Cryogenic Optical Microscope. Faculty Advisor: Prof. Greg Kowalski Sponsors: Dale Larson , James Hogle, Ph.D. (Harvard Medical School) Design Team : Mohammad Ali John Delcore Sarah Kaufmann David Rezac. Problem Statement.

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about ' Cryogenic Optical Microscope' - scott-montoya


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
cryogenic optical microscope

Cryogenic Optical Microscope

Faculty Advisor: Prof. Greg Kowalski

Sponsors: Dale Larson , James Hogle, Ph.D.

(Harvard Medical School)

Design Team:

Mohammad Ali

John Delcore

Sarah Kaufmann

David Rezac

problem statement
Problem Statement
  • Infected, frozen sample to be analyzed in TEM
    • Very high resolution
    • Required to better understand cell behavior
  • Imaging complications
    • Small field of view leads to lots of time searching
    • Electron bombardment
    • Congested area around cell

(Source: www.brockhouse.mcmaster.ca)

problem statement1
Problem Statement
  • Infected, frozen sample to be analyzed in TEM
    • Very high resolution
    • Required to better understand cell behavior
  • Imaging complications
    • Small field of view leads to lots of time searching
    • Electron bombardment
    • Congested area around cell`

(Source: www.lifesci.ucsb.edu)

slide4

Problem Statement (cont.)

  • It is necessary to first image the sample in an optical microscope (OM) and identify areas of interest
  • Currently, frozen samples cannot be viewed in the OM

(Source: webphysics.davidson.edu)

design requirements
Design Requirements
  • Maintain specimen below -140°C
  • Provide means to image the sample (microscope)
  • Isolate sample from significant vibrations
  • Protect sample grid from stresses that may cause deformation
  • Prevent contamination by water contact (condensation)
impact statement
Impact Statement
  • Aid research and development for improved therapeutic advancements
  • Improve quality and quantity of TEM images
  • Enable microscopists
    • Label molecular components in OM
    • Analyze with high resolution of TEM
  • Requires no additional expertise
    • Familiar operations for microscopists
slide7

Why below -140oC?

  • The sample is embedded in vitreous ice
    • Stable below -140oC
  • Vitreous ice is the “glassy” amorphous solid form of water
    • Does not scatter electrons
    • Low vapor pressure

(Source: www.nims.go.jp)

slide8

Optical Microscope

  • Numerical Aperture (NA)
    • Indicates the resolving power of the lens
    • Larger NA = better resolution
    • Inversely related: WD NA

(Source: www.microscopyu.com)

slide9

Optical Microscope

  • Numerical Aperture (NA)
    • Indicates the resolving power of the lens
    • Larger NA = better resolution
    • Inversely related: WD NA

(Source: www.microscopyu.com)

slide10
Design #1

Preliminary Designs

Design #2

Design #3

slide11

Thermal Modeling

TGU=16oC

Design #3

slide12

Cold Temperature Objective Lens

  • Able to function at LN2 temperature
  • Manufactured by Microthek Corp. in Germany
  • 80x magnification
  • 0.8 Numerical Aperture
  • .96 mm working distance
  • Withstands cyclic testing to liquid helium temperatures
slide13

Design Evolution

Window

(Image Path)

Cold Lens

Position

LN2 Level

Vacuum Chamber

Inner Skirt to

Hold N2 Gas

Post to Hold Sample

for Imaging

Outer Housing

slide14

TEM Cryo-Transfer Apparatus

  • Cold Finger
  • LN2 Workstation
slide15

Current Design

Upper Assembly

(Contains Objective Lens)

Cold Finger

Work Station

slide18

Current Design

LN2 Reservoir

slide19

Current Design

Sample

Position

slide22

Boiling

CHF

(Source: www.alamthermal.com)

(Source: www.spaceflight.esa.int)

(Source: www.nuc.berkeley.edu)

slide23

Boiling

  • qin=9.47W
  • SS in 116 minutes
  • 1.98kg LN2
  • qin=3.08W
  • SS in 26 minutes
  • 0.075kg LN2
slide24

Boiling

  • qin=9.47W
  • SS in 116 minutes
  • 1.98kg LN2
  • qin=3.08W
  • SS in 26 minutes
  • 0.075kg LN2
slide25

Testing

Complete Assembly in

Imaging Position

External Nitrogen

Flush

Thermocouples in

Key Locations

slide26

Testing

Test #1

Test #2

Pre-Cooled

Copper

=Tsample

Test #3

Test #4

slide27

Testing

Test #1

Test #2

Pre-Cooled

Copper

=Tsample

Test #3

Test #4

slide28

Testing

Test 1

Test 2

Test 3

Test 4

Imaging

Cutoff

slide29

Microscope

Aberrations

Partial Focus

Focused: 2.7µ

10µ Line Spacing

(filters omitted)

Special Thanks to:

Antoine van Oijen

Resolution Target

slide30

Current Prototype

  • Allows for imaging in the optical microscope
    • Stable cold environment maintains specimen below -140°C for 10-15 minutes
  • Provides imaging resolution down to 2.7µ
  • Incorporates TEM cold finger into the design
    • Protects sample from additional stresses or possible water contamination
    • Saves time and effort
  • Whole system is on a 1,000 lb optical table
    • Reduces vibrations
slide31

Recommendations

  • Continue design stage with thermal mass concept
  • Refine microscope operation
  • Reduce footprint and isolate optical components
  • Function specific improvements:
    • Automate XY and Z stages
    • Purchase camera suited to application