CENTRO DE METROLOGIA QUIMICA. CALCULATION OF UNCERTAINTIES FOR THE RESULTS REPORTED IN SIM.QM-P2. As, Cu, Fe, Mn, Ni, Pb, (Zn) Perkin-Elmer AAnalyst700 Atomic Absorption Spectrophotometer equipped with a longitudinal graphite furnace. ANALYTICAL MEASUREMENT. Cd

Download Presentation

CENTRO DE METROLOGIA QUIMICA

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript


Centro de metrologia quimica

CENTRO DE METROLOGIA QUIMICA


Calculation of uncertainties for the results reported in sim qm p2

CALCULATION OFUNCERTAINTIESFOR THE RESULTSREPORTED IN SIM.QM-P2


Analytical measurement

As, Cu, Fe, Mn, Ni, Pb, (Zn)

Perkin-Elmer AAnalyst700 Atomic Absorption Spectrophotometer equipped with a longitudinal graphite furnace

ANALYTICAL MEASUREMENT


Analytical measurement1

Cd

Flame Atomic Absorption Spectrophotometry

Perkin-Elmer AAnalyst100 and using high sensitivity nebulizer

ANALYTICAL MEASUREMENT


Analytical measurement2

NIST-SRMs were used for the preparation of calibration standards.

The quantification method used was bracketing

Four independent aliquots of the sample were analyzed, in all cases.

NIST-SMR 1643d was used as control (except for Cd)

ANALYTICAL MEASUREMENT


Centro de metrologia quimica

THE UNCERTAINTY ESTIMATION PROCESS

Specify associated model

Identify Uncertainty Sources

QuantifyUncertainty Sources

Convert to standarduncertainty

Calculate combinedstandarduncertainty

Reporting Uncertainty


Centro de metrologia quimica

ASSOCIATED MODEL:

BRACKETING

AH

AS

AL

CL

CS

CH

Sample concentration is calculated through the equation


Identyfing uncertainty sources

IDENTYFING UNCERTAINTY SOURCES

AL, AH, As

d

repeatability

repeatability

calibration

temperature

Csample

F(dil)

U-SMR

repeatability

calibration

CL, CH

temperature


Quantifying uncertainty sources

QUANTIFYING UNCERTAINTY SOURCES

AL, AH, As

d

repeatability

repeatability

calibration

temperature

Csample

F(dil)

U-SMR

repeatability

calibration

CL, CH

temperature


Centro de metrologia quimica

Repeatability of A(S,L,H)

QUANTIFYING UNCERTAINTY SOURCES

The uncertainty due to random variation of the EAA signal measurement of samples and bracketing standards was calculated as:

Fe Determination


Centro de metrologia quimica

QUANTIFYINGUNCERTAINTY SOURCES

AL, AH, As

d

repeatability

repeatability

calibration

temperature

Csample

F(dil)

U-SMR

repeatability

calibration

CL, CH

temperature


Centro de metrologia quimica

Csample

F(dil)

U-SMR

repeatability

calibration

temperature

CL, CH

Equation for concentration assessment

Equation for the uncertainty calculation


Centro de metrologia quimica

Csample

F(dil)

U-SMR

repeatability

calibration

temperature

CL, CH


Centro de metrologia quimica

PREPARATION OF STANDARDS

Values and uncertainties

CL = 9.827 mg/L

uC = 0.055 mg/L


Centro de metrologia quimica

QUANTIFYINGUNCERTAINTY SOURCES

AL, AH, As

d

repeatability

repeatability

calibration

temperature

Csample

F(dil)

U-SMR

repeatability

calibration

CL, CH

temperature


Centro de metrologia quimica

CALCULATING THE CONCENTRATION


Centro de metrologia quimica

CALCULATING THE COMBINEDSTANDARDUNCERTAINTY


Centro de metrologia quimica

KRAGTEN(*)

Finally

(*) J. Kragten, Analyst, 119, 2161-2166 (1994)


Centro de metrologia quimica

CALCULATION OF COMBINEDSTANDARDUNCERTAINTY


Uncertainty contributions

UNCERTAINTY CONTRIBUTIONS


Centro de metrologia quimica

Finally, the concentration and uncertainty values for Fe in the sample SIM.QM-P2 are:

Concentration CS = 93.19 mg/L

Combined uncertainty ucom = 0.92 mg/L

Expanded uncertainty U = 1.84 mg/L

U=u*k, (k=2) (1.97%)


Centro de metrologia quimica

RESULTS OF Intec IN SIM.QM-P2

SUMMARY OF RESULTS OBTAINED ON SAMPLE SIM-P2

Element

Average concentration

Expanded Uncertainty

REU

(%)

(

g/L)

(

g/L)

m

m

Ni

62,86

2,52

4,0

Fe

93,19

1,84

2,0

Cu

17,06

0,44

2,6

As

50,94

1,02

2,0

Pb

21,86

0,66

3,0

Mn

39,44

1,58

4,0

Cd

144

4

2,8


Centro de metrologia quimica

RESULTS OF SIM.QM-P2


Centro de metrologia quimica

RESULTS OF SIM.QM-P2


Centro de metrologia quimica