Validation of a quantitative analytical procedure accuracy total error profile
Download
1 / 53

Validation of a Quantitative Analytical Procedure – Accuracy (total error) profile - PowerPoint PPT Presentation


  • 149 Views
  • Uploaded on

Validation of a Quantitative Analytical Procedure – Accuracy (total error) profile. Federaal Agentschap voor de Veiligheid van de Voedselketen. Dr. Jacques O. DE BEER Workshop IPH 27th April 2007 Scientific Institute of Public Health - Brussels (Belgium).

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about ' Validation of a Quantitative Analytical Procedure – Accuracy (total error) profile' - nuwa


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
Validation of a quantitative analytical procedure accuracy total error profile

Validation of a Quantitative Analytical Procedure – Accuracy (total error) profile

Federaal Agentschap voor de Veiligheid van de Voedselketen

Dr. Jacques O. DE BEER

Workshop IPH 27th April 2007

Scientific Institute of Public Health - Brussels (Belgium)


Method validation general concepts
Method Validation – General Concepts

  • Different regulations relating to GLP, GMP, GCP (OECD, EU)

  • Normative or Regulatory documents (ISO 17025, ICH, EMEA, FDA, dir. 2002/657/EG)

    both suggest that analytical procedures have to comply to certain acceptance criteria.

  • This request imposes that these procedures are to be validated.

    - Some documents define the validation criteria

    - No proposals on experimental approaches !!

    - Limited to general concepts !!


Introduction definition
Introduction - Definition

  • Method Validation: is the confirmation by examination and the provision of objective evidence that the particular requirements for a specific intended use are fulfilled. EN ISO/IEC 17025 § 5.4.5.1

  • Methods need to be validated or revalidated:

  • before introduction into routine application

  • whenever conditions change for which the method has been validated (e.g. Instrument with different characteristics)

  • whenever the method is modified and modifications are outside original scope of the method.


European and international regulatory bodies and their guidelines on different aspects of qa
European and International regulatory bodies and their guidelines on different aspects of QA


Objectives of an analytical procedure
Objectives of an analytical procedure guidelines on different aspects of QA

  • Able to quantify as accurately as possible each unknown quantity to be determined.

  • After analysis: the difference between returned result x and the unknown “true value µT” be small or < acceptance limit λ:

    - < x - µT < λ  x - µT  < λ (eq.1)

  • λ : depends on objective of analytical procedure e.g. 1-2 % on bulk, 5 % on pharmaceuticals, 15 % for biological samples previously defined


Objectives of an analytical procedure1
Objectives of an analytical procedure guidelines on different aspects of QA

  • Analytical procedures characterized by (cfr. def.):

  • “true bias”dM = systematic error (unknown)

  • “true precision”σ²M = random error measured by a standard deviation or variance (unknown)

  • Estimates of bias and precision obtained by experiments during the validation

  • Reliability of these estimates depends on adequacy of experiments on known samples (Valid. Stds),experimental design, number of experiments

  • These estimates  an intermediary but obligatorystep to evaluate if procedure is likely or not to quantify with sufficient accuracy the unknown quantities; not objectives per se


Examples of procedures having the same guidelines on different aspects of QAacceptance limits l = ± 15%

Procedure 1

Procedure 2

Bias = +7 %

RSD = 3%

Bias = +1 %

RSD = 8%

Procedure 3

Procedure 4

Bias = 0 %

RSD = 20%

Bias = +7 %

RSD = 12%


Objectives of an analytical procedure2
Objectives of an analytical procedure guidelines on different aspects of QA

  • Figure: 4 different (hypothetical) methods giving the distribution of 95% of the measures

  • Each method has a true bias dM , a true precision σ²M, a common acceptance limit λ (= 15%  bioanalytical procedure):

  • Procedure 3: negligible bias (0%); unsatisfactory precision (20% CV); too many measures beyond +/- 15% of the true value; does not fulfill objective

  • Procedure 4:bias(7%); precision(12%); important proportion outside acceptance limits; does not fulfill objective; but both < 15%: required by Washington Conf.

  • Procedures 1 and 2: fulfill (valid): at least 95% of results inside acceptance limits


Objectives of an analytical procedure3
Objectives of an analytical procedure guidelines on different aspects of QA

  • Procedure 1: presents a bias (+ 7%), but is very precise (3% CV)

  • Procedure 2: presents a negligible bias (+ 1%), but is less precise (8% CV)

    FIRST CONCLUSION:

    Differences between these two procedures don’t matter since results are never too far from true values of the sample to quantify.

    Quality of results is far more important than the intrinsic characteristic properties of procedure in terms of bias or precision.


Objectives of an analytical procedure4
Objectives of an analytical procedure guidelines on different aspects of QA

  • To develop a procedure without bias and error considerable cost; not acceptable strategy

  • Analyst has to take minimal risks, compatible with the analytical objectives (within reasonable time!!)

  • Set up acceptable maximum proportion of measurements that might be outside acceptance limits (±):

  • e.g. 5% or 20 % of measurements outside (±) as maximum risk.

  • inside triangles (next fig.)  space of acceptable procedures characterized by “true bias” dM and a “true precision” s²M

  • Acceptable procedures: 95, 80, 66% of measurements within ± 15% limits (recommendations Washington Conference)  proportion depends on objectives!!!


% measurements within guidelines on different aspects of QA± 15% bias-precision limits

Proc.3

20

(0,20)

15

Proc.4

True precision (%)

66%

(7,12)

10

(1,8)

80%

Proc.2

5

95%

(7,3)

-10

-5

0

5

10

0

Proc.1

True bias (%)


Objectives of an analytical procedure5
Objectives of an analytical procedure guidelines on different aspects of QA

  • Interior triangle = area of all analytical procedures of which 95% of result X should be included within acceptance limits (±), set according constraints of analytical domain

  • 2 other triangles: proportions of 80% and 66% of measurements included within ± (accept. limits)

  •  procedure with true bias =0 ; true precision = 15% : only 66% will fall within acceptance limits (±)

  •  procedure with true bias =0 ; true precision = 8% : 95% will fall within acceptance limits (±)


Objectives of an analytical procedure6
Objectives of an analytical procedure guidelines on different aspects of QA

  • Figure: procedures 1 and 2 located inside region of acceptance

  • this region guarantees that at least resp. 95% and 80% of the results are within acceptance limits (±)

  • for the same risk of the measurements outside acceptance limits, procedures 3 and 4 not considered as valid

  • for more important risk, procedures 3 and 4 could be valid.


Objectives of an analytical procedure7
Objectives of an analytical procedure guidelines on different aspects of QA

FURTHER CONCLUSION:

  • Procedure qualified as acceptable if:

    it “guarantees” that the difference between every sample measurement (x) and its “true value” (µT) is inside the predefined acceptance limits (± l)

  • In equation: P(x - µT  < l)  b (eq. 2)

    b = proportion of measurements inside acceptance limits

    l = acceptance limit, fixed a priori according objectives of the method

  • Expected proportion of measurements falling outside the acceptance limits  risk of an analytical procedure


Objective of the validation
Objective of the validation guidelines on different aspects of QA

  • What ?

    to give to the laboratories as well to the regulatory bodies“guarantees” that every single measurement performed in routine is close enough to the unknown “true value” of the sample: x - µT  < acceptable limit l

  • Objective of validation: not simply to obtain estimates of bias and precision; it is to evaluate these guarantees and risks

  • These estimates of bias and precision are required to evaluate risks


Objective of the validation1
Objective of the validation guidelines on different aspects of QA

  • With respect to this objective, 2 basic notions should be considered:

  • “close enough”(eq. 1) meaning that routine measure will be less than the acceptance limit λfrom its unknown “true value”

  • “guaranteed”,(eq. 2) meaning that it is very likely that analysis result will be close enough to the true unknown value.


Objective of the validation2
Objective of the validation guidelines on different aspects of QA

  • decision tools are needed giving “guarantees” that future measurements are reasonably inside acceptance limits 


Decision rules
Decision rules guidelines on different aspects of QA

  • Current position with respect to the decision rules used in the phase of validation  most of them based on use of the null hypothesis:

    H0 : bias = 0  H0 : relative bias = 0 %  H0 : recovery = 100 %

  • Bias = x - µT

  • Relative bias = 100 (x - µT)/µT

  • Recovery = 100 x/µT

  • A procedure wrongly declared adequate when the 95% C.I. of the average bias includes 0

  • Test inadequate in validation context of analytical procedures because decision based on computation of rejection criterion of Student t-test


Test based on H guidelines on different aspects of QA0 = bias = 0

20

(0,20)

Proc.3

PROCEDURES VALID

15

(7,12)

Proc.4

10

True precision (%)

(1,8)

Proc.2

5

(7,3)

NOT VALID

Proc.1

0

-15

-10

-5

5

10

15

0

True bias (%)


Decision rules1
Decision rules guidelines on different aspects of QA

  • According to the decision rule based on the null hypothesis H0 in fig.: procedures 2, 3 and 4 are valid and procedure 1 is rejected

  • But: procedure 1 shows reduced bias (+ 7%) and a small RSD (3%)  outside triangle: rejected !!

  • procedure 3 has high RSD (20%), procedure 4 has bias of 7% and RSD of 12%  accepted !!

  •  bad precision  large C.I.  contains 0 as bias value  method accepted

  •  good precision  small C.I.  may not contain 0 as bias value  method rejected

    null hypothesis H0 inadequate in analyt. validation


Test based on acceptance limits ( guidelines on different aspects of QA± 15%)

20

(0,20)

Proc.3

PROCEDURES NOT VALID

15

Proc.4

(7,12)

10

True precision (%)

(1,8)

Proc.2

5

β≥ 80%

(7,3)

0

-15

-10

-5

5

10

15

0

Proc.1

True bias (%)


Decision rules2
Decision rules guidelines on different aspects of QA

  • According to the decision rule based on use of acceptance limits triangle in fig. with acceptible valid procedures

  • Triangle in fig. corresponds to procedures with measurementproportion inside acceptance limits (±λ) ≥a priorichosen proportion (e.g. 80%) as given by equation: P(x - µT  < l)  b (eq. 2)

  •  more sensible decision rule: procedures with good precision  accepted

    bad precision  rejected

  • Biased procedure  small variance: acceptable !!

  • Procedure with higher variance  needs small bias


Decision rules accuracy profile
Decision rules: Accuracy profile guidelines on different aspects of QA

  • easy and visual decision rule: use of the accuracy profile within the acceptance limits (± l)

  • Accuracy profile constructed from the β-expectation intervals on the expected measurements:

    - allows to decide on capability of analytical procedure to give results inside ± l

    - describes dosage interval (range) in which the procedure is able to quantify with known accuracy and a fixed risk at the end of the validation

  • e.g. risk of 5% “guarantee” that 95/100 future measurements will be included in acceptance limits, fixed according requirements (1-2 % on bulk, 5 % on pharmaceut., 15 % in bioanalysis)


Decision rules3
Decision rules guidelines on different aspects of QA

  • Accuracy profile by concentration level (C1, C2, ...) obtained by computing β-expectation tolerance interval allows evaluating the proportion of expected measurements inside acceptance limits

  • This interval is obtained from available validated estimates of the bias and precision of the procedure (by concentration level)

  • This interval of measurements expected within level b (= proportion of measurements inside l) has b-expectation confidence limits


Decision rules4
Decision rules guidelines on different aspects of QA

  • If for each concentration level jβ-expectation toleranceinterval are included within acceptance limits  method accepted!

  • Tolerance interval calculation:

    - what matters is: the guarantee of the results, expected in the future by the same analytical procedure in routine

    - estimation of µj, s²B,j, s²W,j at every conc. j are used to estimate the expected proportion of observations within the predifined acceptance limits [-l,+l], i.e.:

    Eµ,s P[x - µT  < l]dM, sM  b


Calculation of expectation tolerance interval
Calculation of guidelines on different aspects of QAβ-expectation tolerance interval

  • estimated bias (mean added concentrations minus mean calculated concentrations)

  • j = conc. level

  • these statistical parameters (trueness, within/between precision) might be calculated for each concentration level from validation standards.


Calculation of expectation tolerance interval1
Calculation of guidelines on different aspects of QAβ-expectation tolerance interval

  • Calculation of the interval in which a proportion β of all samples with a certain real concentration is observed (method of Mee): β expectation tolerance interval:

ISO 5725-2: calculation of within and between variance


Calculation of expectation tolerance interval2
Calculation of guidelines on different aspects of QAβ-expectation tolerance interval

n = degrees of freedom (Satterthwaite):

p = number of series (days)

n = number of replicates per series

Qt = β quantile of the Student’s t-distribution with ν degrees of freedom


Calculation of expectation tolerance interval3
Calculation of guidelines on different aspects of QAβ-expectation tolerance interval

interval representing in % the region containing β % of analysis results for a certain concentration level j :

after rearrangement:


Calculation of expectation tolerance interval4
Calculation of guidelines on different aspects of QAβ-expectation tolerance interval

  • Interval consists of two terms:

    % bias +/- coefficient of variation for intermediate precision = expression of method accuracy

    method is accurate for this concentration level if obtained tolerance interval is included within acceptance limits [-λ,λ]


Accuracy profile
Accuracy profile guidelines on different aspects of QA

bias (%)

+ l

mean relat. bias

0

acceptance limits

concentration

bias limits of

confidence

- l

C1

C2

C3

C4

LLQ

ULQ

RANGE

dosage interval


Decision rules5
Decision rules guidelines on different aspects of QA

  • Estimates of bias and variance are essential to compute evaluation of the expected proportion of measurements within acceptance limits

  • Accuracy profile obtained by connecting the lower or upper limits of confidence (cfr. fig)

  • If a subsection (concentration range) falls outside the acceptance limits new limits of quantification be defined and a new dosage interval (Upper and Lower Limits of Quantification)


Decision rules conclusion
Decision rules (conclusion) guidelines on different aspects of QA

  • Accuracy profile represents limits “ULQ and LLQ” in agreement with definition of criterion:

    LLQ = smallest quantity of the substance that can be measured with defined accuracy

  • Accuracy profile as single decision tool:

  • Allows reconciling the objectives of the procedure and those of the validation

  • Allows to visually grasp the capacity of the procedure to fulfill its analytical objective


Validation protocols life cycle
Validation Protocols: Life Cycle guidelines on different aspects of QA

  • Validation has to be considered as an element intervening after the development of a new analytical procedure

  • Objective of procedure = to be used in routine

  • Usage in routine must be coupled with a quality control (QC) of which the 2 objectives are:

  • the validity of the found results on the unknown samples

  • the assessment of the continuity of the performances of the procedure at the time of its exploitation


Protocols in validation phase
Protocols in validation phase guidelines on different aspects of QA

Main objectives in validation phase:

  • demonstrate specificity/selectivity

  • validate the response function (or calibration model used in routine)

  • estimate precision (repeatability and intermediate precision), trueness, accuracy

  • validate the quantitation limits, validate the range (dosage interval); cfr. accuracy profile!

  • assess linearity of the analytical procedure (results directly proportional to concentration in the sample – cfr. definitions)


Protocols in validation phase1
Protocols in validation phase guidelines on different aspects of QA

  • preparation of calibration standards (CS) with fixed number of concentration levels and repetitions by level

  • preparation of the validation standards (VS) in the matrix; are independent samples

  • VS prepared and treated independenly as future samples  essential for good estimation of between-series variance.

  • To estimate intermediate precision, VS analyzed on different days, equipment and by different operators.

  • Validation phase is ultimate stage before exploitation;  allows to estimate procedure’s performances in the expected experimental conditions

     allows to check procedure’s capability to quantify unknown sample


Protocols in validation phase2
Protocols in validation phase guidelines on different aspects of QA

  • Question whether or not presence of a matrix effect.

  • If no matrix effect, question is which concentration levels will be used for calibration  apply described validation protocols (V1 and V2)

  • Evidence of matrix effect: apply protocol V5

  • In case of doubt: apply protocols V3 and V4 according to calibration levels (cfr.Table)

  • Which types of standards (CS and VS), concentration levels?

  • VS prepared in matrix and independent; must similate future samples



Description of protocol v1
Description of protocol V1 guidelines on different aspects of QA

Series 1

Series 2

Series 3

R

Calibration

standards

Validation

standards

(1)

(1)

Additional validation standards (linearity ICH)

(1)

Conc


Description of protocol v2
Description of protocol V2 guidelines on different aspects of QA

Series 1

Series 2

Series 3

R

Calibration

standards

(

)

(

)

(

)

Validation

standards

(1)

(1)

Additional validation standards (linearity ICH)

(1)

Conc


Description of protocol v3
Description of protocol V3 guidelines on different aspects of QA

Series 1

Series 2

Series 3

Calibration

Standards without matrix

R

Calibration

Standards within matrix

Validation

standards

(1)

(1)

Additional validation standards (linearity ICH)

(1)

Conc


Possible concentation levels by type of procedure e g 6 comparative procedures
Possible concentation levels by type of procedure (e.g. 6 comparative procedures)

  • Determination of single chemical substance; reference available or determination of active ingredient in a pharmaceutical speciality (matrix)

  • Determination of available synthesis impurity in an active substance or pharmaceutical speciality (matrix) at concentration levels > LOQ

  • Determination of available synthesis impurity in an active substance or pharmaceutical speciality (matrix) around impurity limit (impurity limit > LOQ)

  • Simultaneous determination of chemical substance and one of its non-available impurities in this substance or pharmaceutical speciality (use substance as tracer to allowed maximum concentration of impurity)


Possible concentation levels by type of procedure e g 6 comparative procedures1
Possible concentation levels by type of procedure (e.g. 6 comparative procedures)

  • Determination of active substance for measuring dissolution kinetics for a dry dosage form (matrix)

  • Determination of active ingredient and its metabolites in plasma (drugs), drug residues, ...

    WHICH CONCENTRATION LEVELS ?

     cfr. TABLE


Examples of possible concentration levels by type of procedure
Examples of possible concentration levels by type of procedure

LA = admitted limit; Cmax = max. conc.; Cmin = min. conc.


Protocols in validation phase3
Protocols in validation phase procedure

  • Identify relationship between response Y and concentration X using calibration standards (response function).

  • Regression models are fitted, accuracy profiles calculated, one model selected  decision about validity of the procedure of interest.

  • Model: depends on  procedure type (pharmaceutical, bio-analytical, immuno-assay)  fixed method objectives

  • Linear regression (origin or not) envisaged.

  • Mathematical transformations applied on X and Y

  • Quadratic regression may be useful


Protocols in validation phase4
Protocols in validation phase procedure

  • Back-calculation of estimated VS concentrations by series by  calibration curve equations

  • For each concentration level estimation of trueness and precision

     calculation of limitsforaccuracy : cfr. CIj (bias) (include large proportion of results)

     accuracy profile for each fitted model

  • Accuracy profile  visual decision tool to evaluate capability of the method  if not within pre-fixed acceptance limits:

    - restrict dosis range  new limits of quantification

    - extend acceptance limits (possible??)


ACCURACY PROFILES with same VALIDATION PROTOCOL (0.01 – 5.0 ng/ml)

A

15

B

15

quadratic regression

Bias (%)

Bias (%)

weighed linear regression

-15

-15

C

D

15

15

Bias (%)

Bias (%)

-15

linear regression

-15

linear regression throug 0

linear regression on log transformed data

E

F

linear regression on square root transformed data

15

15

Bias (%)

Bias (%)

-15

-15

Concentration

Concentration

0

1

2

3

4

5

0

1

2

3

4

5


Protocols in validation phase5
Protocols in validation phase 5.0 ng/ml)

  • Figure: Accuracy profiles for validation of dosing procedure of chemical substance in biological matrix.

  • Protocol V5 applied + some concentration levels

  • Essentially low levels  good estimation of LOQ

  • 2 of 6 response functions (A: quadratic regress. + B: weighed regression) answer objective: acceptance limits ± 15%

     accuracy profile allows to decide about method capability:

    Quantifiable dosing range with known accuracy: 0.01 – 5.0 ng/ml at risk  = 5%


Conclusions
CONCLUSIONS 5.0 ng/ml)

  • Lack of generalisation between different validation protocols  harmonized approach

  • Proposal to review objectives of the validation according to objectives of the analytical procedure

  • Distinction between diagnosis rules and decision rules

  • Objectives of validation not simply to obtain estimates of bias and precision but also:

  • To evaluate risks or confidences that any single measurement is close enough to unknown true value

  • Trueness, precision, linearity, ..., no longer sufficient to make these guarantees.


Conclusions1
CONCLUSIONS 5.0 ng/ml)

  • Adapted decision tool  accuracy profile of the analytical procedure, based on:

    • -expectation tolerance interval at each concentration level

    • concept of total error (bias + standard deviation)

  • Allows to bring together objectives of the procedure and those of validation

  • Allows to visually grasp the capacity of the procedure  to fulfil its objectives

     to control risk associated with its use in routine


References
References 5.0 ng/ml)

  • C. Hartmann et al., An analysis of the Washington Conference Report on bioanalytical method validation

    J. Pharm. Biomed. Anal., 12(11) (1994) 1337-1343

  • Ph. Hubert et al., The SFSTP guide on the validation of chromatographic methods for drug bioanalysis: from the Washington Conference to the laboratory.

    Anal. Chim. Acta, 391 (1999) 135-148

  • P. Chiap et al., Validation of an automated method for the liquid chromatographic determination of atenolol in plasma: application of a new validation protocol.

    Anal. Chim. Acta, 391 (1999) 227-238


References1
References 5.0 ng/ml)

  • B. Boulanger et al., An analysis of the SFSTP guide on validation of chromatographic bioanalytical methods: progress and limitations.

    J. Pharm. Biomed. Anal., 32 (2003) 753-765

  • Ph. Hubert et al., Validation des procédures analytiques quantitatives. Harmonisation des démarches.

    STP Pharma Pratiques, 13(3) (2003) 101-138

  • Ph. Hubert et al., Harmonization of strategies for the validation of quantitative analytical procedures. A SFSTP proposal – part I

    J. Pharm. Biomed. Anal., 36 (2004) 579-586


References2
References 5.0 ng/ml)

  • Ph. Hubert et al., Validation des procédures analytiques quantitatives. Harmonisation des démarches. Partie II - Statstiques

    STP Pharma Pratiques, 16(1) (2006) 28 – 58

  • Ph. Hubert et al., Validation des procédures analytiques quantitatives. Harmonisation des démarches. Partie III – Exemples d’application

    STP Pharma Pratiques, 16(2) (2006) 87 – 121

  • M. Feinberg et al., New advances in method validation and measurement uncertainty aimed at improving the quality of chemical data

    Anal. Bioanal. Chem 380 (2004) 502-514

  • M. Feinberg et al., A global approach to method validation and measurement uncertainty

    Accred. Qual. Assur 11 (2006) 3-9