1 / 44

BME 311: BIOMEDICAL INSTRUMENTATION I Lecturer: Ali Işın

FACULTY OF ENGINEERING DEPARTMENT OF BIOMEDICAL ENGINEERING. BME 311: BIOMEDICAL INSTRUMENTATION I Lecturer: Ali Işın. Lecture Note 4: ECG Systems. ECG Systems. What is an E C G?. The electrocardiogram (E C G) is a representation of the electrical events of the cardiac cycle.

yreynosa
Download Presentation

BME 311: BIOMEDICAL INSTRUMENTATION I Lecturer: Ali Işın

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. FACULTY OF ENGINEERING DEPARTMENT OF BIOMEDICAL ENGINEERING BME 311: BIOMEDICAL INSTRUMENTATION ILecturer: Ali Işın Lecture Note 4: ECG Systems BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  2. ECG Systems BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  3. What is an ECG? The electrocardiogram (ECG) is a representation of the electrical events of the cardiac cycle. Each event has a distinctive waveform, the study of which can lead to greater insight into a patient’s cardiac pathophysiology. BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  4. The Normal Conduction System BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  5. The Normal Conduction System BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  6. ECG ECG is a surface measurement of the electrical potential generated by electrical activity in cardiac tissue. BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  7. ECG The human heart can be considered as a large muscle whose beating is simply muscular contraction. Therefore contractions of the heart cause a potential to be developed. The measurement of the potential produced by cardiac muscle is called electrocardiology. BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  8. What types of pathology can we identify and study from ECGs? • Arrhythmias • Myocardial ischemia and infarction • Pericarditis • Chamber hypertrophy • Electrolyte disturbances (i.e. hyperkalemia, hypokalemia) • Drug toxicity (i.e. digoxin and drugs which prolong the QT interval) BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  9. Waveforms and Intervals BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  10. 3. Ventricular repolarization 2. Ventriculardepolarization 1. Atrial depolarization BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  11. Different Segments of ECG P wave: the sequential activation (depolarization) of the right and left atria  QRS complex: right and left ventricular depolarization (normally the ventricles are activated simultaneously)  ST-T wave: ventricular repolarization  U wave: origin for this wave is not clear - but probably represents "afterdepolarizations" in the ventricles  PR interval: time interval from onset of atrial depolarization (P wave) to onset of ventricular depolarization (QRS complex)  QRS duration: duration of ventricular muscle depolarization  QT interval: duration of ventricular depolarization and repolarization  RR interval: duration of ventricular cardiac cycle (an indicator of ventricular rate)  PP interval: duration of atrial cycle (an indicator or atrial rate BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  12. ECG basics • Amplitude: 1-5 mV • Bandwidth: 0.05-100 Hz • Largest measurement error sources: • Motion artifacts • 50/60 Hz powerline interference • Typical applications: • Diagnosis of ischemia • Arrhythmia • Conduction defects BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  13. ECG Leads Leads are electrode setups which measure the difference in electrical potential between either: 1. Two different points on the body (bipolar leads) 2. One point on the body and a virtual reference point with zero electrical potential, located in the center of the heart (unipolar leads) BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  14. Cardiac Axis by Different Leads BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  15. Einthoven Triangle:Note potential difference for each lead of triangle BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  16. Each lead gives a slightly different representation of electrical activity of heart BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  17. ECG Leads The standard ECG has 12 leads: 3 Standard Limb Leads 3 Augmented Limb Leads 6 Precordial Leads The axis of a particular lead represents the viewpoint from which it looks at the heart. BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  18. Standard Limb Leads BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  19. Precordial Leads Adapted from: www.numed.co.uk/electrodepl.html BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  20. 12-Lead ECG measurement • Most widely used ECG measurement setup in clinical environment • Signal is measured non-invasively with 9 electrodes + 1 reference electrode (right leg) Einthoven leads: I, II & III Goldberger augmented leads: VR, VL & VF Precordial leads: V1-V6 BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  21. 12 Lead ECG System BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  22. 12 Lead ECG System BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  23. Block Diagram of a Basic ECG System LeadSelector IsolatedPower Source BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  24. ECG Pre-Amplifier ECG pre-amplifier is a bioelectricdifferentialamplifier. Itincludes; • High impedance input of bioelectric amplifier • Lead selector switch • 1mV calibration pulsesource • Means of protecting amplifier from high voltage discharge such as a defibrillator dischargeused on a patient • Pre-Amplifier contains an instrumentation amplifier as well as an isolation amplifierforpatientsafety BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  25. Isolation Amplifier • Needed for safety! Isolates the patient from high voltages and currents to prevent electric shock by introducing a specific barrier between passage of current from the power line to the patient. • Can be done by using light (photo emitter and photo detector) or a transformer (set of inductors that are used in a step up / step down configuration) BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  26. Common Mode Rejection BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  27. ECG Amplifier Circuit With Right Leg Driver BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  28. Common Mode Voltage (CMV) • If 2 inputs are hooked together into a differential amplifier driven by a common source with respect to ground, then the common mode voltage should be the same and the ideal output should be zero. However practically you will see a voltage. • CMV is composed of 2 parts: • DC electrode offset potential • 50Hz AC induced interference caused by magnetic and electric fields from power lines and transformers • Capacitively coupled into circuit BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  29. Analysis to reduce noise in ECG • Right leg driver circuit is used in a feedback configuration to reduce 50 Hz noise and to drive the noise on the patient to a lower level. BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  30. “6” ways to reduce Noise in ECG • Common Mode Rejection (differential Amplifier) • Right Leg Drive (feedback loop to decrease noise) • Shielding of wires • Isolation amplifier • Notch filter to reduce 50 Hz noise • Bandpass filter to reduce noise below and above ECG bandwidth (0.05-100 Hz) BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  31. High Pass Active Filters Attenuates frequencies lower than the cutoff frequency. cutoff frequency is 1/(2)=1/ 2RiCi Rf Ci Ri - A Vinput + Voutput Ri Ci Voutput Rf Vinput 0 Ii IRf BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  32. Low Pass Active Filters = Integrator Cf Attenuates frequencies higher than the cutoff frequency Rf Ri - A Vinput Voutput + Cf ICf Ri Rf Voutput Vinput 0 Ii IRf BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  33. Defibrillator Protection Circuit • Defibrillator = a high voltage electrical heart stimulator used to resuscitate heart attack victims • When the physician uses defibrillator, the high voltages and currents discharged onto patient can cause damage to medical equipment(specificly pre-amplifier circuitry), BUT physician still needs to view ECG of the patient while defibrillating. • How do you protect your medical equipment from excessive voltages and currents? BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  34. Protection Devices in ECGs: Neon Glow Lamps • Neon Glow Lamps are pair of electrodes mounted in a glass envelope filled with low pressured neon gas or a mix of other inert gases. • Normally impedance across the electrodes is very high (so lamps are not conductive) but when thepotential across the electrodes reachesto the ionization point of the gas, impedance suddenly drops making the lamps conductive (they light up) creating a short circuit to ground,transferring excess energy safely to the ground without damaging the amplifier. BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  35. Protection Devices in ECGs: Zener Diodes • Zener Diode: is a diode which allows current to flow in the forward direction in the same manner as an ideal diode, but also permits it to flow in the reverse direction when the voltage is above a certain value known as the breakdown voltage. • In this configuration when a largerpotentialthan the breakdown voltage is applied to the system (i.e. Defibrillator discharge) the zener diode allow the current to flow in reverse direction and shunts it to the ground thus preventing any damage to the amplifier. BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  36. Protection Devices in ECGs: Current-Limiting Diodes • Current Limiting Diodes are electronic devices that limit current to a maximum specified value for the device. • These diodes consists of a n-channel JFET-transistor • They act as a resistor as long as the current level remains below the limiting value. If current tends to rise above the limit (as in the case of a Defibrillator discharge), it will beclamped and limited preventing any excess current reaching the amplifier thus protecting it. BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  37. Electro-Surgery Unit (ESU) Filtering • During surgery simultaneous ECG recording is also required. • ESU can introduce high frequency signals with frequencies ranging from 100KHz to 100 MHz and with magnitudes up to few kVolts into the ECG signal. This interferance can heavily distort the ECG signal. • ESU introduces: • DC offsets • and obscures the ECG signal BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  38. Even though the ECG frequency range is between 0.05-100 Hz and all the higher frequency signals are attenuated by using low-pass filters at filtering stage (slide 32), ESU signal still manages to disort ECG. • Because; • Internal junction points of the ECG amplifier can rectify high frequency signals like ESU signal and the parasitic capacitance between the leg connections of the amplifier further filters this signal to create a dc ofset potential. (Whenever the ESU device is triggered during the surgery ECG baseline oscilates up and down.) BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  39. And the ESU signal can leak through amplifier and low pass filter layers obscuring the ECG signal. • ECG needs to be of diagnostic quality sothe ESU noise should be eliminated • Common technique is to use a pi-type 3-layered RC filter BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  40. RC Filters • Low Pass Filters will pass frequencies lower than cutoff frequency of FH =1/2RC Vs FH Frequency Vs FL • High Pass Filters will pass frequencies greater than cutoff frequency of FL =1/2RC BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  41. 3-Layered RC ESU Filter RC Filter RA Defibrillator Protection Circuit ECG Differential Amplifier LA LC filter design can also be used BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  42. Some Sample ECG Circuits High-precision analog front end of a portable ECG application. 3 patient electrodes (only 3 bipolar leads) BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  43. ECG system with 5-patient electrodes (7-leads) BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

  44. Standalone ECG with Compact Design BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014

More Related