Inside the black box. fluidics. electronics. light. optics. 1) Blocking capillary. 2) Depth of focus. hoy@cf.ac.uk. Flow Cytometry : Problems with early systems. Detector. Hydrodynamic focusing. V. Kachel, H. Fellner-Feldegg & E. Menke -. MLM. Chapt. 3. Fluidics –

tamika
Download Presentation

Inside the black box

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Inside the black box

  2. fluidics electronics light optics

  3. 1) Blocking capillary 2) Depth of focus hoy@cf.ac.uk Flow Cytometry : Problems with early systems. Detector

  4. Hydrodynamic focusing V. Kachel, H. Fellner-Feldegg & E. Menke - MLM Chapt. 3

  5. Fluidics– Particle Orientation and Deformation V. Kachel, et al. - MLM Chapt. 3

  6. Sheath SampleStream Cell The Flow Cell The introduction of a large volume into a small volume in such a way that it becomes “focused” along an axis is called Hydrodynamic Focusing. Original from Purdue University Cytometry Laboratories, Modified by R. Duggan

  7. Sample differential pressure 10 psi 10 psi 10 psi 10.2 psi 10.4 psi 10.8 psi • If differential is too large, cells will no longer line up single file • Results in wider CV’s and increase in multiple cells passing through the laser at once. No more single cell analysis!

  8. What can we measure • Scattered light • Fluorescence Thats it

  9. Forward Angle Light Scatter Laser FALS Sensor Intensity is said to be proportional to cellsize

  10. 90 Degree Light Scatter Laser FALS Sensor Intensity is said to be proportional to Granularity or internal complexity of the cell 90LS Sensor

  11. Lymphozyt Basophiler Monozyt Neutrophiler Eosinophiler FACSAcademy

  12. Granulocytes Lymphocytes SSC Monocytes RBCs, Debris, Dead Cells FSC Why Look at FSC v. SSC • Since FSC ~ size and SSC ~ internal structure, a correlated measurement between them can allow for differentiation of cell types in a heterogenous cell population

  13. HO O C CO2H How to visualize a specific antibody fluorescein

  14. Fluorescence Stokes Shift is 25 nm Fluorescein molecule 520 nm 495 nm Fluorescnece Intensity Wavelength low energy high energy wavelength (nm)

  15. 350 457 488 514 610 632 Spectra of Common Fluorochromes Laser Lines (nm) PE-Texas Red Texas Red PI Ethidium PE FITC cis-Paranaric Acid 300 400 500 600 700 Original from Purdue University Cytometry Laboratories, Modified by R. Duggan

  16. Fluorescence Detectors Laser FALS Sensor Fluorescence detector (PMT3, PMT4 etc.)

  17. The photoreceptors of a flow cytometer longpass filter bandpass filter

  18. Flow Cytometry Optics dissection of fluorescence PMT 4 Dichroic Longpass filters PMT 3 Flow cell PMT 2 Bandpass Filters PMT 1 Laser

  19. Signal Detection - PMTsphoto multiplier tube Secondary emission Cathode Anode Amplified Signal Out Photons in End Window Dynodes Voltage

  20. 350 457 488 514 610 632 The color blindness of a flow cytometer Laser Lines (nm) PE-Texas Red Texas Red PI Ethidium PE FITC cis-Paranaric Acid 300 400 500 600 700 Original from Purdue University Cytometry Laboratories, Modified by R. Duggan

  21. Signal time Laser

  22. Signal Laser time

  23. Laser Signal time

  24. Pulse Width Pulse Area Pulse Height Measurements of the Pulse Voltage Intensity Time

  25. A mixed cell population

  26. One parameter display (Histogram)

  27. Two parameter display (Dot Plot) 320 FL2 480 FL1 FSC SSC FL1 FL2 FL3 FL4 75 42 320 10 10 9 FSC SSC FL1 FL2 FL3 FL4 20 41 7 480 8 8 FSC SSC FL1 FL2 FL3 FL4 20 41 320 480 8 8

  28. Two parameter display (Dot Plot) CD45PE CD14 FITC

  29. Häufigkeit Parameter2 Parameter1 dot plot Two parameters displayS 3D plot contour plot density plot

  30. Data Analysis Concepts Regions A certain characteristic (blond, little, man, woman) Gates Logical combination of characteristics (blond women, little men)

  31. Characterizationofsubpopulations Gate Auswertung CD45PE CD14FITC

  32. 11-color, 13-parameter flow cytometry: Identification of human naive T cells by phenotype, function, and T-cell receptor diversity S C De Rosa, L A Herzenberg, L A Herzenberg & M Roederer Nature Medicine 7, 245 - 248 (2001).

  33. http://www.cyto.purdue.edu/ Jean Paul Robinson