1 / 12

INSTRUCTIONS FOR AEROCHAMBER MINI* AEROSOL CHAMBER (AC)

INSTRUCTIONS FOR AEROCHAMBER MINI* AEROSOL CHAMBER (AC). For Use with Invasive Ventilation (ET Tube). For Use with Invasive Ventilation (ET Tube). Remove the Metered Dose Inhaler (MDI) aerosol canister from the MDI boot. For Use with Invasive Ventilation (ET Tube).

tress
Download Presentation

INSTRUCTIONS FOR AEROCHAMBER MINI* AEROSOL CHAMBER (AC)

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. INSTRUCTIONS FOR AEROCHAMBER MINI* AEROSOL CHAMBER (AC)

  2. For Use with Invasive Ventilation (ET Tube)

  3. For Use with Invasive Ventilation (ET Tube) • Remove the Metered Dose Inhaler (MDI) aerosol canister from the MDI boot.

  4. For Use with Invasive Ventilation (ET Tube) • Insert the MDI canister into the AeroChamber mini* AC canister port.

  5. For Use with Invasive Ventilation (ET Tube) • Prime the MDI prior to use by shaking the MDI and chamber and then depressing the MDI. Repeat this step a second time.

  6. For Use with Invasive Ventilation (ET Tube) • Attach the manual resuscitator to the AeroChamber mini* AC 15mm Outside Diameter port located at the back of the Chamber.

  7. For Use with Invasive Ventilation (ET Tube) • Disconnect the Ventilator Circuit from the Endotracheal Tube 15mm connector.

  8. For Use with Invasive Ventilation (ET Tube) • Attach the AeroChamber mini* AC with the canister and Manual Resuscitator to the Endotracheal tube 15 mm connector.

  9. For Use with Invasive Ventilation (ET Tube) • While holding the MDI canister vertically, depress the MDI canister downward into the AeroChamber mini* AC actuator and introduce one puff of medication into the Chamber.

  10. For Use with Invasive Ventilation (ET Tube) • Manually ventilate the patient with the required volume for 15 -20 Breaths.

  11. For Use with Invasive Ventilation (ET Tube) NOTE: For more than one dose, remove the MDI canister from the AeroChamber mini* AC, shake the canister and re-insert into the AeroChamber mini*AC canister actuator. Repeat steps 7 through 8.

  12. For Use with Invasive Ventilation (ET Tube) • Disconnect the AeroChamber mini* AC from the Endotracheal Tube connector and connect the ventilator to the Endotracheal tube and commence the normal ventilation

More Related