February, 2012 Calibration issues and answers Service Department GfG Instrumentation - PowerPoint PPT Presentation


  • 86 Views
  • Uploaded on

February, 2012 Calibration issues and answers Service Department GfG Instrumentation 1194 Oak Valley Drive, Suite 20 Ann Arbor, MI 48108 Ph: +1 734 769 0573 Internet: www.gfg-inc.com. Calibration. Verify accuracy on a regular basis to guard against any unexpected loss of sensitivity

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about ' February, 2012 Calibration issues and answers Service Department GfG Instrumentation' - abra-weaver


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
February 2012 calibration issues and answers service department gfg instrumentation

  • February, 2012

    • Calibration issues and answers

    • Service Department

    • GfG Instrumentation

    • 1194 Oak Valley Drive, Suite 20

    • Ann Arbor, MI 48108

    • Ph: +1 734 769 0573

    • Internet: www.gfg-inc.com


Calibration
Calibration

  • Verify accuracy on a regular basis to guard against any unexpected loss of sensitivity

  • Document!


Mandatory to use a calibrated instrument
Mandatory to use a "calibrated" instrument

  • USA OSHA 1910.146(c)(5)(ii)(C) “Permit required confined spaces” requires:

    • Before an employee enters the space, the internal atmosphere shall be tested, with a calibrated direct-reading instrument

    • What does OSHA accept as a "calibrated" direct reading instrument?

      • A testing instrument maintained and calibrated in accordance with the manufacturer's recommendations

      • The best way for an employer to verify calibration is through documentation


Why do instruments need to be tested and or calibrated
Why do instruments need to be tested and / or calibrated?

  • The response of gas detecting sensors can change over the life of the sensor

  • The changes may be sudden, or can be gradual

  • Substances or conditions present in the atmosphere can have an adverse effect on the sensors

  • Different types of sensors have different constraints and conditions which can lead to loss of sensitivity or failure

  • Important to know how sensors detect gas to understand conditions that can lead to inaccurate readings


Make sure the instrument has been calibrated
Make sure the instrument has been calibrated!

  • Follow manufacturer recommendations

  • “Zero” instrument in fresh air prior to use

  • Verify Accuracy Daily!

  • Functional “bump” test sufficient

  • Adjust “span” only if necessary


Loss of sensitivity can be due to
Loss of sensitivity can be due to:

  • Aging or desiccation of the sensors,

  • Mechanical damage due to dropping or immersion

  • Exposure to sensor poisons present in the atmosphere being monitored

  • Loss of sensitivity due to other causes


Regulatory requirements
Regulatory Requirements

  • OSHA 1910.146 requires use of a “calibrated” instrument

  • This means (per OSHA CPL 2.100) that the instrument must be maintained and calibrated according to manufacturer guidelines


Manufacturer guidelines
Manufacturer Guidelines

  • OSHA holds instrument users accountable to maintain, calibrate and operate their instruments according to manufacturer guidelines


Additional canadian requirements

00000000000

Additional Canadian Requirements

  • Canadian Standard CSA 22.2 requires mandatory “bump test” before each day’s use for all instruments equipped with a sensor for the measurement of combustible gas

  • Operation manuals of instruments Certified for sale in Canada carry this warning

Small “c” indicates the certification includes conformity with Canadian requirements


Calibration frequency
Calibration Frequency

  • The safest course of action is to expose the sensors to known concentration test gas before each day’s use!

  • This test is very simple and takes only a few seconds to accomplish


Functional bump test vs calibration
Functional “Bump” Test vs. Calibration

  • Functional “bump” test only provides verification of sensor performance

  • Calibration includes adjustment

  • Only necessary to adjust sensor sensitivity if readings are off

  • Most manufacturers recommend adjustment if readings are off by more than 10% of expected values


Industrial safety equipment association isea guidelines
Industrial Safety Equipment Association (ISEA) Guidelines

  • Provides procedure for lengthening the interval between calibration checks

    • During period of initial use of at least 10 days in the intended atmosphere, check the response daily to be sure there is nothing in the atmosphere which is poisoning the sensor(s).

    • Period of initial use must be of sufficient duration to ensure that the sensors are exposed to all conditions which might have an adverse effect on the sensors.


Industrial safety equipment association isea guidelines1
Industrial Safety Equipment Association (ISEA) Guidelines

  • If these tests demonstrate that it is not necessary to make adjustments, the time between checks may be lengthened

  • This interval should not be lengthened beyond thirty days

  • History of the instrument should be tracked or logged


Industrial safety equipment association isea guidelines2
Industrial Safety Equipment Association (ISEA) guidelines

  • Any conditions, incidents, or exposure to contaminants which might have an adverse effect on the sensors should trigger immediate re-verification before further use

  • Any changes in the work being done, or environment in which the instrument is being used should trigger re-verification by means of daily checking that it is safe to lengthen the interval between calibration checks


Conditions which should trigger immediate re verification of calibration
Conditions which should trigger immediate re-verification of calibration

  • If there is any doubt at any time as to the accuracy of the sensors, verify the calibration of the sensors by exposing them to known concentration test gas before further use!


Don t be afraid of calibration
Don’t be afraid of calibration! calibration

  • Modern designs make calibration easy and automatic

  • Keep the Calibration Materials With the Instrument!

  • All-In-One Calibration Mixtures Make Functional Testing Easy!


Is the concentration of gas used to calibrate instruments dangerous
Is the concentration of gas used to calibrate instruments dangerous?

Not at all!

Consider the concentration if you leaked an entire cylinder of CO cal gas into the interior space of a typical passenger van

Approximate interior volume of Honda Odyssey EX-L = 300 cubic feet

Typical cal gas cylinder holds 34 liters = 1.2 cubic feet @ 50 ppm CO

1.2 cubic feet = 0.4% of the volume of the entire vehicle

50 ppm X .004 = concentration in vehicle = 0.2 ppm CO


What about a cylinder that contains 25 ppm h 2 s
What about a cylinder that contains 25 ppm H dangerous?2S? 

34 liters = 1.2 cubic feet @ 25 ppm H2S

10 X 10 X 10 room = 1,000 cubic feet

1.2 cubic feet = 0.12% of the volume of the entire room

25 ppm X .0012 = concentration in room = 0.03 ppm H2S

10’

10’


Record keeping
Record Keeping dangerous?

  • Documentation is critical!

  • Without good records you cannot defend or explain your procedures

  • If you don’t have the records to prove it was being done right -- it wasn’t!


Atmospheric hazards are frequently invisible to human senses
Atmospheric hazards are frequently invisible to human senses dangerous?

  • You don’t know whether it’s safe until it’s been tested!


Bump test manual procedure
Bump Test (Manual Procedure) dangerous?

  • Make sure the instrument is located in fresh air, turn on, and allow to warm up

  • Note readings, and perform fresh air zero if necessary

  • Attach calibration adapter; instrument will display "AutoCal menu"

  • Press "Exit" to show normal gas reading screen

  • Flow gas to sensors; alarms should activate, and readings should stabilize at expected values (if sensors fail to respond properly, instrument should be calibrated before further use)

  • Turn off gas, remove cal adapter and allow readings to stabilize at fresh air values


Bump test
Bump Test dangerous?

  • Response of sensors to Quad Mix (graphs)

  • Readings recorded while instrument operated in normal gas reading mode

  • Simultaneous response to all four gases, as well as match between cal gas concentrations and readings


Bump test1
Bump Test dangerous?

  • Response of sensors to Quad Mix (table)

  • Readings recorded while instrument operated in normal gas reading mode

  • When a sensor is in alarm readings are recorded in red


Autocal
AutoCal dangerous?

  • AutoCal allows instrument to be “Fresh air” or “Calibration” (span) adjusted if needed


Attach cal cap to enter fresh air and span autocal

Attach cal cap to unit; dangerous?

instrument automatically enters “AutoCal “ menu

Choose ZERO or CAL and apply gas (if calibrating), e.g. an H2S/CO mix.

Attach Cal Cap to Enter Fresh Air and Span “AutoCal”

Adjustment is automatic

Display shows when cal adjust has been successfully completed


Can also enter autocal mode by pushing reset and zoom buttons at same time

Push “Reset” and “Zoom” at same time; instrument automatically enters same “AutoCal “ menu

Choose ZERO or CAL and apply gas (if calibrating), e.g. an H2S/CO mix.

Can also enter “AutoCal” mode by pushing “Reset” and “Zoom” buttons at same time

Adjustment is automatic

Display shows when cal adjust has been successfully completed

Push at same time to enter “AutoCal” Mode


Calibration gas concentrations
Calibration gas concentrations automatically enters same “AutoCal “ menu

  • Best to use the default cal gas concentrations

  • Default GfG concentrations used to calibrate instrument:

    • 200 ppm CO

    • 20 ppm H2S

    • 50% LEL methane (CH4)

  • If you use different concentrations you must change instrument settings!


How does the accuracy of the cal gas affect the accuracy of the readings
How does the accuracy of the cal gas affect the accuracy of the readings?

  • Standard GfG Quad Mix cal gas has 25 ppm H2S, 200 ppm CO, 50% LEL CH4; +/- 10% accuracy, w/ 12-months shelf life dating

  • GfG Quad Mix also available with +/- 3% accuracy, w/ 6-months shelf life dating

  • Accuracy of sensors = +/- 5% of reading

  • Using +/- 3% accuracy gas means the accuracy statement for the combined instrument and cal gas system becomes +/- 8% of reading, or the minimum resolution of sensor, whichever is greater


Range and resolution of gfg instruments used to detect h2s
Range and resolution of GfG instruments used to detect H2S the readings?

  • Standard full range is 0 – 100 ppm. (Users can optionally choose a 0 – 500 ppm extended full range version of the product.)

  • Standard setup displays readings in 0.2 ppm steps over the 0 – 1.0 ppm range.

  • From 1.0 – 99.9 ppm measurements are always incremented in 0.1 ppm steps.


Effect of 5 sensor accuracy on readings
Effect of ± 5% sensor accuracy on readings the readings?

At 1.0 ppm actual, reading between 1.05 and 0.9 ppm, so accuracy becomes +/- minimum resolution = +/- 0.2 ppm

0 5 10 15 25 30


Effect of calibration gas 3 and sensor accuracy 5 on readings combined effects 8
Effect of calibration gas (± 3%) and sensor accuracy (±5) on readings: combined effects = ±8%

At 1.0 ppm actual, reading between 1.08 and 0.92 ppm, so accuracy becomes +/- minimum resolution = +/- 0.2 ppm

0 5 10 15 25 30


So how accurate are the readings
So how accurate are the readings? on readings: combined effects = ±8%

  • Thus, on a standard basis, from 0 – 4 ppm the accuracy is +/- 0.2 ppm; from 4.0 to 100 ppm the accuracy is +/- 5% of reading.

  • If you choose the 0.1 ppm resolution over the 0 – 1.0 ppm range, from 0 – 2.0 ppm the accuracy is +/- 0.1 ppm, from 2.0 to 100 ppm the accuracy is +/- 5% of reading


Use ds400 docking station for daily bump check and or periodic calibration
Use DS400 Docking Station for daily bump check and / or periodic calibration

  • Standalone operation:

  • No PC required

  • Automatic bump-test

  • Automatic span calibration

  • Success / failure indication after each test

  • Economic use of test gas

  • Easy collection of unit test and logged data

  • Reduced maintenance cost


Use ts 400 test station for daily bump check
Use TS-400 Test Station for daily bump check periodic calibration

  • Standalone operation - No PC required!

  • Automatic bump-test only

  • Success / failure indication after each test

  • Economic use of test gas

  • Easy collection of unit test and logged data


Comprehensive test reports

Comprehensive Test Reports periodic calibration

Easy to generate custom test reports



Ds 400 docking station for daily bump check and or periodic calibration
DS-400 Docking Station for daily bump check and / or periodic calibration

  • Standalone operation:

    • No computer required!

  • Docking station controlled by instrument’s control buttons:

    • Push "Test" for Bump Test

    • Push "Cal" for Auto Cal

    • Push "Cancel" to charge only


  • Ds 404 multi inlet docking station for bump check and periodic calibration
    DS-404 Multi-inlet Docking Station for bump check and periodic calibration

    • Inlets for 4 cylinders of gas

    • Automatic Bump and Cal for 5, 6 and 7 channel instruments

    • Supports:

      • Quad mix (O2, LEL, CO, H2S)

      • 5-mix with SO2

      • 5-mix with CO2

      • Isobutylene

      • HCN

      • NO2

      • And other individual gases


    Using ds400 docking station for daily bump check and or periodic calibration
    Using DS400 Docking Station for daily bump check and / or periodic calibration

    • Standalone operation: DS-400 does not require connection to an external computer

    • Test results and other messages displayed on instrument LCD


    Using ds 400 docking station for daily bump check and or periodic calibration
    Using DS-400 Docking Station for daily bump check and / or periodic calibration

    • Using DS-400 allows instrument to record "Bump Test" as specific event in instrument's memory

    • It is possible to set a "Bump Test Due" date in the instrument's memory

    • Only way to reset next "Bump Test Due" date is by means of DS-400 Docking Station


    Using ds 400 docking station for daily bump check and or periodic calibration1
    Using DS-400 Docking Station for daily bump check and / or periodic calibration

    • Make sure DS-400 attached to test gas and power

    • Make sure the instrument is located in fresh air, turn on, and allow to warm up

    • Note instrument readings, and perform fresh air zero if necessary

    • Place instrument in DS-400

    • Docking station controlled by instrument’s control buttons:

      • Push "Test" for Bump Test

      • Push "Cal" for Auto Cal

      • Push "Cancel" to charge only


    How to do bump test
    How to do bump test periodic calibration

    • Make sure instrument turned on

      • Best if instrument allowed to warm-up at least 5 minutes before bump test

    • Make sure Docking Station plugged in and connected to gas

      • Higher inlet is for fresh air

      • Lower inlet is for span gas

    • Place G460 in Docking Station

    • Instrument display will indicate status and results of test


    How to do bump test1
    How to do bump test periodic calibration

    • After 10 second count-down; Docking Station automatically performs bump test

    • Instrument screen shows a check mark besides each sensor as bump check completed

    • If G460 is not removed from Docking Station; 5 minutes after test completed G450 turns off and goes into charging mode


    Ds400 docking station

    DS400 Docking Station periodic calibration

    Bump-Test includes:

    Visual alarm function

    Audible alarm function

    Time for activation to alarm 1

    Time for activationto alarm 2

    Time to t50

    Calibration Test Includes:

    Fresh air zero adjustment

    Span calibration adjustment

    All test results:

    Stored to instrument memory

    Stored to flash memory card in Docking Station


    What to do if g460 fails bump test
    What to do if G460 fails bump test periodic calibration

    • Red color warning screen and message indicates bump test failed

    • This means G460 needs to be “Autocal” adjusted before further use

    • Remove instrument from Dock, and press “Reset” (center) G460 control button to clear alarm message

    • Place G460 back in Dock

    • Before end of count down press Autocal button


    Questions
    Questions? periodic calibration