An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. MARCAPASOS Y CDI IMPLICACIONES ANESTÉSICAS Enrique Anaya Residente Anestesia CES

  2. HISTORIA • Primer marcapasos implantado el 8 de Octubre de 1958 • 1960-1970: menor tamaño, batería de Litio, implantación venosa. • 1962: Desfibrilación externa como terapia para taquiarritmia • 1980: Primera demostración exitosa de Desfibrilación en un humano con ICD – Derivadas liberadoras de esteroides • 1985: FDA aprueba ICD para uso práctica médica Assessment and management of patients with pacemakers and implantable cardioverter defibrillators Crit Care Med 2004 Vol. 32, No. 4

  3. EPIDEMIOLOGÍA • 3´000.000 personas tienen Marcapasos. • Mas de 300.000 CDI The patient with a cardiac pacemaker or implanted defibrillator and management during anaesthesia Current Opinion in Anaesthesiology 2007, 20:261–268

  4. EPIDEMIOLOGÍA USA UK • 720.000 MC • 500.000 CDI • 115.000 nuevos dispositivos por año • 239.000 MC • 2.000 CDI • 30.000 nuevos dispositivos cada año Pacemakers and implantable cardioverter defibrillators Anaesthesia, 2006, 61, pages 883–890

  5. COMPOSICIÓN MARCAPASOS • Generador de pulso • Batería de Litio • Chip silicona • Circuito electrónico • Sistema de derivadas • Permite paso de corriente desde el generador hasta las cavidades cardíacas Assessment and management of patients with pacemakers and implantable cardioverter defibrillators Crit Care Med 2004 Vol. 32, No. 4

  6. POLARIDAD • Unipolares • Punta de electrodo actúa como cátodo • Superficie de generador de pulso: ánodo • Bipolares • Ambos electrodos están dentro del Corazón • Ánodo 1-2 cms proximal a cátodo • Menor probabilidad de interferencia • Single pass Assessment and management of patients with pacemakers and implantable cardioverter defibrillators Crit Care Med 2004 Vol. 32, No. 4

  7. MODOS DE ESTIMULACIÓN • Código NBG desde 1987 • Cinco posiciones universalmente aceptadas • En la actualidad 3 primeras mas comúnmente encontradas Cardiac Rhythm Management Devices (Part I) Anesthesiology 2001; 95:1265–80

  8. NOMENCLATURA The revised NASPE ⁄ BPEG generic code for antibradycardia, adaptiverate, and multisite pacing. Pacing and Clinical Electrophysiology 2002; 25: 260–4.

  9. MODOS DE PROGRAMACIÓN (1998) • VVI más utilizado • AAI: De elección en pacientes con disfunción sinusal y nodo AV intacto • VDD: Ideal en casos de nodo sinusal sano con Bloqueo AV • DDD: ideal Cardiac Rhythm Management Devices (Part I) Anesthesiology 2001; 95:1265–80

  10. MODOS DE PROGRAMACIÓN (2006) • VVI (16.9%) • VVIR (24.8%) • DDD (27.3%) • DDDR (25.4%) Pacemakers and implantable cardioverter defibrillators. Anaesthesia, 2006, 61, pages 883–890

  11. VVI

  12. DDD

  13. MODOS DE PROGRAMACIÓN • Rate Adaptative: Elimina interferencia de movimiento y la vibración y puede censar la ventilación minuto, temperatura, SVO2, para ajustar la FC • VVIR: Fibrilación Auricular crónica con lenta respuesta ventricular • AAIR: Disfunción sinusal pura • DDDR: Alteración NS y AV concomitantes Cardiac Rhythm Management Devices (Part I) Anesthesiology 2001; 95:1265–80

  14. MODO ASINCRÓNICO • AOO – VOO – DOO • NO hay sensibilidad • Respuesta es fija a frecuencia deseada • De elección en pacientes sometidos a procedimientos quirúrgicos Assessment and management of patients with pacemakers and implantable cardioverter defibrillators Crit Care Med 2004 Vol. 32, No. 4

  15. TERAPIA DE RESINCRONIZACIÓN CARDIACA

  16. MARCAPASOS BICAMERALES • Ventajas • Aumentan GC y PAM • Preservación flujo coronario • Disminuye síntomas de Síndrome Marcapasos • Menor incidencia de ICC y dilatación auricular • Disminuyen aparición de Fibrilación Auricular (menos remodelamiento) The patient with a cardiac pacemaker or implanted defibrillator and management during anaesthesia Current Opinion in Anaesthesiology 2007, 20:261–268

  17. MARCAPASOS BIVENTRICULARES • SOLO en pacientes con Falla cardiaca congestiva • Fracción de Eyección baja • QRS prolongado • No mejoría pese a manejo médico óptimo The patient with a cardiac pacemaker or implanted defibrillator and management during anaesthesia Current Opinion in Anaesthesiology 2007, 20:261–268

  18. SINDROME MARCAPASOS • Descrito desde 1969 • Efectos hemodinámicos adversos con marcapasos NORMAL que generan limitación para realizar actividades de rutina • AsincroníaAuriculo-ventricular • CONGESTIVOS: • Semejan falla cardiaca, disnea, ortopnea, edemas, estertores pulmonares y hepatomegalia. • HIPOTENSIVOS • Presíncope, síncope y disminución de las cifras tensionales sistólicas en un 20%. • INESPECÍFICOS: • Dolor precordial, cefalea y fatiga Pacemakers and defibrillators: anaesthetic implications British Journal of Anaesthesia 93 (1): 95±104 (2004

  19. RESPUESTA A MAGNETO • Varía de acuerdo a marca de dispositivo • Unicamerales entran en pacing asincrónico • Bicamerales usualmente entran en asincrónico DDO • NUNCA se apaga Marcapasos Cardiac Rhythm Management Devices (Part II) Anesthesiology 2001; 95:1492–1506

  20. CDI • Generador • Cubierta de Titanio • Circuitos • Microprocesador • Baterías • Capacitor • Sistema de Derivadas • ATP Assessment and management of patients with pacemakers and implantable cardioverter defibrillators Crit Care Med 2004 Vol. 32, No. 4

  21. CDI Pacemakers and implantable cardioverter defibrillators Anaesthesia, 2006, 61, pages 883–890

  22. CDI Funcionamiento • Programación para suministrar energía de acuerdo a ritmo • Modos de pacing similares a los marcapasos modernos Assessment and management of patients with pacemakers and implantable cardioverter defibrillators Crit Care Med 2004 Vol. 32, No. 4

  23. RESPUESTA A MAGNETO • Se impide liberación de descarga a pacientes • Tiempo mínimo de 20 a 30 segundos • Cambio de tono de timbres emitidos • NO se altera la función de marcapasos de estos dispositivos Cardiac Rhythm Management Devices (Part II) Anesthesiology 2001; 95:1492–1506

  24. INDICACIONES MARCAPASOSAHA/ACC • Alteraciones Nodo Sinusal • Disfunción del Nodo sinusal con bradicardia sintomática • Incompetencia cronotrópica sintomática • Bradicardia por medicamentos cuando no hay otra alternativa efectiva (?) The patient with a cardiac pacemaker or implanted defibrillator and management during anaesthesia Current Opinion in Anaesthesiology 2007, 20:261–268

  25. INDICACIONES MARCAPASOSAHA/ACC • Bloqueo AV Adquirido • Bloqueo AV de III grado asociado a: • Bradicardia sintomática • Bradicardia por medicamentos • Asistolia ≥ 3 segundos • Ritmo de escape de menos de 40 Ipm • Enfermedades neuromusculares • Bloqueos postoperatorios • Bloqueo de II grado con bradicardia sintomática The patient with a cardiac pacemaker or implanted defibrillator and management during anaesthesia Current Opinion in Anaesthesiology 2007, 20:261–268

  26. INDICACIONES MARCAPASOSAHA/ACC • Bloqueo Bifascicular y Trifascicular • Bloqueo AV completo intermitente • Bloqueo AV II grado tipo 2 • Síndrome Q-T largo • Cardiomiopatía hipertrófica y dilatada. The patient with a cardiac pacemaker or implanted defibrillator and management during anaesthesia Current Opinion in Anaesthesiology 2007, 20:261–268

  27. INDICACIONES CDIAHA/ACC • Pacientes que han sobrevivido episodio de taquiarritmia ventricular espontánea cuya causa no es reversible • Cardiomiopatía con Fracción de Eyección < 35% • Post-IAM con FE < 30% y QRS > 120 mseg • FV sostenida espontánea • Síndrome de Brugada. • Terapia puente para trasplante cardiaco The patient with a cardiac pacemaker or implanted defibrillator and management during anaesthesia Current Opinion in Anaesthesiology 2007, 20:261–268

  28. INTERFERENCIA

  29. QUÉ NO DA INTERFERENCIA • Radios de banda civil • Taladradoras eléctricas • Mantas eléctricas • Afeitadoras eléctricas • Radios de aficionados • Almohadillas eléctricas • Equipos de seguridad aeroportuaria • Detectores de metales • Hornos de microondas • Transmisores de televisión • Controles remotos de televisión • Aparatos de rayos X CardiacRhythm Management Devices (Part II) Anesthesiology 2001; 95:1492–1506

  30. QUÉ SI DA INTERFERENCIA • Grupos electrógenos • Equipos para soldar • Algunos aparatos utilizados por los dentistas • Máquinas de resonancia magnética • Máquinas de radioterapia para el tratamiento del cáncer • Equipos pesados o motores con imanes potentes CardiacRhythm Management Devices (Part II) Anesthesiology 2001; 95:1492–1506

  31. INTERFERENCIA HOSPITALARIA • Electrocauterio • Monitoria • Movimiento • Potenciales de Acción Muscular • Medicamentos • Estimulador de Nervio periférico Cardiac Rhythm Management Devices (Part II) Anesthesiology 2001; 95:1492–1506

  32. INTERFERENCIA Cardiac Rhythm Management Devices (Part II) Anesthesiology 2001; 95:1492–1506

  33. EFECTOS DE INTERFERENCIA • Respuestas Potenciales • Inhibición • Disparo • “Reset” a modo de fabricación • Modo asincrónico • Disparo innecesario de choques • Daño físico del dispositivo The patient with a cardiac pacemaker or implanted defibrillator and management during anaesthesia Current Opinion in Anaesthesiology 2007, 20:261–268

  34. EFECTOS DE INTERFERENCIA • Consecuencias Clínicas: • Bradicardia • Asistolia • Taquiarritmias • Disminución de Gasto cardíaco • Daño endocárdico The patient with a cardiac pacemaker or implanted defibrillator and management during anaesthesia Current Opinion in Anaesthesiology 2007, 20:261–268

  35. MANEJO PERIOPERATORIO

  36. CONSULTA PREANESTESICA • Identificar pacientes con marcapasos/ICD • Estimar magnitud de morbilidad cardiológica • Determinar dependencia de Marcapasos • Examen físico adecuado • Marca y Modelo de Dispositivo Pacemakers and defibrillators: anaesthetic implications British Journal of Anaesthesia 93 (1): 95±104 (2004

  37. CONSULTA PREANESTESICA • Electro fisiólogo tratante • Indagar respuesta a Magneto • Determinar si existe alta probabilidad de Interferencia electromagnética • Saber si se cuenta con equipo necesario para situaciones emergentes Pacemakers and defibrillators: anaesthetic implications British Journal of Anaesthesia 93 (1): 95±104 (2004

  38. IDENTIFICACIÓN

  39. IDENTIFICACIÓN

  40. IDENTIFICACIÓN

  41. CONDUCTA PREANESTESICA • Reprogramar a modo Asincrónico • Disminución de Sensibilidad • Instauración de Frecuencias mayores que actividad intrínseca • Desactivar pacingsensible a frecuencia (movimiento muscular, movimientos respiratorios) • Desactivar ATP Pacemakers and defibrillators: anaesthetic implications British Journal of Anaesthesia 93 (1): 95±104 (2004

  42. MANEJO INTRAOPERATORIO • Placa de electrocauterio lo mas cerca de sitio operatorio y lo mas lejos de MP/ICD • Verificar que la placa haga contacto pleno con piel de paciente • Monitoría básica (Indicador de pulso) • Bisturí Frío, Bipolar o Armónico • Planes de contingencia disponibles Pacemakers and defibrillators: anaesthetic implications British Journal of Anaesthesia 93 (1): 95±104 (2004

  43. MANEJO INTRAOPERATORIO • Evitar acidosis, alcalosis, Hiperkalemia, Hipoxemia, isquemia miocárdica, Hiperglicemia severa. • Anti arrítmicos. • Succinilcolina. • NINGÚN agente anestésico genera interferencia directa con estos dispositivos. Pacemakers and defibrillators: anaesthetic implications British Journal of Anaesthesia 93 (1): 95±104 (2004

  44. MANEJO POSTOPERTARIO • Monitoria estricta • Equipo de desfibrilación disponible • Remitir a electrofisiología para verificar persistencia de modo de programación necesario Pacemakers and defibrillators: anaesthetic implications British Journal of Anaesthesia 93 (1): 95±104 (2004

  45. PACIENTE URGENTE • Identificar tipo dispositivo • Determinar si dispositivo funciona • Determinar si EMI genera inestabilidad hemodinámica • Magneto • Remisión en postoperatorio inmediato a Electrofisiología Pacemakers and defibrillators: anaesthetic implications British Journal of Anaesthesia 93 (1): 95±104 (2004