Enhanced myocardial efficacy and removal by aspiration of liberated debris
This presentation is the property of its rightful owner.
Sponsored Links
1 / 6

Enhanced Myocardial Efficacy and Removal by Aspiration of Liberated Debris PowerPoint PPT Presentation


  • 71 Views
  • Uploaded on
  • Presentation posted in: General

EMERALD Trial. Enhanced Myocardial Efficacy and Removal by Aspiration of Liberated Debris. Presented at American College of Cardiology Scientific Sessions 2004 Presented by Dr. Gregg W. Stone. EMERALD Trial. 505 patients presenting within six hours of chest pain onset

Download Presentation

Enhanced Myocardial Efficacy and Removal by Aspiration of Liberated Debris

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript


Enhanced myocardial efficacy and removal by aspiration of liberated debris

EMERALD Trial

Enhanced Myocardial Efficacy and Removalby Aspiration of Liberated Debris

Presented at

American College of Cardiology

Scientific Sessions 2004

Presented by Dr. Gregg W. Stone


Emerald trial

EMERALD Trial

505 patients presenting within six hours of chest pain onset

with ≥2 mm ST elevation or left bundle branch block

and planned use of primary or rescue PCI

Randomized, open-label, multicenter

Control

No distal protection

n=246

GuardWire Distal Protection

0.028” GuardWire Plus System

n=252

  • Endpoints:

    • ST resolution 30 minutes post-procedure

    • Infarct size assessed by Tc-99m sestamibi imaging at days 5-14

Presented at ACC Scientific Sessions 2004


Enhanced myocardial efficacy and removal by aspiration of liberated debris

EMERALD Trial

MACE at 30 Days

p=0.79

  • Visible debris was removed in 73% of patients in the GuardWire arm.

  • There was a significantly higher number of diabetic patients in the control arm (17.3% vs. 7.5%, p=0.001).

  • Use of GP IIb/IIIa inhibitors was 82.9% and 83.5% in the GuardWire and Control arms, respectively.

  • Major adverse cardiac events (MACE) at 30 days was similar in both groups, as were the components of MACE: death (2.1% vs. 2.9%, p=0.77), heart failure (2.1% vs. 0.4%, p=0.12), new onset hypotension (11.1% each), and readmission for left ventricular failure (0.8% vs. 0%, p=0.25).

%

Presented at ACC Scientific Sessions 2004


Enhanced myocardial efficacy and removal by aspiration of liberated debris

EMERALD Trial

Post-Procedure

TIMI Frame Count

p=0.48

Post-Procedure

TIMI Grade 3 Flow

p=0.36

Post-Procedure

Myocardial Blush 3

p=0.13

%

Guard

Wire

Control

Guard

Wire

Control

Guard

Wire

Control

  • Post-procedure epicardial flow did not differ between groups, as assessed using TIMI grade 3 flow or TIMI frame count. Myocardial blush grade 3 trended higher in the GuardWire arm.

Presented at ACC Scientific Sessions 2004


Enhanced myocardial efficacy and removal by aspiration of liberated debris

EMERALD Trial

Mean ST Resolution

p=0.39

Complete ST Resolution (≥70%) 30 Minutes Post-Procedure

p=0.77

Infarct Size

p=0.09

%

Guard

Wire

Guard

Wire

Guard

Wire

Control

Control

Control

  • The frequency of complete ST resolution (≥70%) 30 minutes post-procedure did not differ between the treatment arms, nor did mean ST resolution. The other coprimary endpoint, infarct size, also did not differ between treatment groups.

Presented at ACC Scientific Sessions 2004


Enhanced myocardial efficacy and removal by aspiration of liberated debris

EMERALD Trial

  • Among patients with STEMI, use of the distal protection GuardWire device was not associated with an improvement in the primary endpoints of post-procedure ST resolution or final infarct size.

  • Despite the capture of visible debris in 73% of patients in the GuardWire arm, no significant measures of myocardial perfusion were observed, with only a trend toward higher rate of myocardial blush grade 3.