Methods for LTQ Orbitrap A Guided Tour with Examples

Methods for LTQ Orbitrap A Guided Tour with Examples. Dr. Michaela Scigelova. Outline. Data acquisition strategies Method setup and examples OT_LTQ_Big6 MS in Orbi + 6 MS/MS in ion trap OT_3OT MS in Orbi + 3 MS/MS in Orbi NL_MS3_AccMass Phosphopeptides with accurate NL

leona
Download Presentation

Methods for LTQ Orbitrap A Guided Tour with Examples

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Methods for LTQ OrbitrapA Guided Tour with Examples Dr. Michaela Scigelova

  2. Outline • Data acquisition strategies • Method setup and examples • OT_LTQ_Big6 • MS in Orbi + 6 MS/MS in ion trap • OT_3OT • MS in Orbi + 3 MS/MS in Orbi • NL_MS3_AccMass • Phosphopeptides with accurate NL • MS in Orbi, MS2 in Orbi, and MS3 in ion trap • OT_MSA • Phosphopeptides • ‘composite’ MS2 and MS3 spectrum = MSA • MS in Orbi, MSA in LTQ

  3. What is the objective of the analysis? • One or more of these: • Maximise the protein ID • Boost the sequence coverage • Phosphorylation • De novo sequencing • Quantitation

  4. Data Acquisition Strategies • Parallel mode • Survey MS in Orbitrap • MS/MS in LTQ For highest protein ID and coverage assuming very complex mixtures • Serial mode • Survey MS in Orbitrap • MS/MS in Orbitrap For de novo sequencing, PTM discovery

  5. 0.9 s 0.1 s 0.35 s 0.2 s Mass Accuracy and Resolution Settings Parallel Mode Serial Mode

  6. AGC settings LTQ • Full MS 3.00e+04 IT 50ms • SIM 1.00e+04 IT 100ms • MSn 1.00e+04 IT 100ms

  7. AGC settings orbitrap • Full MS 5.00e+05 - 1.00e+06 IT 500ms • SIM 1.00e+05 - 2.00e+05 IT 500ms • MSn 1.00e+05 - 2.0e+05 IT 500ms

  8. Settings for HCD, Lock mass • Act.Q 0.13-0.14 • Normalized Collision Energy 60-70 • specify the FT lock mass abundance (%) • Diagnostics/Set Device • Default value 10%

  9. OT_LTQ_Big6 MS in Orbi + 6 MS/MS in ion trap

  10. Parallel Mode – Method Setup • Step-by-step guide • Method BIG 6 • Tune settings • Max fill time 100 ms • Number of microscans 1

  11. Parallel Method – starting the method

  12. Parallel Method – Survey MS

  13. Parallel Method – Add another 6 scan events

  14. Parallel Method – BIG 6 Save yourself the typing: use copy and paste!

  15. Can be used for directing MS/MS to a preferred region – e.g. 800-1300 in glycopeptide analysis Parallel Method – DDA Scan event 2

  16. Parallel Method – DDA Scan event 2

  17. Ignore Parallel Method – DDA Scan event 2

  18. Parallel Method – DDA Scan event 2 This is the key to PARALLEL mode

  19. Parallel Method – DDA Scan event 2

  20. Parallel Method – DDA Scan event 2 Import lists of target parent ions (parent list) or contaminants (reject list)

  21. Reject ‘1+’ and ‘unassigned’ for max number of PROTEIN ID Parallel Method – DDA Scan event 2

  22. Leave unticked if going for max sequence COVERAGE Parallel Method – DDA Scan event 2

  23. Parallel Method – DDA Scan event 2 Even better COVERAGE: if you inject sample 3 times and set the following:

  24. Parallel Method – DDA Scan event 2

  25. Parallel Method – DDA Scan event 3 Default settings: WRONG!!!!!

  26. Parallel Method – DDA Scan event 3

  27. Parallel Method – DDA Scan event 4

  28. Parallel Method – DDA Scan event 5

  29. Parallel Method – DDA Scan event 6

  30. Parallel Method – DDA Scan event 7

  31. When you have built your method.. Check it!

  32. Who Is Working this Way? • John Yates (Scribbs Inst.) • Ruedi Aebersold (ETH Basel) • David Goodlet (ISB Washington) Yates et al., AC 2006, 78, 493-450

  33. OT_3OT MS in Orbi + 3 MS/MS fragmented in LTQ and measured in OT

  34. Can resolve 4+ ions Serial Method Setup • Full scan in Orbi + 3 MS/MS in Orbi • Tune settings • Max fill time 100 ms • Number of microscans 2 to improve S/N in MSn 0.35 s 0.2 s

  35. Serial Method – Full scan MS

  36. Serial Method – MS + 3 MS/MS Save yourself the typing: use copy and paste!

  37. Serial Method – Major change in ‘Current Segment’ There is NO box ticked!

  38. Serial Method – DDA Scan event 2

  39. Serial Method – DDA Scan event 3

  40. Serial Method – DDA Scan event 4

  41. Who Is Doing It This Way? • Matthias Mann (MPI Martinsreid) • Lock mass ‘trick’ • Divide the mass range is sections • Roman Zubarev (Uppsala) Olsen et al., MPC 2005, 4, 2010-2021

  42. Want ppb Mass Accuracy? • Use Lock mass in the method Olsen et al., MPC 2005, 4, 2010-2021

  43. Lock Mass - Caveats • Lock mass should be present in your analysis all the time • Can be a background ion • Or spike in a lock mass solution via a Y-connector (P-773,17 nL swept volume) with fused silica tubing attached to a syringe pump • Use multiple lock masses • For calibrating MS/MS spectra the lock mass closest to your parent mass of interest is taken (good for peptides) • Once you fill in the lock mass dialog box, you can export it, save it, and import it to other methods

  44. NL_MS3_AccMass MS in Orbi + MS/MS in Orbi + MS3 in LTQ

  45. Strategy for Phosphopeptides • Enrichment!!! • Use prominent neutral loss of phosphate in MS2 to trigger MS3 • Accurately defined neutral loss • Use Orbi for MS and MS2 • Cuts down on false positive MS3 triggers • MS3 is highly informative for peptide sequence and PTM position • Data review is fast and easy – search MS3 scans only • Thorough search (MS2 + MS3) provides high confidence ID Olsen and Mann, PNAS 2004, 101, 13417-22

  46. NL MS3 Method – starting the method

  47. NL MS3 Method – Survey MS

  48. NL MS3 Method – MS2

  49. Accurate NL determination avoids false triggers of MS3  more efficient analysis NL MS3 Method – MS2 NOTE: phosphopeptide analysis

  50. NL MS3 Method – MS2 NOTE: phosphopeptide analysis