1 / 119

Mentor Corporation’s Silicone Gel-Filled Breast Implants (P030053)

Mentor Corporation’s Silicone Gel-Filled Breast Implants (P030053). April 13, 2005. FDA Presenters. CDR Samie Allen, USPHS Sam Arepalli, Ph.D. David Berkowitz, Ph.D., V.M.D. Herbert P. Lerner, M.D. Sahar Dawisha, M.D. Phyllis Silverman, M.S.

stesha
Download Presentation

Mentor Corporation’s Silicone Gel-Filled Breast Implants (P030053)

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Mentor Corporation’sSilicone Gel-Filled Breast Implants (P030053) April 13, 2005

  2. FDA Presenters • CDR Samie Allen, USPHS • Sam Arepalli, Ph.D. • David Berkowitz, Ph.D., V.M.D. • Herbert P. Lerner, M.D. • Sahar Dawisha, M.D. • Phyllis Silverman, M.S. Mentor P030053

  3. Device Description & Preclinical Testing Overview Samie Allen

  4. Device Description • High, moderate, & moderate plus profiles • Smooth & textured surfaces • Round • Single lumen • Components: shell, patch, filler, & silicone adhesive Mentor P030053

  5. Preclinical Testing • Modes and causes of rupture • Fatigue testing • Gel bleed • Gel cohesion • Shelf life Mentor P030053

  6. Modes & Causes of Rupture • Retrieval studies of explanted devices • Physical property / crosslink density testing • Assessment of manufacturing processes • Assessment of surgical techniques • Review of literature Mentor P030053

  7. Modes & Causes of Rupture (cont.) PE Database Analysis: • Includes failed, retrieved devices in 3 categories • Iatrogenic User Related – surgical damage • Rent Unknown Cause (RUC) • Not Apparent Etiology Unknown (NAEU) • Focused on RUC and NAEU categories • 203 available for analysis Mentor P030053

  8. Modes & Causes of Rupture (cont.) Mentor P030053

  9. Modes & Causes of Rupture (cont.) Mentor P030053

  10. Modes & Causes of Rupture (cont.) Mentor P030053

  11. Modes & Causes of Rupture (cont.) Mentor P030053

  12. Modes & Causes of Rupture (cont.) Mentor P030053

  13. Modes & Causes of Rupture (cont.) Mentor’s Proposed Next Steps: • In-vitro study to determine optimum incision size • Develop inserter instrument • Assess alternate texturing process • Investigate patch design changes • Labeling and physician training Mentor P030053

  14. Fatigue Testing • 20-30 lbs endurance load limit • Analyzed raw data • Estimated median life of 25-47 years • Appropriateness of test set-up and accuracy of estimate unknown based on lack of pure cyclic failures observed in modes and causes of rupture studies Mentor P030053

  15. Gel Bleed Testing • ASTM F703 testing • Gel loss analysis • Gel bleed study Mentor P030053

  16. Gel Bleed Testing (cont.) ASTM F703 Testing: • Smooth Moderate Profile implants (350cc) • Incubated for 15 (not 8) weeks at 110°F • Ave weight gain rate – 0.0011 g/cm2/week • ASTM F703 test method does not mimic in-vivo conditions Mentor P030053

  17. Gel Bleed Testing (cont.) Gel Loss Analysis: • 74 devices randomly selected and re-weighed to obtain post-explantation weight • Calculated % implant weight • 40 smooth; 0.2-15.2 in-vivo years; 101% (97-107) • 34 Siltex; 0.1-9.4 in-vivo years; 101% (99-104) • Does not identify and quantify the rate of bleed for all gel bleed constituents Mentor P030053

  18. Gel Bleed Testing (cont.) Gel Bleed Study: • 125cc implants incubated in 225ml porcine serum at 37°C • Detected D4, D5, and D6. Cumulative bleed rate was 0.95 ng/cm2/day by day 30 • Issues with testing: • Applicability of 120 days to in-vivo conditions • Implant size vs. volume of porcine serum medium • Decrease in release amount for D5 and D6 after day 30 • GC analysis correction due to volatility • No rate of diffusion for each gel bleed constituent Mentor P030053

  19. Gel Cohesion Testing • Gel cohesion testing of final gel • ASTM F703 (<4.5cm & no gel separation) • Results: passed • Penetrometer testing of in-process gel • No standard (internal specification) • Results: passed; 63.4 (range of 61-64) Mentor P030053

  20. Shelf Life • Device and package testing • 5-year shelf life date on package label Mentor P030053

  21. Summary – Preclinical Testing • Modes and Causes of Rupture: • Characterize through 10 years • Not predictive of lifetime rupture rate • Proposed labeling and training to address failures related to surgical procedure • Proposed several design/manufacturing changes • Proposed to develop introducer instrument • Fatigue Testing: • Adequacy of testing and estimate lifetime cannot be validated Mentor P030053

  22. Summary – Preclinical Testing (cont). • Gel Bleed: • Outstanding issues that should be able to be addressed by sponsor • Gel Cohesion: • Adequate to address issue • Shelf Life: • Adequate to support 5-year shelf life on label Mentor P030053

  23. Chemistry Overview Sam Arepalli, Ph.D.

  24. Device Materials • Shell, middle (barrier) layer: Diphenyldimethyl-siloxane copolymer, 15 mole% diphenyl • Shell, inner/outer (base) layers: Dimethylsiloxane polymer • Patch Assembly: Dimethylsiloxane Pt (platinum) cure polymer • Silicone Gel: Two-part platinum cure gel • Silicone adhesive: RTV silicone Sn (tin) cure Mentor P030053

  25. Degree of Crosslinking • Shell: 7.90 chains/cm3 (Sol Fraction Method) • Gel: • 8.81 chains/cm3 • 3.0-10.0 mm (Penetrometer) Mentor P030053

  26. Volatiles • Shell: total volatiles (11.1 ppm) • IPA (1 ppm) • Xylenes (0.1 ppm) • Methoxymethylsilane (3 ppm) • Dodecane (3 ppm) • Undecane (1.3 ppm) • Gel: total volatiles (2.8 ppm) • D3 (0.18 ppm) • D4 (0.5 ppm) • D5 (1.6 ppm) • Undecane (0.3 ppm) Mentor P030053

  27. Extractables • Gravimetric analysis • Gel permeable chromatography • FTIR analysis • Qualitative and quantitative analysis (GC-MS Analysis) Mentor P030053

  28. GC-MS Analysis • Shell and gel • LMW oligosiloxanes (up to D10) present at<10ppm • High MW cyclic (>D11) oligosiloxanes concentrations comparable to those of saline-filled breast implants. Mentor P030053

  29. Metal Analysis • Extracted residue: • Shell: Sn (ND); Pt (0.133 ppm) • Gel: Sn (ND); Pt (0.323 ppm) • Unextracted: • Shell: Sn (0.03 ppm); Pt (8.8 ppm) • Gel: Sn (0.01 ppm); Pt (4.8 ppm) Mentor P030053

  30. Silica Analysis • Amorphous silica (X-ray diffraction) • No free silica present (Raman & Photoelectron spectroscopy) Mentor P030053

  31. Summary - Chemistry • Shell and gel tested separately • Degree of crosslinking • Volatiles • Metals • Extractables • Gravimetric analysis • GPC • FTIR • GC-MS Mentor P030053

  32. Toxicology Overview David B. Berkowitz, Ph.D., V.M.D.

  33. Five Testing Categories • Pharmacokinetics • Biocompatibility • Immunotoxicology • Reproductive and Teratogenicity Testing • Genotoxicity and Carcinogenesis Mentor P030053

  34. Pharmacokinetics • Elastomer: Orthopedic implants in dogs and humans could be recovered after years. • Gel: Long-term gel implants in rats remained at subcutaneous implantation sites for at least 450 days. 99.97% of 14-C-labeled gel remained in place in mice for at least 56 days. • LMW Gel Components: Subcutaneous implants of low molecular weight cyclic siloxanes (D3-D7) were distributed to tissues over a year. Mentor P030053

  35. Biocompatibility Testing • Cytotoxicity • Irritation and Short-Term Implantation • Acute Systemic Toxicity • Hemolysis • Pyrogenicity Mentor P030053

  36. Immunotoxicity Testing • Sensitization Testing • Other Immunotoxicity Testing Endpoints • Body, Spleen, and Thymus Weights • Hematology • Splenic T-Cells (CD4+and CD8+) • T-Cells Response to Mitogens • Mixed Lymphocyte Response • IgM Antigen Forming Cells Mentor P030053

  37. Reproductive & Teratogenicity Testing • A 1-Generation Study • 100 F0 Females – 4 groups of 25 each • No Reproductive Effects in F0 • No Reproductive or Teratogenic Effects in F1 Mentor P030053

  38. Genotoxicity & CarcinogenesisTesting • Salmonella Reverse Mutation Assay • Unscheduled DNA Synthesis • Chromosomal Aberration Assay • In Vivo Mouse Micronucleus Test • Carcinogenicity Testing Mentor P030053

  39. Toxicology Summary • Mentor has provided the toxicology information recommended in the breast implant guidance document. • No safety issues were raised by data. Mentor P030053

  40. Clinical Data Overview Herbert P. Lerner, M.D. Medical Officer

  41. Summary of Studies • Core Study—Started 2000. • Adjunct Study—Started 1992. All open label, prospective, multicenter. • Yearly F/U in Core Study. • Both collected local complications. Mentor P030053

  42. Core Study • Majority of Safety and Effectiveness data. • Augmentation, Reconstruction, Revision. • Yearly F/U to 10 years after implantation. • Study includes prospective MRI screening for silent rupture in 420 of 1007 patients. • QOL and CTD signs/symptoms collected. Mentor P030053

  43. Adjunct Study • Intended to make the implants available for reconstruction and revision patients. • Collected local complications at 1, 3, and 5 years after implantation surgery. • No MRI Screening. • Unlimited sample size. • Enrollment is ongoing. Mentor P030053

  44. Core Study Results Mentor P030053

  45. Core Study Demographics: Age Mentor P030053

  46. Core Augmentation Cohort Mentor P030053

  47. Patient Disposition - Core Aug • 551 patients (1110 devices) enrolled. • 439 (80%) theoretically due at 3 years. • 94% patient follow-up rate at 3 years. • 0 deaths • 22 patients with implant removals/replacements • 9 lost to follow-up Mentor P030053

  48. By-Patient 3-Year Cumulative KM Complication Rates - Core Aug Mentor P030053

  49. Reoperations - Core Augmentation • 160 additional surgical procedures in 98 reoperations in 79 patients. • Primary reasons for reoperation: • Capsular Contracture - 44% • Patient Request – 32% • Types of additional surgical procedures • Capsule procedures - 36% • Implant removal/replacement – 28% Mentor P030053

  50. Primary Procedure for Given Reoperation Mentor P030053