a real “converged” network specification • NTP is essential even for a Clinical Gateway • Security Infrastructure including role model should be harmonized • Common Requirements Glossary • + recommendations for Transport QoS, Security,.. based on UC More Info • http://mdcx73-working.wikispaces.com/MDCx73_UseCases-UCs An architecture for distributed systems of medical devices in high acuity environments

  21. openSDC Update Agenda openSDC Update Convergence Project Standardization Roadmap An architecture for distributed systems of medical devices in high acuity environments

  22. Standardization RoadmapOR.NET project OR.NET Partners http://www.ornet.org/ An architecture for distributed systems of medical devices in high acuity environments

  23. Standardization RoadmapOR.NET project Source: http://ws4d.e-technik.uni-rostock.de/2014/or-net-and-ws4d-org-at-cebit-2014/ An architecture for distributed systems of medical devices in high acuity environments

  24. Standardization RoadmapopenSDC & OR.NET • Summary of a telco in April • The OR.NET standardization project & Dräger closely work together to publish a technical specification for a service-oriented architecture implementation for a distributed system of medical devices in the OR and related areas. • The technical specification for medical device communication should be published under the umbrella of IEEE 11073. • The technical specification is based on “openSDC” & OR.NET work An architecture for distributed systems of medical devices in high acuity environments

  25. Standardization RoadmapOR.NET project – Protocol View Medical Devices IT (EHR, PDMS,…) Endos. Cams OR Workstation Light srcs Device Connectivity Insufflator OR table HF dissector Tracking Navigation Media Ctrl Device 2 Enterprise (HL7, DICOM,…) Imaging According to J. Dehm (OR.NET – Standardization)

  26. Standardization Roadmap MDPWS + BICEPS Spec 2012 MDPWS + BICEPS Spec 2014 • openSDC/OR.NET • Technical Specification • Nomenclature Additions • Device Specializations • Clinical Gateway Spec • ICE Implementation Guide An architecture for distributed systems of medical devices in high acuity environments

  27. Contact Dr. Stefan Schlichting Research Unit, Drägerwerk AG stefan.schlichting@draeger.com Thank you for your attention.

  28. Use case discussion with MDPnP • Follow-up discussion on the HIMSS 2014 An architecture for distributed systems of medical devices in high acuity environments

More Related