Voyager Gas Chromatograph Concept. Voyager Basics. Truly portable Only 15 pounds (6.8kg) with battery installed 15.4” (39cm) long x 10.6” (27cm) wide x 5.9” (15cm) high Multiple Column, Dual Detector Portable GC 3 Columns (Heavy, Middle, and Light Compounds)

Download Presentation

Voyager Gas Chromatograph Concept

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript


Voyager gas chromatograph concept

Voyager

Gas Chromatograph

Concept


Voyager basics

Voyager Basics

  • Truly portable

    • Only 15 pounds (6.8kg) with battery installed

    • 15.4” (39cm) long x 10.6” (27cm) wide x 5.9” (15cm) high

  • Multiple Column, Dual Detector Portable GC

    • 3 Columns (Heavy, Middle, and Light Compounds)

    • PID (Photoionization Detector) and optional ECD (Electron Capture Detector)

  • 3 Levels of Instrument Operation

    • Factory programmed methods

    • User modified methods

    • Total VOC quick screen mode

  • Intrinsically Safe (Class 1, Division 1, Groups A,B,C & D)

    • The only portable GC with this classification


Voyager basics1

Voyager Basics

  • Completely self-contained

    • Internal carrier gas cylinder

    • Oven

    • Battery

  • Internal data logger

    • On-board data handling provides immediate results

    • No laptop needed in field

  • Comprehensive Windows softwarefor data download and reports

  • Easy to use

    • Minimal training required for point & press operation


Voyager chromatography

Voyager Chromatography

  • Carrier Gas:

    • High Purity Nitrogen or Ultra Zero Air

    • Carrier Life: 8 hours at 1800 psi

    • Carrier Flow Control via Software (Pressure)

  • Column Oven:

    • Isothermal 55oC to 80oC

    • Higher temperatures help clean out columns and avoid contamination

  • Detector & Injection Port Heated to Column Temperature

  • Two Sampling Methods

    • Variable Volume Sample Loop

    • Syringe Injection


Voyager chromatography1

Voyager Chromatography

  • Benefits of the 3 column set

    • Overlapping selectivities cover

      • Heavy compounds (C7 to C12)

      • Middle compounds (C3 to C7)

      • Light compounds (C1 to C3)

  • Simplified method development

  • Columns do not degrade

    • Vapor phase injections only

    • Moderate temperatures with no ramping


Column scheme voyager gc assay 1

Column Scheme: Voyager GC Assay # 1

PID

Carrier

Gas (N2)

Column A

70%

Column B

ECD

30%

Pre-Column

Column C

Make-Up Carrier Gas

Precolumn: 4m x 0.53mm x 2.0 um SPB-35

Column A (HEAVY): 8m x 0.25mm BLANK Fused Silica

Column B (MID-RANGE): 20m x 0.32mm x 1.0 um Supelcowax10 (PEG)

Column C (LIGHT): 15m x 0.32mm x 12 um Quadrex 007-1


Column scheme voyager gc assay 2 7

Column Scheme: Voyager GC Assay # 2 - 7

Column A

Carrier Gas

(N2 or UZ Air)

Column B

PID

Precolumn

Column C

Precolumn: 2.5m x 0.53mm x 2.0 um SPB-35

Column A (Heavy): 8m x 0.32mm BLANK Fused Silica

Column B (Mid-Range): 20m x 0.32mm x 1.5 um Supelcowax 10 (PEG)

Column C (Light): 1.2m x 1.6mm 1% SP1000 on 60/80 Carbopack B


Voyager assays

Voyager Assays

  • Target Compounds grouped into Assays (1-7)

  • Each Assay includes

    • GC Method

    • Integration Settings

    • Compound Library

  • Assay 1 Environmental Assay

    • Most comprehensive with 40 compounds

  • Assay 2-7 Occupational Health and Safety

    • Targeted by industry segment


Assay 1 environmental compound list

Chloromethane

Vinyl Chloride

Bromomethane

Chloroethane

Methylene Chloride

1,1-Dichloroethylene

Acetone

Carbon Disulfide

Vinyl Acetate

t-1,2-Dichloroethylene

Methyl Ethyl Ketone

c-1,2-Dichloroethylene

Chloroform

1,1,1-Trichloroethane

1,1,2-Trichloroethane

Dibromochloromethane

Chlorobenzene

Ethylbenzene

m-Xylene

p-Xylene

o-Xylene

Styrene

Bromoform

1,1,2,2-Tetrachloroethane

1,3-Dichlorobenzene

1,4-Dichlorobenzene

1,2-Dichlorobenzene

Assay #1 Environmental Compound List

Carbon Tetrachloride

Benzene

1,2-Dichloroethane

Trichloroethylene

2-Chloroethyl Vinyl Ether

1,2-Dichloropropane

Bromodichloromethane

Methyl Isobutyl Ketone

c-1,3-Dichloropropene

Toluene

t-1,3-Dichloropropene

2-Hexanone

Tetrachloroethylene


Voyager gas chromatograph concept

Assays #2-7: Occupational Health & Safety

Assay #2: Petroleum and Petrochemical

  • Methanol

  • Ethanol

  • MTBE

  • TAME

  • Benzene

  • Toluene

  • Ethylbenzene

  • m-, o-Xylenes

  • Naphthalene

    Assay #6: Surfactant Manufacture

  • Ethylene Oxide

  • Propylene Oxide

Assay #4: ABS Rubber

  • Acrylonitrile

  • 1,3-Butadiene

  • Styrene

  • Assay #5: Pulp & Paper

    • Hydrogen Sulfide

    • Methyl Mercaptan

    • Ethyl Mercaptan

    • Dimethyl Sulfide (DMS)

    • Dimethyl Disulfide (DMDS)

    • Acetone

    • Methyl Ethyl Ketone (MEK)

    • a-Pinene

    • Methanol

  • Assay #7: Latex Paint Monomers

    • Methyl Acrylate

    • Methyl Methacrylate

    • Ethyl Acrylate

    • n-Butyl Acrylate

    • iso-Butyl Acrylate

    • Acrylic Acid

    • Vinyl Acetate

    • Ethylene

    • Styrene


Assay customization

Assay Customization

  • Preprogrammed assays can be customized in the SiteChart software for specific applications

  • Existing assays are modified by adding or deleting compounds

  • Full control of all hardware and software methods development and calibration parameters

  • Customized assays are then sent to Voyager from the PC


  • Login