1 / 14

Review of spectroscopic diagnostic at COMPASS tokamak

Review of spectroscopic diagnostic at COMPASS tokamak. Presented by Diana Naydenkova Institute of Plasma Physics AS CR, v.v.i., Association EURATOM/IPP.CR Za Slovankou 3, 182 00 Prague 8 Thanks for contributions to COMPASS team & collaborators. ADAS Workshop 2012 / 23 – 25 September 2012.

mckenzie
Download Presentation

Review of spectroscopic diagnostic at COMPASS tokamak

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Review of spectroscopic diagnostic at COMPASS tokamak Presented by Diana Naydenkova Institute of Plasma Physics AS CR, v.v.i., Association EURATOM/IPP.CR Za Slovankou 3, 182 00 Prague 8 Thanks for contributions to COMPASS team & collaborators ADAS Workshop 2012/23 – 25 September 2012

  2. Outline • COMPASS tokamak • Spectroscopic diagnostic • Thomson scattering • Two fast VIS cameras • Plasma rotation • Bolometry • Soft X-rays • Visible light detection • Spectrometer for VIS and UV • Bremsstrahlung • Summary Review of spectroscopic diagnostic at COMPASS tokamak / ADAS Workshop 2012/23 – 25 September 2012 01

  3. COMPASS tokamak Major radius[m] 0.56 Minor radius [m] 0.23 x 0.38 Plasma current [kA] <180 (400) Magnetic field [T] <1.8 (2.1) Triangularity ~ 0.5 Elongation ~ 1.6 Pulse length [s] < 0.3 (0.5) Review of spectroscopic diagnostic at COMPASS tokamak / ADAS Workshop 2012/23 – 25 September 2012 01

  4. Spectroscopic diagnostic Review of spectroscopic diagnostic at COMPASS tokamak / ADAS Workshop 2012/23 – 25 September 2012 02

  5. Laser Tokamak vesel edge & core collection optics beam dump Thomson scattering Nd:YAG (1.5 J, 30 Hz, 1064nm) Fast digitizers Optical fibers bundle Polychromator Detectors (Avalanche Photodiodes) • Spatial resolution: • Central ~10mm • Edge ~4mm Optical delay line Review of spectroscopic diagnostic at COMPASS tokamak / ADAS Workshop 2012/23 – 25 September 2012 03

  6. Fast VIS camera Cypress LUPA-1300 CMOS sensor Resolution: 1280 x 1024 Speed: 541 frames/sec In collaboration with KFKI-RMKI and SIU EURATOM/HAS Review of spectroscopic diagnostic at COMPASS tokamak / ADAS Workshop 2012/23 – 25 September 2012 02

  7. Fast VIS camera Separatrix Cypress LUPA-1300 CMOS sensor Resolution: 1280 x 1024 Speed: 541 frames/sec In collaboration with KFKI-RMKI and SIU EURATOM/HAS Review of spectroscopic diagnostic at COMPASS tokamak / ADAS Workshop 2012/23 – 25 September 2012 02

  8. Plasma rotation Resolution 0.004nm Range ~8nm Readout rate <10MHz QE >90% In collaboration with IPP, EURATOM/IST Review of spectroscopic diagnostic at COMPASS tokamak / ADAS Workshop 2012/23 – 25 September 2012 02

  9. Tomographic system AXUV photodiodes (bolometers) temporal resolution 1MHz spectral range: 7eV – 6 keV Photodiodes (windowless) temporal resolution 1MHz spectral range: 0.5 (Be foil) – 10 keV Review of spectroscopic diagnostic at COMPASS tokamak / ADAS Workshop 2012/23 – 25 September 2012 02

  10. Visible light measurements • two angular port plugs • interference filters(line selection/Zeff) with FWHM of 10 nm/2nm • optical fibers (high OH silica) of diameter200mm, length~20 m • photomultipliers /two photodiode arrays (35 channels, 1ms) Review of spectroscopic diagnostic at COMPASS tokamak / ADAS Workshop 2012/23 – 25 September 2012 02

  11. Minispectrometers Review of spectroscopic diagnostic at COMPASS tokamak / ADAS Workshop 2012/23 – 25 September 2012 02

  12. Minispectrometers Review of spectroscopic diagnostic at COMPASS tokamak / ADAS Workshop 2012/23 – 25 September 2012 02

  13. Spectroscopic diagnostic ~523nm ~5 mV noise C III H a bremsstrahlung All VIS Review of spectroscopic diagnostic at COMPASS tokamak / ADAS Workshop 2012/23 – 25 September 2012 02

  14. Summary • Plasma rotation • Zeff • Impurities influx and concentration • L-H transition • Pedestal parameters • Charge exchange spectroscopy • IR camera Review of spectroscopic diagnostic at COMPASS tokamak / ADAS Workshop 2012/23 – 25 September 2012 02