OPTIMIZATION OF CR IMAGES. The impression that CR images can always be adjusted after exposing the CR with x-rays is not necessarily true. There are several factors affecting the quality of CR images, and radiographers or technologists are the key persons who are responsible in delivering good quality radiographs, with reasonable radiation dose given to the patients. Quality control of the technical parameters and radiographic positioning are therefore critical to a CR image. Optimization of 9448

Pat_Xavi
Download Presentation

Digital Artifacts Imaging ERRORS

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


    1. Digital Artifacts & Imaging & ERRORS The advantages of CR are its large dynamic range, digital format, portability, and post-processing capability

    3. Positioning and collimation Exposure techniques Image processing selection Lifetime of the PSP

    4. Collimation Proper collimation reduces scattered radiation in the region of interest and reduces the noise that degrades the radiographic contrast. This good practice is still valid with CR, and most image processing software employed in CR relies on the fact that the image collimator edge is detected, so that the contrast may be optimised. Failure of the software to define the image boundary may be caused by a number of factors.

    5. Positioning of Collimation The F/S method of two projections of a hand radiograph in one 18 cm x 24 cm film. Is not a good practice with CR technique, since double or multiple exposures on a single photostimulable phosphor (PSP) can lead to a failure of the image processing software to detect the image boundary. Matching the positioning and collimation with the image processing parameters is also crucial. Some radiographers may take a radiograph of a lumbosacral spine without collimation, thus making the radiograph looks more like an image of the Kidney-Ureter-Bladder (KUB) technique. Image processing will eventually fail to process since the input information is totally different.

    6. Exposure techniques In order to introduce CR as a replacement for film-screen technique, the common thinking is that it would be reasonable to adhere to the same exposure techniques to help the radiographers to adapt to the newer technology. But this is not necessarily the case. CR may be operated at a different film speed, and then optimizing the exposure technique accordingly. Existing CR has a speed similar to medium speed film-screen system (200) while spatial resolution is still generally inferior

    7. Exposure cont. The idea of reducing radiation dose to patients when switching from screen-film system to CR may not always be valid. To keep the same signal to noise ratio, CR needs 20% more radiation exposure as we treat CR as medium speed film. Reduction of radiation dose to the patient will then results from reduction of reject rate due to poor exposure technique. As a result of poorer intrinsic spatial resolution of the PSP, radiographers need to make sure that when they set up exposure factors, i.e., the mA station from small focal spot should be selected when imaging bones or other high resolution required body parts.

    8. Some radiographers may still use too low kVp for chest radiographs. Employing a standard high kVp technique when the pre-set kVp range for image processing may be higher prevents optimization of the image quality.

    9. proper adjustment of exposure technique is therefore still crucial in any radiography practice. Although an increased radiation exposure would yield a higher signal-to-noise ratio and better low contrast detectability in PSP, this would clearly violate the “As low as Reasonably Achievable” (ALARA) principle.

    10. Some examples of artifacts in CR (a) an image with loss of contrast as a result of improper selection of image processing; the same image now shows acceptable image quality as a result of proper selection of image processing.

    11. Proper selection of an image processing algorithm specific to each type of x-ray examination is thus important. The technical skills of radiographers definitely play a crucial role in determining the quality of the radiographic image. Even though a CR image may be adjusted to improve the image visibility in the cases of over- or under-exposures, it would still be impossible for an image processing to improve the visibility of clinical features that were not available in the raw image.

    12. Lifetime of the PSP One of the major advantages of CR is that the phosphor plate is reusable. However, there are a number of factors that may affect the lifetime of an imaging plate. The plates are subjected to normal wear and tear from scratches, scuffs, cracks, and contamination with dust and dirt, which may interfere with the production of a good image. The establishment of a well-organized quality control program will play an important role in assessing the clinical quality of the imaging plate. This may easily be carried out by artifact assessment and uniformity evaluation across the plate.

    13. Digital artifacts Some information obtained from: AR Online » Current Issue » Artifacts and misadventures in digital radiography By Charles E. Willis, PhD, DABR; Stephen K. Thompson, MS, DABR; S. Jeff Shepard, MS, DABR Volume: 33 Number: 1 January 2004

    14. CR artifacts require special attention. This is due to the fact that CR artifacts may be produced from various components of the CR system itself Artifacts may also be generated by the users who are not aware of the proper imaging techniques or selection of appropriate image processing protocols Since CR is also very sensitive to scattered radiation, it is vital that anti-scattered grids be used as in conventional radiography. Radiographers should be concerned since these may generate unwanted artifacts that could not be corrected by any image processing algorithm.

    17. Halo effect with Edge enhancement

    27. Dirt on screens

    28. Acquiring good quality images Regardless of the acquisition technology, good radiographic images can be produced only when certain fundamental requirements are met. Appropriate radiographic technique must be used, proper tube potential (kVp), beam current (mAs), source-to-image distance (SID), collimation, alignment of the X-ray central ray, and positioning of detector and subject for the specific anatomic projection.

    29. The detector must receive enough X-rays to make a good image. Although DR is more tolerant of incorrect exposure factor selection, it cannot make up for extra noise, loss in subject contrast, and signal out of its range of adjustment.

    30. Whether from underexposure or misalignment of a scatter reduction grid, too few X-rays produce noisy images

    31. Too many X-rays are a disservice to the patient and may also produce poor images (Figure 3).

    32. Exposure factor creep" is a well-known phenomenon related to the wide exposure latitude of DR. Observers tend to complain about noise in DR images exposed at one-fourth to one-half of the appropriate level. On the other hand, artifacts are generally not apparent until the exposure exceeds 10 times the appropriate level. Technologists soon learn to avoid the unpleasant circumstance of repeating an underexposed study by routinely increasing the radiographic technique. Thus, the potential for gross overexposure exists in DR. Image optical density (OD), the usual indicator of proper exposure, is arbitrary in DR. Managment of exposure factor in DR must rely on the value of a derived exposure indicator, which itself is subject to interference. However, programs that monitor the exposure indicator have been shown to be effective in controlling exposure factor in DR.10

    33. Processing algorithms used by the DR system must be designed to anticipate the use of this display function in order to properly render the image for display. Not all vendors adhere to this standard. Calibrated, high-quality QC monitors are essential on every acquisition system and QC workstation. Adjusting image processing on an uncalibrated monitor leads to unsatisfactory images.

    37. Double exposure is a classic operator error that constitutes approximately 2% of all rejected images. The consequence of double exposure can be either a single repeated examination, when an inanimate object is involved (Figure 11), or two repeated examinations when two patients are involved (Figure 12). In DR, double exposures can also be caused by power interruptions and communications errors, as well as by inadequate erasure secondary to overexposure or erasure mechanism failure.

    38. Double exposure (Figure 11),

    39. Double exposure # 12

    40. Abdomen not cleared completely before chest Used large cassette to balance small for patella on merchant boardAbdomen not cleared completely before chest Used large cassette to balance small for patella on merchant board

    41. Appropriate digital image processing is key to producing good DR images. All DR systems have extremely wide latitude, which means that connected to a display system with a relatively narrow dynamic range, DR images have extremely low contrast. The primary purpose of image processing is to maximize the contrast of the part of the image that contains relevant clinical details. To do this, the DR system locates either the boundary of collimation or the border of the projected anatomy, and disregards details outside this boundary.

    42. Errors in collimation can cause mistakes in detection of the boundary, with a dramatic loss of image contrast (Figures 13 and 14).

    43. Quality control processes must be in place to detect and correct unsatisfactory images (Figures 8, 9, and 10).

    44. #8

    45. #9

    46. #10

    47. The secondary function of image processing is to customize contrast in the region of interest (Table 1). This type of image processing includes modifying the image to enhance the contrast and sharpness of some features while compromising the contrast and sharpness of others, as well as modifying the image to make it appear more like a conventional transilluminated film. This secondary image processing is applied in a manner that is usually specific to the anatomic projection.

    48. Errors in the selection of the anatomic projection can cause inappropriate processing (Figure 15).

    49. Different exam parameters in different rooms

    50. An auxiliary purpose of image processing is to improve the usability of the digital image. This includes imprinting demographic overlays, adding annotations, applying borders and shadow masks, flipping and rotating, increasing magnification, conjoining images for special examinations like scoliosis, and modifying the sequence of views. This processing may require a separate QC workstation

    51. Misuses of image processing include compensating for inappropriate radiographic technique, compensating for poor calibration of acquisition and display devices, and surreptitious deletion of nondiagnostic images. Image processing to recover nondiagnostic images to prevent re-exposure should be a last resort, not a routine activity. Routine reprocessing indicates a problem with automatic image processing or technical practice. Access to image-processing software is essential to develop and maintain appropriate processing parameters

    52. Automatic image processing involves assumptions about the radiographic technique, the composition of anatomic region imaged, and the use of collimation. A number of factors can interfere with the automatic detection of the boundaries of the radiation field, including nonparallel collimation, use of multiple fields on a single imaging plate, poor centering, implants (especially when they overlie the boundary), and violation of collimation rules provided by the vendor.

    53. For example, placement of gonadal shields is no longer trivial, but may adversely affect image quality.

    54. What is the problem?

    56. Wider dynamic range means that technologists have to pay attention to exposure indicator values, instead of brightness and contrast. Without this attention, patient dose will escalate. If exposure indicator logs are available, they need to be evaluated. If they aren't, this will need to be done manually.

    57. New technologies should be developed for dealing with pediatric examinations and patients with prosthetic devices. New image processing strategies may be needed with these special patients, as well. Thorough training and active onsite support of the technical staff are crucial. For many, this is a completely different way of thinking about the imaging process. Technologists are generally eager to become involved and master this new technology, but they need proper training and guidance to use it effectively to produce diagnostic-quality images.

    58. Exhausted developer Every time a film is inserted into the processor, a small amount of replenisher is released into the developer tank to keep the developer active. Practically all automatic x-ray processors were designed to process at least 30 films per day. The developer needs that many films going through it to give the developer enough replenisher to keep from becoming rapidly exhausted. Many facilities do not process nearly this many films in a day, and then they wonder why their chemicals are losing strength in a week and a half! This is even more noticeable in a small counter-top processor

    59. Same pt – what improved this image?

    60. Screens can be check with a UV light for problems

    74. Single filed per plate - better