Determination of Uranium in Urine using EIChrom Resins

Determination of Uranium in Urine using EIChrom Resins. Helen E Carter Loughborough University UK Arthur E Lally OBE RTC Ltd UK Michaela Langer EIChrom Europe France Peter Warwick Loughborough University UK. Evaluation of some methods. Reason for the study

xarles
Download Presentation

Determination of Uranium in Urine using EIChrom Resins

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Determination of Uranium in Urine using EIChrom Resins Helen E Carter Loughborough University UK Arthur E Lally OBE RTC Ltd UK Michaela Langer EIChrom Europe France Peter Warwick Loughborough University UK EIChrom User Group Meeting 15/11/99

  2. Evaluation of some methods • Reason for the study • Inconsistency in results from certain methods • Attribute possible causes EIChrom User Group Meeting 15/11/99

  3. Problem areas? • A satisfactory analytical result requires • Combination of • Analyst • Method • Equipment • Instrumentation • Laboratory environment • Materials and chemicals EIChrom User Group Meeting 15/11/99

  4. Analyst • Trained • limited experience • Limited time to test effectiveness • Manual procedures EIChrom User Group Meeting 15/11/99

  5. Analytical Methodology • For this study all were similar • Method A no apparent problems • Method B inconsistencies in results • Method C less of a problem than B • Method D small inconsistencies EIChrom User Group Meeting 15/11/99

  6. Equipment • Materials EIChrom resin TRU • Batch to batch variation • Storage EIChrom User Group Meeting 15/11/99

  7. Study protocol Part 1 • PhD student - initial work • Familiarisation with techniques using • Synthetic urine • Genuine urine • Co-precipitation • Electro deposition • Actual methods EIChrom User Group Meeting 15/11/99

  8. Study protocol Part 2 • Evaluation of data • PhD student finished her work and left • Final tests on hydrolysis to be carried out • Conclusions to be made EIChrom User Group Meeting 15/11/99

  9. Outline - Method A • Mineralisation with HNO3 • Copptn on Ca phosphate • Muffle at 900Oc 8 hrs • Dissolution HNO3+ Al nitrate • Reduction with Fe(II) sulphate • TRU resin column • Elution with Amm. Oxalate • Source preparation • Reduction with TiCl3 • Micro pptn on CeF3 EIChrom User Group Meeting 15/11/99

  10. Outline – Method B • Mineralise with HNO3/H2O2 • Muffle at 500oc 6 hrs • Hydrolyse 1.4 M HNO3 • Copptn Ca phosphate • Ferrous sulphate/ Al2(NO3)3 • TRU resin • U elution • Micro pptn with LaF3 EIChrom User Group Meeting 15/11/99

  11. Outline- Method C • Similar to method B but • Muffle 12hrs at 500oC • Separation on AGMP1 resin • Pu on column • U in load + 8M HNO3 wash • U fraction on to TRU resin • As for method B EIChrom User Group Meeting 15/11/99

  12. Outline- Method D • Mineralisation and evaporation on hotplate • Hotplate to 500oC 5hrs • Dissolution / hydrolysis in 2M HNO3 • Co pptn Ca phosphate • Dissolution HCl • Co pptn Fe phosphate • As per method C EIChrom User Group Meeting 15/11/99

  13. Results - same batch of resin • Method A, Mock Urine • Batch TR8B resin • CeF3 pptn • % recovery • 84, 75, 77, 94, 88, 92, • 89, 95, 92, 87, 93, 77, EIChrom User Group Meeting 15/11/99

  14. Results – DifferentBatches of Resin EIChrom User Group Meeting 15/11/99

  15. Results using different methods EIChrom User Group Meeting 15/11/99

  16. Results- Different methods (real urine) EIChrom User Group Meeting 15/11/99

  17. Conclusions – to date • Consistency for Method A and analyst • satisfactory • Different batches ofresin • No real evidence of effect • Different methods • A, C and D reasonable • B poor EIChrom User Group Meeting 15/11/99

  18. What is going on? Analyst or method ! • Method A always OK • Method B always poor • Give the analyst a chance! • What else could be wrong? • HYDROLYSIS ? EIChrom User Group Meeting 15/11/99

  19. HydrolysisAlways been a problem • Actinides U and Pu love pyrophosphates better than any ion-exchange resin • (yes even EIChrom!) • Experiment with hydrolysis procedures • New analyst on job • Old analyst in years!! EIChrom User Group Meeting 15/11/99

  20. Hydrolysis proceduresevaluated using:- • Mock Urine • Method B • CeF3 micro pptn • U-233 spike • Very large prayer mat. EIChrom User Group Meeting 15/11/99

  21. Method B amendments 1 • Samples 1 and 2 • Standard 500oC 6hrs • Standard 1.4M nitric hydrolysis and Fe(II) SO4 • Samples 3 and 4 • As above but 500oC 18hrs EIChrom User Group Meeting 15/11/99

  22. Method Bamendments 2 • Samples 5 and 6 • Standard 500oC 6hrs • Standard 1.4M nitric but- with ferrous sulphamate and Fe(II) SO4 • Samples 7 and 8 • As above but 500oC 18hrs EIChrom User Group Meeting 15/11/99

  23. Results of hydrolysischanges to Method B EIChrom User Group Meeting 15/11/99

  24. Conclusions of hydrolysis changes • No significant differences • Method B appears to give good results EIChrom User Group Meeting 15/11/99

  25. Overall Conclusions • The resin appears to make no difference with regards to • Age • Batch • Packing • Methods all appear OK at times • What about the operator? • Think about method, operator and equipment combination! EIChrom User Group Meeting 15/11/99

  26. Summary of all experiments • No effect of resin • All methods satisfactory • Human factor • Bored staff? • Staff in wrong method? • Automation • Who wants an old chemist? EIChrom User Group Meeting 15/11/99