An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Alarms Externalisation Dr Signouret Thomas

  2. Intensive care unit • Patients • Vital prognosiscommitted • Multiorganfailure • Cares : • Invasives, painfull, 24 h/24 • Life support equipment: ventilator, dyalisis, infusion pumps • Monitoring: 24 h/24 • Caregivers • Monitoring device

  3. Noise: World Health Organisation guidelines: 35→40 dB • Currentlyaverage noise: 60 → 70 dB • Alarms: 70 → 80 dB !!! • Staff conversations: 59 → 90 dB Bush-VishniacJ acoust soc Am 2005 • 10 à 25 % of alarms do induce a therapeutic modification Lawless ST Crit Care Med 1994 • Patients: • Disturbance • Tachycardia, hypertension • Pituitaryand adrenal gland stimulation XieCrit care 2009 • Nurses: Concentration, stress, Phenomen of alarm fatigue Morrison Crit Care Med 2003 Chambrin MC crit care 2001

  4. Link between noise and sleepdisturbance ? • Sleepdisturbance: • Slowerhealing • Cognitive function • Décreaseimmune response • ICU delirium : 20 →80 %  ↗lenght of stay ↗morbi-mortality Ely EW JAMA 2004 • Noise is one of manyfactors of sleepdisturbance: • 17→ 56 % XieCrit care 2009

  5. Historic Remove the cardiacs monitor alarmsfrom the room to the centrale station Close the patient door’s room Pb: What about the othersdevicesAlarm’s ? Need a translation of all the alarm in the samelanguage (HL7): Capsultech® Need to bring and allocate the alarm to the caregiver: Connexall®

  6. Aims of the Externalisation alarms • Decrease the noise due to the alarms in the patient room. • Allow the doorclosing: • Decrease the noise from the nursing station - 4 dB: Lawson Am J Crit Care 2010 • Overpressurelevelrespected in the unit • Aspergillose prevention R113: Référentiel structVdef2011 • Alarmssend to caregivers • Tracability

  7. Définition of gravitylevel • Cardiacarrest(vital priority): Vital emmergencywhoneed an immediatrespons of all the caregivers( nurses and doctor) of intensive care unit • Critical(highpriority ) : Urgent situation whorequire an immediatresponse of caregivers • Grave(medium priority) :Imminent riskwhorequires a rapidresponse of caregivers • Message(lowpriority) : Attention of medical staff isrequire Europeanstandard medicaldevices.EN 475:1995

  8. Mappingalarms

  9. Workflow

  10. The Daily Job of the Nurse

  11. Hardware connectivity

  12. Check devices connexion • Specificconnection by DIM Capsultech® on device RS232 outlet • Connection IT wires on DATACAPTOR®module for sending the information on the network

  13. Sélect and Assign the patients 303 304 305 301 302 R2.2 R2.1 • Connexall central station: • Assignement of each patient on an iPoDwhichisdedicated to a nurse

  14. Assignment

  15. Connexall Central stationMain Screen

  16. Alarms on mobile Device Arrêt cardiaque box 1 Acknowledgment

  17. Currently • Golive on a five beds unit after a training of all nurses of the unit • Connection of all devices • Good expression of the sonorAlarmswith right tonality, gravity and workflow • Visual alarms: Good rendering • Doorsclosedrecently • Weleft the sound in the box due to an IT server issue

  18. Alarm Externalisation Risk • Log out the devicesfrom the datacaptor • 4 or 5 network • Drager • Capsultech • Connexall • Hopital européen Marseille • Apple RegulatedMedicalDevices Class II FDA

  19. Whatweexpected ? Sécurity • A visual indicator of log out on connexall central station • An automatic detection of IT server issue by heart bit then an alarm message with a workflow to: • Drager IT engineer • Biomed • Doctor • Chief nurse • A new device who avoid apple network • Evaluation the quality of caregivers réaction

  20. And Also…: Dévellopment • Transitory (2min) Inhibition of the transmission of all alarm on the Connexall system when the Nurse active “a nursing function” • Showing the waves in live in case of alarm: Link patient watch • Sending vocal message with: • 1st : the sound characteristic of level of gravity • 2nd : the number of room • 3rd : the message of alarm

  21. Link on patient watch

  22. Evaluation Safety Efficacy Perception

  23. Safety • Description of differentsalarm : • Number of alarm / hour and /12 h • Number by device • Number by nurse • Number by gravity • Delay betweenalarm and acknowledgment • Delay betweenalarm and resolution

  24. Efficacy • Averageintensity noise duringday and night : • In the room • In the nurse station • Number of Peaklevel> 80 dB/ h duringday and night : • In the room • In the nurse station

  25. Perception • Patient questionnaire : • Sleepquality : Richards Campbell Sleep Questionnaire • Noise discomfort • Care giversquetionnaire : • Noise discomfort • Satisfaction

  26. A prospective open interventionnel studyévaluatea reducingstratégy of noise by an alarm notification system Primarypurpose: réduction of10 dB by currentlyaverage noise level 60 +/- 10 dB Weneed16 patients in eachstategies Power: 80 % alpha risk : 5 %

  27. Middle and Long terms Plan • Extended the externalisation of the alarm to other medical Device • ECMO, Bed, Labs, etc… • Extended the externalisation of the alarm to other medical specialties • Anesthesia : An anesthesiologist works on several OR • Cardiology • Special device to show the alarm: Samsung gear, Google Glasses

  28. Thankyou for you attention

  29. Functionnal Basics of Third-Party Alarm Notification Systems B MoormanBiomed Instrumentation and Technology 2011