An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Application of new photo-detectors to calorimetry and medical physics Nicola D’Ascenzo, Erika Garutti University of Hamburg – DESY Alexander Tadday University of Heidelberg

  2. The Geiger Mode Avalanche Photodiode Why is this detector so interesting ? • High gain (Typically 105 - 106) • Low bias voltage (30 V -70 V) • Small area (1 mm2) • Up to 1600 pixels per mm2 • Not sensitive to magnetic field Devices from MEPHI, Hamamatsu and SensL currently avalaible in DESY: Good possibility of comparison and testing PD07 Kobe

  3. ILC Physics: challenges in calorimetry • Electron-positron accelerator • Energy: 500 GeV up to 1 TeV • ILC physics places premium on jet measurements and flavoure tagging, in an environment where event reconstruction is possible • tth  8 jets • hZ  2l + 2 jets, 4 jets • hhZ  2l + 4 jets, 6jets Aim to fully reconstruct final state • + SUSY, quark, τ tagging, lepton/hadron id • Missing energy signatures of SUSY and extra-dimensions This requires a calorimeter which is: COMPACT – Inside the magnetic coil HIGH GRANULAR – Imaging calorimeter HERMETIC PD07 Kobe

  4. HCAL ECAL TPC The scintillator tile hadron calorimeter • Sampling calorimeter : absorber stainless steel, active medium plastic organic scintillator. • High granular, compact, hermetic. • Along the magnetic field direction it extends in -220 cm < z <220 cm • It covers the detector along the radius in the region 180 cm< r < 290 cm • The test beam prototype: 1 m3 . • Test beam prototype : 38 layers, 216 channel per layer. • Fully equipped with MEPHI/PULSAR SiPM PD07 Kobe

  5. SiPM: a good photodetector for hadron calorimetry Test Beam CERN. 10 GeV Pion in the HCAL The minimum signal in 1 cell is a m.i.p. • The good SiPM for this application: • Low noise above the suggested cut on the mip spectrum (Controlled dark rate and cross talk) • Light Yield not exceeding the saturation behaviour of the device, but high enough to detect a m.i.p. with at least 95% efficiency. (good PDE in the light peak spectrum) • Stable and controlled production • Compactness, unsensitivity to magnetic field up to 5 T, radiation hardness ... The test beam prototype, fully equipped woth MEPHI SiPMs, is showing good performances for a long time operation of the device: the high granular calorimeter operated with SiPMs is starting to be a suitable option. Test Beam data are under analysis PD07 Kobe

  6. New approach to the calorimeter design:the direct readout of the scintillator R&D Present status • Scintillator tile (Uniplast, Vladimir (RU)) • Green wavelength shifter fiber (Kurakay) • Mirror • In order to: • Assure Light yield (MEPHI SiPM is green sensitive) • Assure Light collection uniformity • The direct readout is easier to extend to • a large scale production. • What is the solution which respects the required bounds • on the parameters: • Blue Sensitive? • Green sensitive and large? PD07 Kobe

  7. The MPPC produced by Hamamatsu Blue sensitive Multi Pixel Photon Counter PD07 Kobe

  8. Direct readout with MPPC • Readout of a plastic scintillator tile directly and comparison with mediated green fiber • Tile wrapped with a 3M reflector. In case of the direct readout, only a window of 1mm x 1mm is open in front of the MPPC • Source: Ru106 • Integration gate 80 ns • External trigger (Photomultiplier tube) • Reproducibility of the measurement +- 3% • systematic (coupling, positioning etc ...) PD07 Kobe

  9. 1600 px 400 px Direct readout with MPPC : Most Probable Value of the mip response spectrum Direct (Blue dots) and Wavelength shifter mediated readout (Green dots) . Light Yield in the calorimeterprototype PD07 Kobe

  10. Direct readout with MPPC: mip collection efficiency 1600 px Cut at3KHz 400 px The cut at 3 KHz (as in the actual prototype) corresponds to a low number of pixels (low dark rate and cross talk). The efficiency is, then, always more than 97 %. PD07 Kobe

  11. 3x3 MPPC: direct readout Ru106 response • About 50 p.e. Per MPV • Good single photoelectron resolution! LED Light PD07 Kobe

  12. The direct readout of MPPC: discussion • MPPCs are a good starting point for the discussion of the direct readout in the hadron calorimetry: • High PDE in the blue emission of the scintillator: good m.i.p. signal determination. • Low dark rate and cross talk: cut for noise even for the stricter requirement of ILC (300 Hz), without loosing m.i.p efficiency (>95%) • High dinamic range (up to 1600 pixels) • Some points under investigation: • Uniformity of the light collection • Done in laboratory setup in ITEP and LNF-FRASCATI • Real prototype (ECAL): Japan • Radiation hardness • Under investigation in ITEP • For a thinner scintillator solution, the tested samples seem to have a too low Light Yield in this readout configuration. PD07 Kobe

  13. Application of Geiger Mode Avalanche Photodiodes to PET: motivations and goals • Why applying these photodetectors to PET? • Compact • Not sensitive to magnetic field (Combination with MRI) • Coupling with cristal (1x1 mm2): high space resolution. • The calorimeter is just a big PET machine: • A calorimeter measures the energy deposited. • The hadron calorimeter has a very high granularity: imaging detector! • It is only needed to: • Change scintillator (from organic to inorganic) • Change granularity (from 3x3 cm2 to ~1x1 mm2) PD07 Kobe

  14. Energy and time resolution in PET The scattered coincidences can be cut with a good energy resolution: only the photoelectric peak is allowed. Good PDE in the blue region. The timing resolution of the module restricts the analysis area. 500 ps (FWHM) -> 7.5 cm (FWHM) PD07 Kobe

  15. Application of MPPC in PET: setup 2 LSO Crystals (Hilger Crystals, emission peak at 420 nm, 40 ns decay time) of 1mm x 1mm x 15mm (3mm x3mmx15mm) 2 LFS crystals (Lebedev institute, Moscow, new crystal, with few specifications) of 3mm x3mmx15mm Read out by 1x1 mm (3x3 mm) Hamamatu MPPC, coupled 1:1 with the photodetector. No amplification needed !!! Timing with 4 GHz Oscilloscope (50 ps resolution) Reproducibility of the measurement : 5% (Coupling is the main systematic problem) & QDC PD07 Kobe

  16. Application of MPPC in PET: energy resolution (LSO) 3mmx3mm LSO crystal Coupled with a 3mmx3mm MPPC 1mmx1mm LSO crystal Coupled with a 1mmx1mm MPPC Resolution (FWHM) ~10 % ~500 p.e. Resolution (FWHM) ~14 % ~300 p.e. The typical value of the single LSO crystal energy resolution (at 511 KeV) measured with the traditional photomultiplier tube is 10 %. PD07 Kobe

  17. Energy resolution: LSO and LFS 3mmx3mm LFS crystal Coupled with a 3mmx3mm MPPC 3mmx3mm LSO crystal Coupled with a 3mmx3mm MPPC Resolution (FWHM) ~11 % Resolution (FWHM) ~10 % Spectra acquired at the same voltage. The difference of the resolution is inside the reproducibility of the measurement. The crystals look equivalent. PD07 Kobe

  18. Timing resolution in PET 2 1 3mmx3mm LSO crystal Coupled with a 3mmx3mm MPPC 1 : Define a coincidence threshold 2 : Define a timing threshold N_cut Dt = t(s1>N_cut) – t(s2>N_cut) sigma150 ps expected from the formula. 150-200 ps typical with PM tube (Bill Moses, LBL) A background is superimposed and ruins the time resolution... Mean1: 4.42 ± 0.02 Sigma1: 0.171 ± 0.020 Mean2: 4.39 ± 0.01 Sigma2: 0.639± 0.015 PD07 Kobe

  19. Application of MPPC in PET:timing resolution (1x1) Mean1: 5.36 ± 0.01 Sigma1: 0.276 ± 0.005 Mean2: 4.9 ± 10.3 Sigma2: 0.570± 0.050 Mean1: 4.42 ± 0.02 Sigma1: 0.295 ± 0.020 Mean2: 4.39 ± 0.01 Sigma2: 0.570± 0.050 Coincidence At ~30 p.e. Coincidence At ~10 p.e. 1mmx1mm LSO crystal Coupled with a 1mmx1mm MPPC 1mmx1mm LSO crystal Coupled with a 1mmx1mm MPPC Mean1: 5.39 ± 0.02 Sigma1: 0.286 ± 0.020 Coincidence At ~50 p.e. Background Signal 1mmx1mm LSO crystal Coupled with a 1mmx1mm MPPC Compton Photoelectric PD07 Kobe

  20. MPPC in PET : discussion • MPPC show for the first time a very promising possibility of application of Geiger Mode avalnche photodiodes to Positron Emission Tomography: • Energy resolution 10% with LSO DONE!!!! • Time resolution (500 ps FWHM) ALMOST THERE!!!! • More studies needed: • Study in magnetic field, combined with Magnetic Resonance. • What crystal ? WORK IN PROGRESS (LSO, LSF, YAP...) • Study spatial resolution of a matrix • Buid a prototype and verify the concept PD07 Kobe

  21. Outlook ... A big effort is running , all over the world, in order to steer the production of the Geiger Mode avalanche photodiodes: • Blue sensitivity • Good dynamic range • Low noise This is not more a dream... The Particle Physics and the Medical Physics communities are profiting of a continous interchange... PD07 Kobe