XIX Congress of the European Society for Pediatric Neurosurgery Rome, May 6-9th 2004

XIX Congress of the European Society for Pediatric Neurosurgery Rome, May 6-9th 2004 Perspectives of the Shunt Technology: iValve and DigiShunt Aschoff A, Hashemi B, Scheihing M, Unterberg A, Kremer P

Download Presentation

XIX Congress of the European Society for Pediatric Neurosurgery Rome, May 6-9th 2004

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. XIX Congress of the European Society for Pediatric Neurosurgery Rome, May 6-9th 2004 Perspectives of the Shunt Technology: iValve and DigiShunt Aschoff A, Hashemi B, Scheihing M, Unterberg A, Kremer P University of Heidelberg, Department of Neurosurgerye-mail: Alfred_Aschoff@med.uni-heidelberg.de Oral presentation on Sunday, May 9th, 2004 Selectedslides

  2. Shunt-Revisions: Quantities Misplaced catheters Hydraulic Mismanagement Infections Disconnections

  3. Ventricular Catheters - Positions Prospective, 63 EVD-catheters 23.4% Anterior horn, ipsilateral 5.8% Lateral ventricle, ipsilateral 5.8% Anterior horn, contralateral 19.3% other positions in the ventricles 2.3% extraventricular intrathecal 2.3% in the brain parenchyma Weis N, Naff N, Hanley D: Accuracy of intraventricular catheter placement ... Poster 221, AANS 2/2003 Phoenix

  4. ±3o Catheter,ideal position= all holesfree in the ventricle Tolerance for lateral deviations ±3o only!

  5. Current Quotes of Shunt-Infections Valve Patients Procedures % n % n Pollack 99 Medos P 9.8 377 Kestle 00 OSV,DP,Delta8.4 367Zemack 01 Medos P 11.4 477 8.5 660 Lundkvist 01 Medos P 11.0 122 Götz 01 Medos P 2.8 143 Richards 01 multiple *2.6/6.0 12,950 (*Erst-OP) Vougioukas 01 OSV I+II 8.6 81 Kiefer 01 G-valves 1.7 120 Cochrane 02 multiple 8.6 12,106 Hanlo 03 OSV II 8.2 557 Vinchon 03 multiple 21.7 1564 6.6

  6. 1. Sterility in theOP-theatre 2. Antibacterial surface modification of implant 3. Systemic antibiotic prophylaxis

  7. 55 rabbits, prospective, randomized Ventricular catheter ± Rifampin (covalent) ± 107 Staph. epiderm., 105 Staph. aureus - Untreated catheters: 96 % Infections of catheters & brain - Rifampin-catheters:0 % Infections of implant p<0,001 Kockro, Aschoff et al. J Med Microbiol 49 (2000):441-450

  8. Simple slit- and diaphragm- valves, Orbis-Sigma: Inaccuracies & long-term-drifts common

  9. ASD, Delta, distal slit ... : Safety deficits Instable valve bodiesReflux …

  10. Adjustable Valves: Change of Technological Leadership ? Miethke ProGAVSophysa-Polaris+ gravitational valve Codman-Medos

  11. Adjustable Valves – Present Use Treatment of self-produced complications (preferebly overdrainage) Fine tuning for individual needs Psychotherapy Training to shunt-independence

  12. Adjustable Valves and Shunt-Removal 1. Stepwise increase of valve pressure to 200-400 mmH2O 2. Shunt removal after 1/2 - 2 years In 59% (71/120) successful!Takahashi Y (2001) Withdrawal of the shunt systems - clinical use of the programmable shunt system and its effcects in hydrocephalus in children. Child´s Nerv Syst 17:472-477

  13. Adjustable Valves – Use in Future Training to shunt-independence Fine tuning for individual needs Psychotherapy Treatment of self-produced complications

  14. G- + adjustable valve adjustable alone Ball- + g-valve 1 2 3

  15. Gravitational Valves - Problems - Complicated stock-keeping in OR 6 supplementary devives (e.g. Shunt-Assistant) 6 complete g-valves (e.g. PaediGAV) - In case of growth, adipositas, pregnancy, or inadequate pressure selection is under- or (residual) overdrainage not excluded. - No individual fine-tuning

  16. Gravitational Valves- Perspectives Adjustable g-Valves Percutaneous variation - Number of “active“ balls - Balls with different weights - Variation of the cone-geometry

  17. Shunt Technology in 2005 1. Adjustable valve 50-350 mmH2OStable in MRI, X-ray control not necessary 2. + gravitational valve, adjustable between 100-350 mmH2O, stable during jogging etc. 3. +low-flow-catheter (ID Ø 0.8 mm) 4. ICP-telemetry (sensor intradural) 5. Incorporated antibiotics

  18. Shunt Technology in 2010 - 2 intracranial microtransducers - 2 extrathecal transducers - 2 gravitational chip sensors - Storage chip: ICP of last month - Telemetry - Electronically controlled valve - CPU, programmable for an “event- controlled” shunt function and training to shunt-independence