Helium D istribution in SM18 RF H orizontal Test Bunker M9 PowerPoint PPT Presentation


  • 104 Views
  • Uploaded on
  • Presentation posted in: General

Helium D istribution in SM18 RF H orizontal Test Bunker M9. T. Koettig, BE-RF-SRF 1 st Aug 2013. Content. Layout of the SM18 RF test area New cryogenic infrastructure => commissioning is ongoing M9 clients and requirements Proposal for the new Interconnection Valve B ox .

Download Presentation

Helium D istribution in SM18 RF H orizontal Test Bunker M9

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript


Helium d istribution in sm18 rf h orizontal test bunker m9

Helium Distribution in SM18 RF

Horizontal Test Bunker M9

T. Koettig, BE-RF-SRF 1stAug 2013


Helium d istribution in sm18 rf h orizontal test bunker m9

  • Content

  • Layout of the SM18 RF test area

  • New cryogenic infrastructure => commissioning is ongoing

  • M9 clients and requirements

  • Proposal for the new Interconnection Valve Box


Helium d istribution in sm18 rf h orizontal test bunker m9

Cryo Infra Upgrade of SM18 RF TF

RF TEST FACILITY AREA

V3

V4

V5

V6

He Guards

TRANSFER LINE GALLERY

M7

M9

LP/RECUP return

The 6 kW Linde cold box

Capacity: 25 g/s (720 l/h) of LHe

6 kW Linde Cryoplant

25 m3 Dewar

1.6 bara sat. LHe

to WPU2


Helium d istribution in sm18 rf h orizontal test bunker m9

Cryo Infra Upgrade: 3D-layout

M7

2 K

M9

2 K - 4.5 K

V3

2 K

V4

2 K

V5

4.5 K

V6

4.5 K


Helium d istribution in sm18 rf h orizontal test bunker m9

SM18 RF TF : General PID for 2013

M9 - Service module

  • Warm pumping unit capacity:

    • 6 g/s @ 10 mbar

    • 12 g/s @ 20 mbar

    • 18 g/s @ 30 mbar

Very low pressure

heater:

32 kW (20 g/s)


Helium d istribution in sm18 rf h orizontal test bunker m9

  • Requirements of customers in M9

  • LHC temporary connection with semi rigid bayonets:

  • LHe supply and GHe return, 4.5 K operation, 80 % reused later.

  • Hie-Isolde CM

  • special procedures for cool down and warm up,

  • 4.5 K operation + shield cooling at 50 K.

  • Two Venturi flow meters

  • CRAB cavity CM operation at T< 2 K, shield maybe at 80 K?

  • Helium guard, Pressure relief devices: SV at 1.8 bara, RD at 2.2 bara

  • Functional specs for He distribution in M9 can be found in EDMS:

  • https://edms.cern.ch/document/1279849/1


Helium d istribution in sm18 rf h orizontal test bunker m9

  • PID M9 helium distribution


Helium d istribution in sm18 rf h orizontal test bunker m9

  • M9 detail: new Interconnection Valve Box


Helium d istribution in sm18 rf h orizontal test bunker m9

  • Test schedule in SM18 RF : horizontal bunkers

  • Projects:

  • 1 test @ 4.5K of LHC cryo module . Provisional schedule : 2013

  • 2 tests @ 4.5 K of HIE-Isolde cryomodule. Provisional schedule : 2014

  • 1 test @ 1.8 K of SPL half cryomodule. Provisional schedule : 2014

  • x tests @ 2K of crab cavity CM. Provisional schedule : 2nd half of 2015


Helium d istribution in sm18 rf h orizontal test bunker m9

  • Thank you very much for your attention


Helium d istribution in sm18 rf h orizontal test bunker m9

Venturi Flow meter:

Cavity circuit at 4.5 K vapor

Shield circuit at 75 K gas


Helium d istribution in sm18 rf h orizontal test bunker m9

Venturi Flow meter

Cavity circuit at 4.5 K vaporShield circuit at 75 K gas


Helium d istribution in sm18 rf h orizontal test bunker m9

SM18 RF TF : bunkers PID for 2013

QLKRF_M7 @ 1.8 K

QLKRF_M9 @ 1.8 K – 4.5 K