Opening your mind to vibrational spectroscopy
This presentation is the property of its rightful owner.
Sponsored Links
1 / 41

Opening your Mind to Vibrational Spectroscopy PowerPoint PPT Presentation

  • Uploaded on
  • Presentation posted in: General

Opening your Mind to Vibrational Spectroscopy. Rodolfo J. Romañach, Ph.D. UPR-Mayagüez [email protected] May 14, 2012. Regions of Infrared Spectrum: Far Infrared: 650 – 25 c m -1 Mid Infrared: 4000 – 650 c m -1 wavelengths from 2.5 m to 25 m).

Download Presentation

Opening your Mind to Vibrational Spectroscopy

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.

- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript

Opening your mind to vibrational spectroscopy

Opening your Mind to Vibrational Spectroscopy

Rodolfo J. Romañach, Ph.D.


[email protected]

May 14, 2012

Infrared vibrational spectroscopy

Regions of Infrared Spectrum:

Far Infrared: 650 – 25 cm-1

Mid Infrared: 4000 – 650 cm-1 wavelengths from 2.5 m to 25 m).

Near Infrared: 12800 – 4000 cm-1(0.8 - 2.5 m, or 800- 2500 nm).

Absorption occurs when the dipole moment of the molecule changes, and the molecule interacts with radiation equal to the frequency of vibration.

Infrared (Vibrational) Spectroscopy

Change in dipole moment during molecular vibrations







Change in Dipole Moment during Molecular Vibrations

  • Must change for IR absorption to occur.

  • The dipole moment is a measure of the degree of polarity of molecule (magnitude of the separated charges times the distance between them).

  • A measurement of degree of unequal distribution of charges in molecule.



Band intensity in ir

Band Intensity in IR

  • Band intensity depends on the rate of change of dipole moment during absorption of IR light.

  • Stronger bands occur when the change in dipole moment is greatest.

Infrared spectroscopy mid ir near infrared spectroscopy

Infrared Spectroscopy (mid-IR) & Near Infrared Spectroscopy

  • Infrared Spectroscopy (mid-IR), studied in organic chemistry courses; principal identification method in pharmaceutical industry.

  • Mid-IR sharp bands that provide structural information on compounds; better than fingerprint.

  • Near IR weak overlapping bands; difficult to interpret (region of 12500 to 4000 cm-1, wavelengths from 0.8 to 2.5 m.)

Mid i r spectroscopy widely used in

Mid-IR Spectroscopy Widely Used in:

  • Identification of Pharmaceutical Raw Materials and Finished Products. Principal method of material identification in a pharmaceutical manufacturing company.

  • Combination with MS and NMR to determine structure of process impurities and degradation products.

  • Characterization of natural products, use of GC/FT-IR.

  • Forensic Analysis, IR-Microscopy.

  • Environmental Analysis: GC/FT-IR.

  • Surface Analysis, Diffuse Reflectance, Attenuated Total Reflectance, Grazing Angle.

  • Studies of Protein Structure and Dynamics.

Group frequencies

Group Frequencies

  • Characteristic of functional groups such as -O-H, -CH3, -COCH3, and COOH.

  • Determined empirically by studying the spectra of many related molecules.

  • Always found in the spectrum of a molecule containing that group, and always occurs in the same narrow frequency range.

  • The form of the bands is nearly always the same in every molecule containing that group.

Dividing line

Dividing Line

  • 1500 cm-1 dividing line: above 1500 cm-1 if a band has a reasonable intensity, it is a group frequency.

  • Below 1500 cm-1 the band may be either a group or fingerprint frequency.

  • Below 1500 cm-1 called the fingerprint region.

Opening your mind to vibrational spectroscopy

Typical Identification in Pharmaceutical Industry with Mid-Infrared Spectroscopy

Opening your mind to vibrational spectroscopy

Checking Fingerprint Region.

Mid ir spectroscopy

Mid-IR Spectroscopy

+ Each compound has a unique mid-IR spectrum. Excellent for identification. Key technique for identifying raw materials in pharmaceutical industry.

+ Information rich technique, spectra can be interpreted – learning that you have an organic compound, carbonyls, aromatic compounds, etc.

- Requires sample preparation, except when attenuated total reflectance is used.

Can your hplc analyze identify this sample in 1 minute

Can your HPLC analyze & identify this sample in 1 minute?

Sample from Dr. Lysa Chismadia’s collection, RUM Geology Department.

Advantages of nirs

Advantages of NIRS

  • Sample preparation is not required leading to significant reductions in analysis time. Waste and reagents are minimized (non-destructive testing).

  • Excellent analytical method for the study of solids. Capable of ID based on both chemical & physical properties.

  • Spectra may be obtained in non-invasive manner.

  • Remote sampling is possible (good for hazardous materials). NIR radiation can be transmitted through silica fiber optics that are relatively inexpensive.

Spectroscopy of solids

non-invasive spectral acquisition & remote sampling

Opening your mind to vibrational spectroscopy

Non-invasive monitoring. Opportunity for following processes.

CDI Lab Scale NIRS system,

Opening your mind to vibrational spectroscopy

IR Source


IR Energy

Delivery Fiber Bundle

Collection Fiber Bundle

Reflected IR Energy


Powder and Solids Probe – Courtesy Bruker Optics

Diffuse Reflection Probe Schematic

Powder & Solids Probe with liquid attachment

Extra-long immersion

depth: 12”

Spectroscopy of the solid state

Spectroscopy of the Solid State

  • Spectroscopy - Interaction between radiation and matter.

  • Spectra – pattern that indicates absorbance or reflection of radiation as a f(λ) or f(ν).

  • NIR – offers possibility of study of interaction of solids with radiation since sample preparation is not required.

Need to visualize interaction between particle and radiation.

May also be used for liquids, but majority of applications for solids avoiding sample preparation.

Characterized by wide bands

Characterized by Wide Bands

Wide bands - 12,000 – 4,000 cm-1, or 800 – 2500 nm

Overtones & Combination Bands of O-H, N-H and C-H groups

Nir fundamentals electromagnetic spectrum

NIR Fundamentals: Electromagnetic Spectrum

12,500 cm-1 (800 nm)

4,000 cm -1 (2500 nm)




X – Ray











Radio, TV Waves


InnerShell Electronic






SpinOrientation in MagneticField




Courtesy Bruker Optics

Advantages of nirs1

Advantages of NIRS

  • Possibility of using it in a wide range of applications (physical and chemical), and viewing relationships difficult to observe by other means.

  • The spectrum may be used to identify the formulation and also to quantify the drug in the formulation.* Cross-sensitivity.

*M. Blanco, J. Coello, A. Eustaquio, H Iturriaga, and S. Maspoch, Development and Validation of a Method for the Analysis of a Pharmaceutical Preparation by Near-Infrared Diffuse Reflectance Spectroscopy, Journal of Pharmaceutical Sciences, 1999, 88(5), 551 – 556.

Cautions in working with nirs

Cautions in Working with NIRS

  • Overlapping bands, not easy to interpret. Differences in spectra are often very subtle.

  • May confuse chemical & physical effects.

  • Calibration requires careful experimental design.

  • Depends on accuracy of reference methods.

  • Usually not for trace level analysis.

  • Implementation of NIR requires a significant investment in Human Resources. Not an HPLC !! Over 60% of instruments installed in pharmaceutical companies are never used.

Nir absorption bands

NIR Absorption Bands

  • Absorption bands in the NIR are the result of combination and overtone bands from the fundamental vibrations of C-H, N-H, and O-H bonds seen in the mid-IR.

  • The overtone and combination bands are 10 – 100 X less intense than the fundamental bands in mid-IR. (Do not confuse energy of vibration with intensity of band).

  • Differences in spectra are usually very subtle. Instruments have a high signal to noise ratio.

Mir vs nir oleic acid

Oleic Acid




Wavenumber cm-1

MIR vs. NIR: Oleic Acid

NIR Bands will be 10 – 100 x weaker than mid-infrared bands

Slide Courtesy Bruker Optics

Opening your mind to vibrational spectroscopy

Fundamentals and Overtones

In the case of the anharmonic oscillator, the vibrational transitions no longer only obey the selection rule n = 1. This type of vibrational transition is called fundamental vibration. Vibrational transitions with n = 2, 3, ... are also possible, and are termed overtones. Called first, second, and so on, overtones.

Subtle differences in spectra

Subtle Differences in Spectra

Opening your mind to vibrational spectroscopy

Local Mode Effect

Subtle differences valuable info

Subtle Differences,Valuable Info.

M. Blanco, D. Valdés, I. Llorente, and M. Bayod, “Application of NIR Spectroscopy in Polymorphic Analysis: Study of Pseudo-Polymorphs Stability”, Journal of Pharmaceutical Sciences, 2005, 94(6), 1336 – 1342.



  • NIR spectra are difficult to interpret and differences in spectra are very subtle, yet it is able to discriminate between spectra of very similar molecules (although the differences are not as easy to discern as in mid-IR).

  • Able to provide both chemical and physical information on a samples in the solid state.

  • Must work cautiously to avoid confusing chemical and physical information.

Developing a calibration model

Developing a Calibration Model

  • Most NIR calibration models are multivariate. The absorbance at multiple wavelengths or frequencies are mathematically related to an analyte concentration or physical property.

  • Multivariate regression models like MLR, PLS are used, unlike the univariate linear least squares method used in most analytical chemistry.

Opening your mind to vibrational spectroscopy

O-H first overtone

Variation Implies Information !!

However, variation could come from differences in moisture content (chemical info) or variation in particle size, porosity, density (physical info).

Absorbance, 1st-derivative, and 2nd derivative spectra.

X. Zhou, P. Hines, and M.W. Borer, “Moisture Determination in a Hygroscopic drug Substance by Near Infrared Spectroscopy”, Journal of Pharmaceutical and Biomedical Analysis, 17(1998), 219-225.

Opening your mind to vibrational spectroscopy

O-H combination band

X. Zhou, P. Hines, and M.W. Borer, “Moisture Determination in a Hygroscopic drug Substance by Near Infrared Spectroscopy”, Journal of Pharmaceutical and Biomedical Analysis, 17(1998), 219-225.

Absorbance, 1st-derivative, & 2nd derivative spectra.

Developing a chemical quantitative method

Developing a Chemical Quantitative Method

  • Used Partial Least Squares (PLS) regression to relate changes in water content to changes in the NIR spectra.

  • PLS is an example of a multivariate method that uses many responses from the spectrum and relates them to an analytical property.1

  • NIR spectra will include variation due to particle size and other physical effects. If you want to measure a chemical property such as concentration then you need to make sure that you are not confusing physical effects with chemical changes. – reason for pretreatment.

1- the methods that we study in analytical chem. class are still univariate)

Developing a chemical quantitative method1

Developing a Chemical Quantitative Method

  • Differences in particle size lead to differences in baseline.

  • Baseline is not related to concentration.

  • May eliminate baseline with 1st or 2nd derivative.

  • The changes observed are more related to chemical changes after the spectral pretreatment.

Opening your mind to vibrational spectroscopy

Raw spectra (1100-2205nm) obtained from ribbons of MCC 200 produced at 15 (straight), 25 (dash), and 45 bars (dot).

Spectra subtracted to zero at one λ (baseline corrected).

However, keep effect of pressure on spectrum to develop a NIR model for density.

D, Acevedo, A. Muliadi ,A. Giridhar, J.D. Litster, R. J. Romañach, AAPSPharmscitech, 2012, 13(3), 1005 – 1012, DOI: 10.1208/s12249-012-9825-0.

Comparison of sampling volume

Comparison of Sampling Volume

Systematic error in the relationship between the optical (NIR) method and the reference method.

First 2 mm sampled by NIR beam vs.

Entire Sampled Analyzed by HPLC or UV method

Calibration with error

Calibration with Error

Three types of error:

  • Random error in the reference laboratory values

  • Random error in the optical data

  • Systematic error in the relationship between these two. (e.g. Differences in the sample size of the two methods – sampling error)

From: Principles and Practice of Spectroscopic Calibration – H. Mark, John Wiley & Sons, 1991, p. 17

Opening your mind to vibrational spectroscopy

The NIR radiation interacts with about 260 mg of sample, the acquisition time is about 0.5 seconds, RMSEP ≈ 0.34 % (w/w).

A.U. Vanarase, M. Alcalà, J.I. Jerez Rozo, F.J. Muzzio and R.J. Romañach, “Real-time monitoring of drug concentration in a continuous powder mixing process using NIR spectroscopy, Chemical Engineering Science, 2010, 65(21), 5728 – 5733.

  • Login