1 / 10

HBD Meeting 4/25/06

HBD Meeting 4/25/06. SB and BNL Crew. Installation of HBD into PHENIX. Transported HBD from USB under gas (CF4) flow (maintained ~ 1Torr overpressure for ~ 2-3hrs)

suzy
Download Presentation

HBD Meeting 4/25/06

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. HBD Meeting4/25/06 SB and BNL Crew

  2. Installation of HBD into PHENIX • Transported HBD from USB under gas (CF4) flow (maintained ~ 1Torr overpressure for ~ 2-3hrs) • Installation was difficult, but went smoothly…Don Lynch suggested the use of temporary lifting handles connected to mounting points during installation of final detector Positioning HBD into place Flash Lamp Also,15% Co Pb converter mounted over ½ area of GEM active area B. Azmoun

  3. HBD Hutch • Set up Shop in Space beneath South-side stairs in assembly hall • Ultimately this space is for the Transmission monitor • Some gas flow controls & H2O and O2 monitors & P/T Monitor • 2 signal cables + Pulser for PA + Flash lamp Trigger • Pad “lucky” #13  Fe55 (2uCi source)  Masked from light • Pad # 49  Open to light B. Azmoun

  4. Daily Checks • Absolute Gain @ 1 GEM voltage: DV = 495V • Gain consistent w/ USB measurements ~ 9x103 • Flash Lamp test (PD mode, gain mode) • PD Flash lamp tests also consistent: pulse ~ 520mV in PD mode • H2O & O2 • PPM levels seem to be a little higher at PHENIX than at USB…calibration error??? PHENIX USB B. Azmoun

  5. HBD Flash-lamp Event Display(pe signal avalanche through GEMs and map onto hex pads) Pad 13 shadowed by Fe55 Known dead strips 2 known dead strips, but there are two others w/ position unknown Sum along strips Sum perp. to strips Known dead strips Possible position of unknown dead strips B. Azmoun

  6. Flash Lamp: Photon Flux Calc’s • PD Mode: Entire PC (No Attenuator) • dI = 500mV/50ohm = 0.01C/sec • ½ * 0.01 C/sec * 1.2usec = 6nC = 3.75x1010pe • 3.75x1010pe / 0.3 (int. QE) =1.25 x1011g (entire PC) • 1.25 x1011g (PC) *[1920 (Pad 49)/73317(PC)] = 3.27x109 g (pad #49) • PC Lifetime…will the flash-lamp harm the PC? • Charge not to exceed 140uC/cm2 • 6nC/(25x25cm2) = 9.6pC/cm2 • Each test: ~2Hz for ½ hour = 3600 flashes • 3600 flashes * 9.6pC/cm2 = 34.6nC/cm2 (each test) • # of tests before PC degradation: (140uC/cm2)/(34.6nC/cm2) > 4000 tests • Gain Mode: Monitor single pad # 49 (GEM Attenuator) • PA signal (pad #49) = 1920mV @ Gain ~ 250 (DV = 427V) • PA calib.: 2000e-/mV • 1920mV * 2000e-/mV = 3.8x106 e-’s  3.8x106 e-’s / 250 = 1.5x104pe’s • 1.5x104pe’s / 0.3 int. QE = 5.12x104g (pad #49) • Attenuation Factor from 3GEM at Pad #49 • 3GEM Attenuator: 5.12x104g , No Attenuator: 3.27x109 g • Attenuation Factor: 3.2x109 / 5.12x104 = 6.25x104 • Attenuation per GEM: (6.25x104)1/3 = 39.7 (Probably not this simple) Flash Lamp Pulse (Noise Free) B. Azmoun

  7. Beam Test: Preparation Baseline • Noise (Baseline ~ 10-20mV p-p) • Occasional Oscillating pulses (Corona?) • PPM: H2O ~25ppm, O2 ~15ppm (total = 40ppm) Oscillating Pulses ~1pe/10ppm MIP/Fe55/ pe Pulses B. Azmoun

  8. MIPS in 100um gap (1.6e-) • Fe55 in 100um gap (109e-) • Other particles • MIPS in 3mm gap (46.6e-) • Fe55 in DG (109e-) • Other particles • MIPS in 3mm gap (46.6e-) • Cherenkov (1-20max pe/pad) • Other particles • MIPS in 100um gap (1.6e-) • Cherenkov (1-20max pe/pad) • Other particles Beam Test: First look at Signals (Qualitative Results) • First Beam this past Saturday (4/22/06) • Primary charge • MIP: 7kEv/cm * 3mm / (54eV e-ion pair) * (1.2 rel. rise) = 46.6 e- • Fe55 in CF4: 5.89keV x-ray/(54 eV/e-ion pair) = 109pe (Pad # 13 only) • Photoelectron yield (Cherenkov spectrum + CsI QE ): 36pe-s total ~15-20 pe per pad max (Pad # 13 masked off) • Rate of electrons  15% radiation length FB  DG=0.8kV/cm Loss of pe coll. Efficiency? B. Azmoun

  9. Fe55 Spectrum w/ HBD in IR: Beam Off • Peak of spectrum is centered on ~109pe as expected calibration check for previous measurements (assumed gain was 9.0x103 @ Vin= 3700V (DV=495V), DG = 0.8kV/cm ) B. Azmoun

  10. Outlook • Quantitative Beam tests to determine No (CF4) with HBD in “dead turtle” position • Method of normalizing and subtracting pad 49 spectra from pad 13 spectra • Continue to monitor ppm’s will increase flow rate if needed in order to improve gas purity. • Looking ahead: if opportunity permits (during short access), we’ll connect to and check electronics chain, and possibly take data. B. Azmoun

More Related