实验三

实验三. 火焰原子吸收光谱法灵敏度 和自来水中钙、镁的测定. 实验原理. 在使用锐线光源条件下,基态原子蒸气对共振线的吸收,符合朗伯 - 比尔定律,即 A=lg ( I 0 /I ) =KLN 0 ,在一定实验条件下,待测元素原子总数目与该元素在试样中的浓度成正比,则 A=KC 。 由标准曲线法算出元素的含量,并计算灵敏度:. 仪器与试剂. 1 .仪器: TAS-990 型原子吸收分光光度仪 2 .试剂: ( 1 ) 50 mg·L -1 镁标准使用溶液 ( 2 ) 100 mg·L -1 钙标准使用溶液

zody
Download Presentation

实验三

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. 实验三 火焰原子吸收光谱法灵敏度 和自来水中钙、镁的测定

  2. 实验原理 在使用锐线光源条件下,基态原子蒸气对共振线的吸收,符合朗伯-比尔定律,即A=lg(I0/I)=KLN0,在一定实验条件下,待测元素原子总数目与该元素在试样中的浓度成正比,则A=KC。 由标准曲线法算出元素的含量,并计算灵敏度:

  3. 仪器与试剂 1.仪器: TAS-990型原子吸收分光光度仪 2.试剂: (1)50 mg·L-1镁标准使用溶液 (2)100 mg·L-1钙标准使用溶液 (3)MgO(GR),无水CaCO3(GR),HCl(AR) 配制用水均为二次蒸馏水。

  4. 实验步骤 1.Ca,Mg标液的配制 (1)Ca:2.0,4.0,6.0,8.0,10.0 mg·L-1 (2)Mg:0.1,0.2,0.3,0.4,0.5 mg·L-1 2.工作条件的设置 (1)吸收线波长 Ca:422.7 nm Mg:285.2 nm (2)空心阴极灯电流 Ca:3.0 mA Mg:2.0 mA (3)狭缝宽度 0.4 mm (4)原电子化器高度 6 mm (5)空气流量4 L·min-1,乙炔气流量1.2 L·min-1

  5. 3.Ca的测定 (1)用10 mL移液管,吸取自来水样于100 mL容量瓶中,用蒸馏水稀释至刻度, 摇匀。 (2)在最佳工作条件下,以蒸馏水为空白,由稀至浓逐个测量Ca系列标准溶液的吸光度,最后测量自来水样的吸光度A。 4.Mg的测定(同上) 5.结束后,用蒸馏水喷洗原子化系统2min,按关机程序关机。 6.绘制钙、镁的标准曲线,由未知样的吸光度求出自来水中的Ca,Mg含量。 7.根据测量数据,计算仪器测定Ca,Mg的灵敏度S。

  6. 思考题 为什么要配制Ca,Mg标液?所配制的Ca,Mg标液可以放置到第二天再使用吗?为什么?

More Related