An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Is Grey the New Black?Image Display Optimisation & Consistency for Softcopy ReportingMark BarnfieldSt James’s University Hospital, Leeds

  2. Softcopy Concerns… • Softcopy more often becoming last component in the ‘Imaging chain’. • In NMed chain can cost in excess of £300K – undertake well developed QA protocols. • … however, workstation monitors relatively inexpensive ~£300 – possible weak link with little or no QA system employed.

  3. Image Quality ApproachSt James’s NMed • Optimisation (currently qualitative) • To improve presentation of radiological images and ensure that important image information is not lost during visual inspection. • Consistency • Images should appear similar on different displays or on the same display at different times.

  4. First step - Display Preparation • Monitor Cleanliness • Display quality is compromised by dust, fingerprints, ink, etc. • Reduction in luminance. • Possible misinterpretation. • Needs cleaning on regular basis and on as-needed basis. • Lint free cloth to reduce static. • Never spray directly onto monitor. • Manufacturer may recommend special cleaner for monitors coated with antireflective or antiglare coatings.

  5. Picture Boarder Linearity & Geometric distortion Uniformity Mid-band streaking Brightness & Contrast Resolution response Focus The SMPTE Test Pattern Ref:Radiology 1985; 154: 519-527

  6. An Optimal Contrast Range Brightness & Contrast Adjustment • Set to ~50% and back off • Bring up until black just turns grey and back off fraction • Set to give reasonably bright white level but avoid loss of image definition The brightness and contrast of yourmonitor are adequately set if the inner 5% squares at both ends of gray scale are visible.

  7. Observations • 17” Iiyama – Vision Master 403 • Visual • Geometric (Aspect ratio) distortion – squares  rectangles • Barrelling • Also observed pincushion on my PC! • Refreshverticalrate66Hz - <72Hz some observers can perceive flicker and greater eye stain is possible. • Known that nearly all parameters are dependent on position on screen - degrading from centre.

  8. DICOM Part14: (GSDF) Grey-Scale Standard Display Function The Consistent image • Aims to mathematically define a display function for all image presentation systems. • Aims for similarity in grey-scale perception between different display systems. • Not a performance or image display standard. • It does not define luminance range.

  9. Measuring Monitor Characteristic CurveAnnex D - DICOM pt14 pp27-36 Background (20% of max lum) Target (10% area of screen) Photometer (Light-O-Meter P10 Std) Step target through greyscale range

  10. Curve Mapping

  11. LUT set-up verification • 5% & 95% contrast equally perceptible. • Linear perception along the colour bar.

  12. Effect of brightness & contrast setting on display characteristic curves - PCs

  13. Reproducibility of monitor set-up using SMPTE test pattern NB. A – B done as consecutive measurements, C – D on separate occasions over a week.

  14. Proposed Frequency of QC Tests • Weekly • Monitor preparation • Brightness & contrast adjustment • Monthly • Geometry • Spatial resolution • Quarterly • Spatial uniformity • Veiling glare

  15. Acknowledgements I would like to thank, Sarah Jervis for allowing the use of her SMPTE and DICOM 14 Matlab programs; David Brettle for guidance and technical support.